VAR-201812-0847

Vulnerability from variot - Updated: 2024-11-23 21:50

A SQL Injection vulnerability exists in EVLink Parking, v3.2.0-12_v1 and earlier, which could give access to the web interface with full privileges. Schneider Electric EVLink Parking is prone to multiple security vulnerabilities. An attacker can leverage these issues to compromise the application, access or modify data, exploit latent vulnerabilities in the underlying database, inject code, execute arbitrary code, or gain access to the affected system. EVLink Parking Versions 3.2.0-12_v1 and prior are vulnerable

Show details on source website

{
  "@context": {
    "@vocab": "https://www.variotdbs.pl/ref/VARIoTentry#",
    "affected_products": {
      "@id": "https://www.variotdbs.pl/ref/affected_products"
    },
    "configurations": {
      "@id": "https://www.variotdbs.pl/ref/configurations"
    },
    "credits": {
      "@id": "https://www.variotdbs.pl/ref/credits"
    },
    "cvss": {
      "@id": "https://www.variotdbs.pl/ref/cvss/"
    },
    "description": {
      "@id": "https://www.variotdbs.pl/ref/description/"
    },
    "exploit_availability": {
      "@id": "https://www.variotdbs.pl/ref/exploit_availability/"
    },
    "external_ids": {
      "@id": "https://www.variotdbs.pl/ref/external_ids/"
    },
    "iot": {
      "@id": "https://www.variotdbs.pl/ref/iot/"
    },
    "iot_taxonomy": {
      "@id": "https://www.variotdbs.pl/ref/iot_taxonomy/"
    },
    "patch": {
      "@id": "https://www.variotdbs.pl/ref/patch/"
    },
    "problemtype_data": {
      "@id": "https://www.variotdbs.pl/ref/problemtype_data/"
    },
    "references": {
      "@id": "https://www.variotdbs.pl/ref/references/"
    },
    "sources": {
      "@id": "https://www.variotdbs.pl/ref/sources/"
    },
    "sources_release_date": {
      "@id": "https://www.variotdbs.pl/ref/sources_release_date/"
    },
    "sources_update_date": {
      "@id": "https://www.variotdbs.pl/ref/sources_update_date/"
    },
    "threat_type": {
      "@id": "https://www.variotdbs.pl/ref/threat_type/"
    },
    "title": {
      "@id": "https://www.variotdbs.pl/ref/title/"
    },
    "type": {
      "@id": "https://www.variotdbs.pl/ref/type/"
    }
  },
  "@id": "https://www.variotdbs.pl/vuln/VAR-201812-0847",
  "affected_products": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/affected_products#",
      "data": {
        "@container": "@list"
      },
      "sources": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/sources#"
        },
        "@id": "https://www.variotdbs.pl/ref/sources"
      }
    },
    "data": [
      {
        "model": "evlink parking",
        "scope": "lte",
        "trust": 1.0,
        "vendor": "schneider electric",
        "version": "3.2.0-12"
      },
      {
        "model": "evlink parking",
        "scope": "lte",
        "trust": 0.8,
        "vendor": "schneider electric",
        "version": "3.2.0-12_v1"
      },
      {
        "model": "evlink parking",
        "scope": "eq",
        "trust": 0.6,
        "vendor": "schneider electric",
        "version": "3.2.0-12"
      },
      {
        "model": "evlink parking 3.2.0-12 v1",
        "scope": null,
        "trust": 0.3,
        "vendor": "schneider electric",
        "version": null
      }
    ],
    "sources": [
      {
        "db": "BID",
        "id": "106807"
      },
      {
        "db": "JVNDB",
        "id": "JVNDB-2018-013119"
      },
      {
        "db": "CNNVD",
        "id": "CNNVD-201812-1093"
      },
      {
        "db": "NVD",
        "id": "CVE-2018-7802"
      }
    ]
  },
  "configurations": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/configurations#",
      "children": {
        "@container": "@list"
      },
      "cpe_match": {
        "@container": "@list"
      },
      "data": {
        "@container": "@list"
      },
      "nodes": {
        "@container": "@list"
      }
    },
    "data": [
      {
        "CVE_data_version": "4.0",
        "nodes": [
          {
            "cpe_match": [
              {
                "cpe22Uri": "cpe:/o:schneider_electric:evlink_parking_firmware",
                "vulnerable": true
              }
            ],
            "operator": "OR"
          }
        ]
      }
    ],
    "sources": [
      {
        "db": "JVNDB",
        "id": "JVNDB-2018-013119"
      }
    ]
  },
  "credits": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/credits#",
      "sources": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/sources#"
        }
      }
    },
    "data": "Vladimir Kononovich and Vyacheslav Moskvin of Positive Technologies",
    "sources": [
      {
        "db": "BID",
        "id": "106807"
      },
      {
        "db": "CNNVD",
        "id": "CNNVD-201812-1093"
      }
    ],
    "trust": 0.9
  },
  "cve": "CVE-2018-7802",
  "cvss": {
    "@context": {
      "cvssV2": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/cvss/cvssV2#"
        },
        "@id": "https://www.variotdbs.pl/ref/cvss/cvssV2"
      },
      "cvssV3": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/cvss/cvssV3#"
        },
        "@id": "https://www.variotdbs.pl/ref/cvss/cvssV3/"
      },
      "severity": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/cvss/severity#"
        },
        "@id": "https://www.variotdbs.pl/ref/cvss/severity"
      },
      "sources": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/sources#"
        },
        "@id": "https://www.variotdbs.pl/ref/sources"
      }
    },
    "data": [
      {
        "cvssV2": [
          {
            "accessComplexity": "LOW",
            "accessVector": "NETWORK",
            "authentication": "SINGLE",
            "author": "nvd@nist.gov",
            "availabilityImpact": "PARTIAL",
            "baseScore": 6.5,
            "confidentialityImpact": "PARTIAL",
            "exploitabilityScore": 8.0,
            "id": "CVE-2018-7802",
            "impactScore": 6.4,
            "integrityImpact": "PARTIAL",
            "severity": "MEDIUM",
            "trust": 1.9,
            "vectorString": "AV:N/AC:L/Au:S/C:P/I:P/A:P",
            "version": "2.0"
          },
          {
            "accessComplexity": "LOW",
            "accessVector": "NETWORK",
            "authentication": "SINGLE",
            "author": "VULHUB",
            "availabilityImpact": "PARTIAL",
            "baseScore": 6.5,
            "confidentialityImpact": "PARTIAL",
            "exploitabilityScore": 8.0,
            "id": "VHN-137834",
            "impactScore": 6.4,
            "integrityImpact": "PARTIAL",
            "severity": "MEDIUM",
            "trust": 0.1,
            "vectorString": "AV:N/AC:L/AU:S/C:P/I:P/A:P",
            "version": "2.0"
          }
        ],
        "cvssV3": [
          {
            "attackComplexity": "LOW",
            "attackVector": "NETWORK",
            "author": "nvd@nist.gov",
            "availabilityImpact": "HIGH",
            "baseScore": 8.8,
            "baseSeverity": "HIGH",
            "confidentialityImpact": "HIGH",
            "exploitabilityScore": 2.8,
            "id": "CVE-2018-7802",
            "impactScore": 5.9,
            "integrityImpact": "HIGH",
            "privilegesRequired": "LOW",
            "scope": "UNCHANGED",
            "trust": 1.8,
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H",
            "version": "3.0"
          }
        ],
        "severity": [
          {
            "author": "nvd@nist.gov",
            "id": "CVE-2018-7802",
            "trust": 1.0,
            "value": "HIGH"
          },
          {
            "author": "NVD",
            "id": "CVE-2018-7802",
            "trust": 0.8,
            "value": "High"
          },
          {
            "author": "CNNVD",
            "id": "CNNVD-201812-1093",
            "trust": 0.6,
            "value": "HIGH"
          },
          {
            "author": "VULHUB",
            "id": "VHN-137834",
            "trust": 0.1,
            "value": "MEDIUM"
          },
          {
            "author": "VULMON",
            "id": "CVE-2018-7802",
            "trust": 0.1,
            "value": "MEDIUM"
          }
        ]
      }
    ],
    "sources": [
      {
        "db": "VULHUB",
        "id": "VHN-137834"
      },
      {
        "db": "VULMON",
        "id": "CVE-2018-7802"
      },
      {
        "db": "JVNDB",
        "id": "JVNDB-2018-013119"
      },
      {
        "db": "CNNVD",
        "id": "CNNVD-201812-1093"
      },
      {
        "db": "NVD",
        "id": "CVE-2018-7802"
      }
    ]
  },
  "description": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/description#",
      "sources": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/sources#"
        }
      }
    },
    "data": "A SQL Injection vulnerability exists in EVLink Parking, v3.2.0-12_v1 and earlier, which could give access to the web interface with full privileges. Schneider Electric EVLink Parking is prone to multiple security vulnerabilities. \nAn attacker can leverage these issues to compromise the application, access or modify data, exploit latent vulnerabilities in the underlying database, inject code, execute arbitrary code, or gain access to the affected system. \nEVLink Parking Versions 3.2.0-12_v1 and prior are vulnerable",
    "sources": [
      {
        "db": "NVD",
        "id": "CVE-2018-7802"
      },
      {
        "db": "JVNDB",
        "id": "JVNDB-2018-013119"
      },
      {
        "db": "BID",
        "id": "106807"
      },
      {
        "db": "VULHUB",
        "id": "VHN-137834"
      },
      {
        "db": "VULMON",
        "id": "CVE-2018-7802"
      }
    ],
    "trust": 2.07
  },
  "external_ids": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/external_ids#",
      "data": {
        "@container": "@list"
      },
      "sources": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/sources#"
        }
      }
    },
    "data": [
      {
        "db": "NVD",
        "id": "CVE-2018-7802",
        "trust": 2.9
      },
      {
        "db": "ICS CERT",
        "id": "ICSA-19-031-01",
        "trust": 2.9
      },
      {
        "db": "SCHNEIDER",
        "id": "SEVD-2018-354-01",
        "trust": 2.1
      },
      {
        "db": "BID",
        "id": "106807",
        "trust": 2.1
      },
      {
        "db": "JVNDB",
        "id": "JVNDB-2018-013119",
        "trust": 0.8
      },
      {
        "db": "CNNVD",
        "id": "CNNVD-201812-1093",
        "trust": 0.7
      },
      {
        "db": "VULHUB",
        "id": "VHN-137834",
        "trust": 0.1
      },
      {
        "db": "VULMON",
        "id": "CVE-2018-7802",
        "trust": 0.1
      }
    ],
    "sources": [
      {
        "db": "VULHUB",
        "id": "VHN-137834"
      },
      {
        "db": "VULMON",
        "id": "CVE-2018-7802"
      },
      {
        "db": "BID",
        "id": "106807"
      },
      {
        "db": "JVNDB",
        "id": "JVNDB-2018-013119"
      },
      {
        "db": "CNNVD",
        "id": "CNNVD-201812-1093"
      },
      {
        "db": "NVD",
        "id": "CVE-2018-7802"
      }
    ]
  },
  "id": "VAR-201812-0847",
  "iot": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/iot#",
      "sources": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/sources#"
        }
      }
    },
    "data": true,
    "sources": [
      {
        "db": "VULHUB",
        "id": "VHN-137834"
      }
    ],
    "trust": 0.85
  },
  "last_update_date": "2024-11-23T21:50:54.394000Z",
  "patch": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/patch#",
      "data": {
        "@container": "@list"
      },
      "sources": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/sources#"
        }
      }
    },
    "data": [
      {
        "title": "SEVD-2018-354-01",
        "trust": 0.8,
        "url": "https://download.schneider-electric.com/files?p_enDocType=Software+-+Release+Notes\u0026p_File_Name=SEVD-2018-354-01_Security+Notification.pdf\u0026p_Doc_Ref=SEVD-2018-354-01"
      },
      {
        "title": "Schneider Electric EVLink Parking SQL Repair measures for injecting vulnerabilities",
        "trust": 0.6,
        "url": "http://www.cnnvd.org.cn/web/xxk/bdxqById.tag?id=88166"
      },
      {
        "title": "Threatpost",
        "trust": 0.1,
        "url": "https://threatpost.com/critical-bug-patched-in-schneider-electric-vehicle-charging-station/140370/"
      }
    ],
    "sources": [
      {
        "db": "VULMON",
        "id": "CVE-2018-7802"
      },
      {
        "db": "JVNDB",
        "id": "JVNDB-2018-013119"
      },
      {
        "db": "CNNVD",
        "id": "CNNVD-201812-1093"
      }
    ]
  },
  "problemtype_data": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/problemtype_data#",
      "sources": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/sources#"
        }
      }
    },
    "data": [
      {
        "problemtype": "CWE-89",
        "trust": 1.9
      }
    ],
    "sources": [
      {
        "db": "VULHUB",
        "id": "VHN-137834"
      },
      {
        "db": "JVNDB",
        "id": "JVNDB-2018-013119"
      },
      {
        "db": "NVD",
        "id": "CVE-2018-7802"
      }
    ]
  },
  "references": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/references#",
      "data": {
        "@container": "@list"
      },
      "sources": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/sources#"
        }
      }
    },
    "data": [
      {
        "trust": 3.0,
        "url": "https://ics-cert.us-cert.gov/advisories/icsa-19-031-01"
      },
      {
        "trust": 2.5,
        "url": "http://www.securityfocus.com/bid/106807"
      },
      {
        "trust": 2.1,
        "url": "https://www.schneider-electric.com/en/download/document/sevd-2018-354-01/"
      },
      {
        "trust": 0.8,
        "url": "https://cve.mitre.org/cgi-bin/cvename.cgi?name=cve-2018-7802"
      },
      {
        "trust": 0.8,
        "url": "https://nvd.nist.gov/vuln/detail/cve-2018-7802"
      },
      {
        "trust": 0.3,
        "url": "https://www.schneider-electric.co.in"
      },
      {
        "trust": 0.3,
        "url": "https://download.schneider-electric.com/files?p_endoctype=software+-+release+notes\u0026p_file_name=sevd-2018-354-01_security+notification.pdf\u0026p_doc_ref=sevd-2018-354-01"
      },
      {
        "trust": 0.1,
        "url": "https://cwe.mitre.org/data/definitions/89.html"
      },
      {
        "trust": 0.1,
        "url": "https://nvd.nist.gov"
      },
      {
        "trust": 0.1,
        "url": "https://threatpost.com/critical-bug-patched-in-schneider-electric-vehicle-charging-station/140370/"
      }
    ],
    "sources": [
      {
        "db": "VULHUB",
        "id": "VHN-137834"
      },
      {
        "db": "VULMON",
        "id": "CVE-2018-7802"
      },
      {
        "db": "BID",
        "id": "106807"
      },
      {
        "db": "JVNDB",
        "id": "JVNDB-2018-013119"
      },
      {
        "db": "CNNVD",
        "id": "CNNVD-201812-1093"
      },
      {
        "db": "NVD",
        "id": "CVE-2018-7802"
      }
    ]
  },
  "sources": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/sources#",
      "data": {
        "@container": "@list"
      }
    },
    "data": [
      {
        "db": "VULHUB",
        "id": "VHN-137834"
      },
      {
        "db": "VULMON",
        "id": "CVE-2018-7802"
      },
      {
        "db": "BID",
        "id": "106807"
      },
      {
        "db": "JVNDB",
        "id": "JVNDB-2018-013119"
      },
      {
        "db": "CNNVD",
        "id": "CNNVD-201812-1093"
      },
      {
        "db": "NVD",
        "id": "CVE-2018-7802"
      }
    ]
  },
  "sources_release_date": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/sources_release_date#",
      "data": {
        "@container": "@list"
      }
    },
    "data": [
      {
        "date": "2018-12-24T00:00:00",
        "db": "VULHUB",
        "id": "VHN-137834"
      },
      {
        "date": "2018-12-24T00:00:00",
        "db": "VULMON",
        "id": "CVE-2018-7802"
      },
      {
        "date": "2019-01-31T00:00:00",
        "db": "BID",
        "id": "106807"
      },
      {
        "date": "2019-02-14T00:00:00",
        "db": "JVNDB",
        "id": "JVNDB-2018-013119"
      },
      {
        "date": "2018-12-25T00:00:00",
        "db": "CNNVD",
        "id": "CNNVD-201812-1093"
      },
      {
        "date": "2018-12-24T16:29:00.827000",
        "db": "NVD",
        "id": "CVE-2018-7802"
      }
    ]
  },
  "sources_update_date": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/sources_update_date#",
      "data": {
        "@container": "@list"
      }
    },
    "data": [
      {
        "date": "2019-02-28T00:00:00",
        "db": "VULHUB",
        "id": "VHN-137834"
      },
      {
        "date": "2019-02-28T00:00:00",
        "db": "VULMON",
        "id": "CVE-2018-7802"
      },
      {
        "date": "2019-01-31T00:00:00",
        "db": "BID",
        "id": "106807"
      },
      {
        "date": "2019-02-14T00:00:00",
        "db": "JVNDB",
        "id": "JVNDB-2018-013119"
      },
      {
        "date": "2019-02-26T00:00:00",
        "db": "CNNVD",
        "id": "CNNVD-201812-1093"
      },
      {
        "date": "2024-11-21T04:12:45.807000",
        "db": "NVD",
        "id": "CVE-2018-7802"
      }
    ]
  },
  "threat_type": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/threat_type#",
      "sources": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/sources#"
        }
      }
    },
    "data": "remote",
    "sources": [
      {
        "db": "CNNVD",
        "id": "CNNVD-201812-1093"
      }
    ],
    "trust": 0.6
  },
  "title": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/title#",
      "sources": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/sources#"
        }
      }
    },
    "data": "EVLink Parking In  SQL Injection vulnerability",
    "sources": [
      {
        "db": "JVNDB",
        "id": "JVNDB-2018-013119"
      }
    ],
    "trust": 0.8
  },
  "type": {
    "@context": {
      "@vocab": "https://www.variotdbs.pl/ref/type#",
      "sources": {
        "@container": "@list",
        "@context": {
          "@vocab": "https://www.variotdbs.pl/ref/sources#"
        }
      }
    },
    "data": "SQL injection",
    "sources": [
      {
        "db": "CNNVD",
        "id": "CNNVD-201812-1093"
      }
    ],
    "trust": 0.6
  }
}



Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Drag nodes to arrange them. Double-click a related vulnerability to open it, or a source to visit its link. The Physics button on the left tunes the layout and the simulation; the legend at the bottom left hides or shows nodes by kind and by sighting type.

Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.

Sightings

Author Source Type Date Other

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Vulnerability timeline

Sightings and recorded dates, grouped by day · UTC

JSON
How durations are calculated

Following the VTEM framework (Black Hat USA 2015), time to patch starts when the vendor or a coordinator was informed, total time to patch at discovery, and disclosure to patch at public disclosure; vendor response time runs from the report to its acknowledgement. Time to exploit and time unpatched start at the earliest discovery or report, falling back to public disclosure. Unpatched time ends at the first recorded patch, or the calculation time when no patch is recorded. Exploitation to patch shows the interval between those two events; a published proof of concept is a public exploit, not exploitation.

Each result shows its source dates. A negative interval means the end event came first; when date-only inputs leave the order open, the result says so. Date-only inputs are approximate. The earliest evidence wins: KEV report dates and sightings are labelled as such. Records describing another vulnerability are listed but never calculated from, and a PoC alone does not establish exploitation. Missing patch evidence does not prove that no patch exists. Filters below change the event list only.

Events included in the timeline

Loading timeline…


    Loading…

    Loading…

    Loading…

    Related by attack behaviour

    Vulnerabilities whose description is nearest to this one in the vector space of the CIRCL/vulnerability-attack-technique-biencoder model. This is a similarity search over the bi-encoder space (plain cosine), not a classification, and it has no measured accuracy.