Search

Find a vulnerability

Search criteria

    8 vulnerabilities found for symbiq_infusion_system_firmware by pifzer

    CVE-2015-3956 (GCVE-0-2015-3956)

    Vulnerability from nvd – Published: 2019-03-25 17:44 – Updated: 2024-08-06 06:04
    VLAI
    Summary
    Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
    Severity
    No CVSS data available.
    CWE
    • CWE-345 - Insufficient verification of data authenticity CWE-345
    Assigner
    References
    Date Public
    2015-06-10 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-06T06:04:02.927Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Plum A+ Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.4"
                }
              ]
            },
            {
              "product": "Plum A+3 Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.6"
                }
              ]
            },
            {
              "product": "Symbiq Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 3.13"
                }
              ]
            }
          ],
          "datePublic": "2015-06-10T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-345",
                  "description": "Insufficient verification of data authenticity CWE-345",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-03-25T17:44:44.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2015-3956",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Plum A+ Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.4"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Plum A+3 Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.6"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Symbiq Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 3.13"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "Hospira"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "Insufficient verification of data authenticity CWE-345"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2015-3956",
        "datePublished": "2019-03-25T17:44:44.000Z",
        "dateReserved": "2015-05-12T00:00:00.000Z",
        "dateUpdated": "2024-08-06T06:04:02.927Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2015-3954 (GCVE-0-2015-3954)

    Vulnerability from nvd – Published: 2019-03-25 16:12 – Updated: 2024-08-06 06:04
    VLAI
    Summary
    Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
    Severity
    No CVSS data available.
    CWE
    • CWE-285 - Improper authorization CWE-285
    Assigner
    References
    Date Public
    2015-06-10 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-06T06:04:01.128Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Plum A+ Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.4"
                }
              ]
            },
            {
              "product": "Plum A+3 Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.6"
                }
              ]
            },
            {
              "product": "Symbiq Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 3.13"
                }
              ]
            }
          ],
          "datePublic": "2015-06-10T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-285",
                  "description": "Improper authorization CWE-285",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-03-25T16:12:01.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2015-3954",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Plum A+ Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.4"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Plum A+3 Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.6"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Symbiq Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 3.13"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "Hospira"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "Improper authorization CWE-285"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2015-3954",
        "datePublished": "2019-03-25T16:12:01.000Z",
        "dateReserved": "2015-05-12T00:00:00.000Z",
        "dateUpdated": "2024-08-06T06:04:01.128Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2015-3953 (GCVE-0-2015-3953)

    Vulnerability from nvd – Published: 2019-03-25 16:02 – Updated: 2024-08-06 06:04
    VLAI
    Summary
    Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
    Severity
    No CVSS data available.
    CWE
    • CWE-259 - Use of hard-coded password CWE-259
    Assigner
    References
    Date Public
    2015-06-10 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-06T06:04:00.957Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Plum A+ Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.4"
                }
              ]
            },
            {
              "product": "Plum A+3 Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.6"
                }
              ]
            },
            {
              "product": "Symbiq Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 3.13"
                }
              ]
            }
          ],
          "datePublic": "2015-06-10T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-259",
                  "description": "Use of hard-coded password CWE-259",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-03-25T16:02:25.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2015-3953",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Plum A+ Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.4"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Plum A+3 Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.6"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Symbiq Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 3.13"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "Hospira"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "Use of hard-coded password CWE-259"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2015-3953",
        "datePublished": "2019-03-25T16:02:25.000Z",
        "dateReserved": "2015-05-12T00:00:00.000Z",
        "dateUpdated": "2024-08-06T06:04:00.957Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2015-3952 (GCVE-0-2015-3952)

    Vulnerability from nvd – Published: 2019-03-25 15:42 – Updated: 2024-08-06 06:04
    VLAI
    Summary
    Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
    Severity
    No CVSS data available.
    CWE
    • CWE-312 - Cleartext storage of sensitive information CWE-312
    Assigner
    References
    Date Public
    2015-06-10 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-06T06:04:01.121Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Plum A+ Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.4"
                }
              ]
            },
            {
              "product": "Plum A+3 Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.6"
                }
              ]
            },
            {
              "product": "Symbiq Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 3.13"
                }
              ]
            }
          ],
          "datePublic": "2015-06-10T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-312",
                  "description": "Cleartext storage of sensitive information CWE-312",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-03-25T15:42:39.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2015-3952",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Plum A+ Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.4"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Plum A+3 Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.6"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Symbiq Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 3.13"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "Hospira"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "Cleartext storage of sensitive information CWE-312"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2015-3952",
        "datePublished": "2019-03-25T15:42:39.000Z",
        "dateReserved": "2015-05-12T00:00:00.000Z",
        "dateUpdated": "2024-08-06T06:04:01.121Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2015-3956 (GCVE-0-2015-3956)

    Vulnerability from cvelistv5 – Published: 2019-03-25 17:44 – Updated: 2024-08-06 06:04
    VLAI
    Summary
    Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
    Severity
    No CVSS data available.
    CWE
    • CWE-345 - Insufficient verification of data authenticity CWE-345
    Assigner
    References
    Date Public
    2015-06-10 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-06T06:04:02.927Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Plum A+ Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.4"
                }
              ]
            },
            {
              "product": "Plum A+3 Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.6"
                }
              ]
            },
            {
              "product": "Symbiq Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 3.13"
                }
              ]
            }
          ],
          "datePublic": "2015-06-10T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-345",
                  "description": "Insufficient verification of data authenticity CWE-345",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-03-25T17:44:44.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2015-3956",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Plum A+ Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.4"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Plum A+3 Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.6"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Symbiq Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 3.13"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "Hospira"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "Insufficient verification of data authenticity CWE-345"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2015-3956",
        "datePublished": "2019-03-25T17:44:44.000Z",
        "dateReserved": "2015-05-12T00:00:00.000Z",
        "dateUpdated": "2024-08-06T06:04:02.927Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2015-3954 (GCVE-0-2015-3954)

    Vulnerability from cvelistv5 – Published: 2019-03-25 16:12 – Updated: 2024-08-06 06:04
    VLAI
    Summary
    Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
    Severity
    No CVSS data available.
    CWE
    • CWE-285 - Improper authorization CWE-285
    Assigner
    References
    Date Public
    2015-06-10 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-06T06:04:01.128Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Plum A+ Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.4"
                }
              ]
            },
            {
              "product": "Plum A+3 Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.6"
                }
              ]
            },
            {
              "product": "Symbiq Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 3.13"
                }
              ]
            }
          ],
          "datePublic": "2015-06-10T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-285",
                  "description": "Improper authorization CWE-285",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-03-25T16:12:01.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2015-3954",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Plum A+ Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.4"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Plum A+3 Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.6"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Symbiq Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 3.13"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "Hospira"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "Improper authorization CWE-285"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2015-3954",
        "datePublished": "2019-03-25T16:12:01.000Z",
        "dateReserved": "2015-05-12T00:00:00.000Z",
        "dateUpdated": "2024-08-06T06:04:01.128Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2015-3953 (GCVE-0-2015-3953)

    Vulnerability from cvelistv5 – Published: 2019-03-25 16:02 – Updated: 2024-08-06 06:04
    VLAI
    Summary
    Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
    Severity
    No CVSS data available.
    CWE
    • CWE-259 - Use of hard-coded password CWE-259
    Assigner
    References
    Date Public
    2015-06-10 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-06T06:04:00.957Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Plum A+ Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.4"
                }
              ]
            },
            {
              "product": "Plum A+3 Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.6"
                }
              ]
            },
            {
              "product": "Symbiq Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 3.13"
                }
              ]
            }
          ],
          "datePublic": "2015-06-10T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-259",
                  "description": "Use of hard-coded password CWE-259",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-03-25T16:02:25.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2015-3953",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Plum A+ Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.4"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Plum A+3 Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.6"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Symbiq Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 3.13"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "Hospira"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "Use of hard-coded password CWE-259"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2015-3953",
        "datePublished": "2019-03-25T16:02:25.000Z",
        "dateReserved": "2015-05-12T00:00:00.000Z",
        "dateUpdated": "2024-08-06T06:04:00.957Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2015-3952 (GCVE-0-2015-3952)

    Vulnerability from cvelistv5 – Published: 2019-03-25 15:42 – Updated: 2024-08-06 06:04
    VLAI
    Summary
    Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
    Severity
    No CVSS data available.
    CWE
    • CWE-312 - Cleartext storage of sensitive information CWE-312
    Assigner
    References
    Date Public
    2015-06-10 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-06T06:04:01.121Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Plum A+ Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.4"
                }
              ]
            },
            {
              "product": "Plum A+3 Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 13.6"
                }
              ]
            },
            {
              "product": "Symbiq Infusion System",
              "vendor": "Hospira",
              "versions": [
                {
                  "status": "affected",
                  "version": "\u003c= 3.13"
                }
              ]
            }
          ],
          "datePublic": "2015-06-10T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-312",
                  "description": "Cleartext storage of sensitive information CWE-312",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-03-25T15:42:39.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2015-3952",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Plum A+ Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.4"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Plum A+3 Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 13.6"
                              }
                            ]
                          }
                        },
                        {
                          "product_name": "Symbiq Infusion System",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "\u003c= 3.13"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "Hospira"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "Cleartext storage of sensitive information CWE-312"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2015-3952",
        "datePublished": "2019-03-25T15:42:39.000Z",
        "dateReserved": "2015-05-12T00:00:00.000Z",
        "dateUpdated": "2024-08-06T06:04:01.121Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }