VAR-202309-2171
Vulnerability from variot - Updated: 2025-01-09 23:11Rockwell Automation FactoryTalk View Machine Edition on the PanelView Plus, improperly verifies user’s input, which allows unauthenticated attacker to achieve remote code executed via crafted malicious packets. The device has the functionality, through a CIP class, to execute exported functions from libraries. There is a routine that restricts it to execute specific functions from two dynamic link library files. By using a CIP class, an attacker can upload a self-made library to the device which allows the attacker to bypass the security check and execute any code written in the function. Rockwell Automation PanelView Plus is a human-machine interface (HMI) product line from Rockwell Automation. These HMI devices are designed to integrate with industrial automation systems to provide operators with an intuitive interface to control and monitor production processes. PanelView Plus has a wide range of applications, especially in manufacturing, industrial control, and process control
Show details on source website{
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/VARIoTentry#",
"affected_products": {
"@id": "https://www.variotdbs.pl/ref/affected_products"
},
"configurations": {
"@id": "https://www.variotdbs.pl/ref/configurations"
},
"credits": {
"@id": "https://www.variotdbs.pl/ref/credits"
},
"cvss": {
"@id": "https://www.variotdbs.pl/ref/cvss/"
},
"description": {
"@id": "https://www.variotdbs.pl/ref/description/"
},
"exploit_availability": {
"@id": "https://www.variotdbs.pl/ref/exploit_availability/"
},
"external_ids": {
"@id": "https://www.variotdbs.pl/ref/external_ids/"
},
"iot": {
"@id": "https://www.variotdbs.pl/ref/iot/"
},
"iot_taxonomy": {
"@id": "https://www.variotdbs.pl/ref/iot_taxonomy/"
},
"patch": {
"@id": "https://www.variotdbs.pl/ref/patch/"
},
"problemtype_data": {
"@id": "https://www.variotdbs.pl/ref/problemtype_data/"
},
"references": {
"@id": "https://www.variotdbs.pl/ref/references/"
},
"sources": {
"@id": "https://www.variotdbs.pl/ref/sources/"
},
"sources_release_date": {
"@id": "https://www.variotdbs.pl/ref/sources_release_date/"
},
"sources_update_date": {
"@id": "https://www.variotdbs.pl/ref/sources_update_date/"
},
"threat_type": {
"@id": "https://www.variotdbs.pl/ref/threat_type/"
},
"title": {
"@id": "https://www.variotdbs.pl/ref/title/"
},
"type": {
"@id": "https://www.variotdbs.pl/ref/type/"
}
},
"@id": "https://www.variotdbs.pl/vuln/VAR-202309-2171",
"affected_products": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/affected_products#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
},
"@id": "https://www.variotdbs.pl/ref/sources"
}
},
"data": [
{
"model": "factorytalk view",
"scope": "lte",
"trust": 1.0,
"vendor": "rockwellautomation",
"version": "13.0"
},
{
"model": "automation factorytalk view",
"scope": "lte",
"trust": 0.6,
"vendor": "rockwell",
"version": "\u003c=13.0"
},
{
"model": "panelview plus",
"scope": null,
"trust": 0.6,
"vendor": "rockwellautomation",
"version": null
}
],
"sources": [
{
"db": "CNVD",
"id": "CNVD-2025-00304"
},
{
"db": "NVD",
"id": "CVE-2023-2071"
}
]
},
"cve": "CVE-2023-2071",
"cvss": {
"@context": {
"cvssV2": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/cvss/cvssV2#"
},
"@id": "https://www.variotdbs.pl/ref/cvss/cvssV2"
},
"cvssV3": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/cvss/cvssV3#"
},
"@id": "https://www.variotdbs.pl/ref/cvss/cvssV3/"
},
"severity": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/cvss/severity#"
},
"@id": "https://www.variotdbs.pl/ref/cvss/severity"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
},
"@id": "https://www.variotdbs.pl/ref/sources"
}
},
"data": [
{
"cvssV2": [
{
"accessComplexity": "LOW",
"accessVector": "NETWORK",
"authentication": "NONE",
"author": "CNVD",
"availabilityImpact": "COMPLETE",
"baseScore": 10.0,
"confidentialityImpact": "COMPLETE",
"exploitabilityScore": 10.0,
"id": "CNVD-2025-00304",
"impactScore": 10.0,
"integrityImpact": "COMPLETE",
"severity": "HIGH",
"trust": 0.6,
"vectorString": "AV:N/AC:L/Au:N/C:C/I:C/A:C",
"version": "2.0"
}
],
"cvssV3": [
{
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"author": "nvd@nist.gov",
"availabilityImpact": "HIGH",
"baseScore": 9.8,
"baseSeverity": "CRITICAL",
"confidentialityImpact": "HIGH",
"exploitabilityScore": 3.9,
"id": "CVE-2023-2071",
"impactScore": 5.9,
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"trust": 2.0,
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
}
],
"severity": [
{
"author": "nvd@nist.gov",
"id": "CVE-2023-2071",
"trust": 1.0,
"value": "CRITICAL"
},
{
"author": "PSIRT@rockwellautomation.com",
"id": "CVE-2023-2071",
"trust": 1.0,
"value": "CRITICAL"
},
{
"author": "CNVD",
"id": "CNVD-2025-00304",
"trust": 0.6,
"value": "HIGH"
}
]
}
],
"sources": [
{
"db": "CNVD",
"id": "CNVD-2025-00304"
},
{
"db": "NVD",
"id": "CVE-2023-2071"
},
{
"db": "NVD",
"id": "CVE-2023-2071"
}
]
},
"description": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/description#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "\nRockwell Automation FactoryTalk View Machine Edition on the PanelView Plus, improperly verifies user\u2019s input, which allows unauthenticated attacker to achieve remote code executed via crafted malicious packets. \u00a0The device has the functionality, through a CIP class, to execute exported functions from libraries. \u00a0There is a routine that restricts it to execute specific functions from two dynamic link library files. \u00a0By using a CIP class, an attacker can upload a self-made library to the device which allows the attacker to bypass the security check and execute any code written in the function. Rockwell Automation PanelView Plus is a human-machine interface (HMI) product line from Rockwell Automation. These HMI devices are designed to integrate with industrial automation systems to provide operators with an intuitive interface to control and monitor production processes. PanelView Plus has a wide range of applications, especially in manufacturing, industrial control, and process control",
"sources": [
{
"db": "NVD",
"id": "CVE-2023-2071"
},
{
"db": "CNVD",
"id": "CNVD-2025-00304"
}
],
"trust": 1.44
},
"external_ids": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/external_ids#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"db": "NVD",
"id": "CVE-2023-2071",
"trust": 1.6
},
{
"db": "CNVD",
"id": "CNVD-2025-00304",
"trust": 0.6
}
],
"sources": [
{
"db": "CNVD",
"id": "CNVD-2025-00304"
},
{
"db": "NVD",
"id": "CVE-2023-2071"
}
]
},
"id": "VAR-202309-2171",
"iot": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/iot#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": true,
"sources": [
{
"db": "CNVD",
"id": "CNVD-2025-00304"
}
],
"trust": 0.06
},
"iot_taxonomy": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/iot_taxonomy#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"category": [
"ICS"
],
"sub_category": null,
"trust": 0.6
}
],
"sources": [
{
"db": "CNVD",
"id": "CNVD-2025-00304"
}
]
},
"last_update_date": "2025-01-09T23:11:01.362000Z",
"patch": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/patch#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"title": "Patch for Rockwell Automation PanelView Plus Code Issue Vulnerability",
"trust": 0.6,
"url": "https://www.cnvd.org.cn/patchInfo/show/648781"
}
],
"sources": [
{
"db": "CNVD",
"id": "CNVD-2025-00304"
}
]
},
"problemtype_data": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/problemtype_data#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"problemtype": "CWE-20",
"trust": 1.0
},
{
"problemtype": "CWE-434",
"trust": 1.0
}
],
"sources": [
{
"db": "NVD",
"id": "CVE-2023-2071"
}
]
},
"references": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/references#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"trust": 1.0,
"url": "https://rockwellautomation.custhelp.com/app/answers/answer_view/a_id/1140724"
},
{
"trust": 0.6,
"url": "https://nvd.nist.gov/vuln/detail/cve-2023-2071"
}
],
"sources": [
{
"db": "CNVD",
"id": "CNVD-2025-00304"
},
{
"db": "NVD",
"id": "CVE-2023-2071"
}
]
},
"sources": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#",
"data": {
"@container": "@list"
}
},
"data": [
{
"db": "CNVD",
"id": "CNVD-2025-00304"
},
{
"db": "NVD",
"id": "CVE-2023-2071"
}
]
},
"sources_release_date": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources_release_date#",
"data": {
"@container": "@list"
}
},
"data": [
{
"date": "2025-01-03T00:00:00",
"db": "CNVD",
"id": "CNVD-2025-00304"
},
{
"date": "2023-09-12T14:15:09.663000",
"db": "NVD",
"id": "CVE-2023-2071"
}
]
},
"sources_update_date": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources_update_date#",
"data": {
"@container": "@list"
}
},
"data": [
{
"date": "2025-01-03T00:00:00",
"db": "CNVD",
"id": "CNVD-2025-00304"
},
{
"date": "2023-09-14T00:45:17.207000",
"db": "NVD",
"id": "CVE-2023-2071"
}
]
},
"title": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/title#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "Rockwell Automation PanelView Plus Code Issue Vulnerability",
"sources": [
{
"db": "CNVD",
"id": "CNVD-2025-00304"
}
],
"trust": 0.6
}
}
Drag nodes to arrange them. Double-click a related vulnerability to open it, or a source to visit its link. The Physics button on the left tunes the layout and the simulation; the legend at the bottom left hides or shows nodes by kind and by sighting type.
Sightings
| Author | Source | Type | Date | Other |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Vulnerability timeline
Sightings and recorded dates, grouped by day · UTC
How durations are calculated
Following the VTEM framework (Black Hat USA 2015), time to patch starts when the vendor or a coordinator was informed, total time to patch at discovery, and disclosure to patch at public disclosure; vendor response time runs from the report to its acknowledgement. Time to exploit and time unpatched start at the earliest discovery or report, falling back to public disclosure. Unpatched time ends at the first recorded patch, or the calculation time when no patch is recorded. Exploitation to patch shows the interval between those two events; a published proof of concept is a public exploit, not exploitation.
Each result shows its source dates. A negative interval means the end event came first; when date-only inputs leave the order open, the result says so. Date-only inputs are approximate. The earliest evidence wins: KEV report dates and sightings are labelled as such. Records describing another vulnerability are listed but never calculated from, and a PoC alone does not establish exploitation. Missing patch evidence does not prove that no patch exists. Filters below change the event list only.
Loading timeline…
The approach is described in our paper Mapping CVEs to MITRE ATT&CK Techniques: A Curated Gold-Set Classifier and the Limits of LLM-Assisted Label Expansion.
Browse all ATT&CK techniques and the vulnerabilities related to each.
Related by attack behaviour
Vulnerabilities whose description is nearest to this one in the vector space of the CIRCL/vulnerability-attack-technique-biencoder model. This is a similarity search over the bi-encoder space (plain cosine), not a classification, and it has no measured accuracy.