FKIE_CVE-2026-17583

Vulnerability from fkie_nvd - Published: 2026-08-05 21:16 - Updated: 2026-08-26 14:50

{
  "affected": [
    {
      "affectedData": [
        {
          "defaultStatus": "unaffected",
          "product": "Applied Biosystems 3500/3500xL Series Data Collection Software",
          "vendor": "Thermo Fisher",
          "versions": [
            {
              "lessThanOrEqual": "4.0.2",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Applied Biosystems 3730/3730xL Series Data Collection Software",
          "vendor": "Thermo Fisher",
          "versions": [
            {
              "lessThanOrEqual": "5.0.2",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software",
          "vendor": "Thermo Fisher",
          "versions": [
            {
              "lessThanOrEqual": "1.2.5",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Applied Biosystems SeqStudio Flex Series Instrument Software",
          "vendor": "Thermo Fisher",
          "versions": [
            {
              "lessThanOrEqual": "1.2.0",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Applied Biosystems GeneMapper ID-X Software",
          "vendor": "Thermo Fisher",
          "versions": [
            {
              "lessThanOrEqual": "1.7.3",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Applied Biosystems 3130 Series Data Collection Software",
          "vendor": "Thermo Fisher",
          "versions": [
            {
              "lessThanOrEqual": "4.1",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "ABI PRISM 3100/3100-Avant Data Collection Software",
          "vendor": "Thermo Fisher",
          "versions": [
            {
              "lessThanOrEqual": "2.0",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "ABI PRISM 310 Data Collection Software",
          "vendor": "Thermo Fisher",
          "versions": [
            {
              "lessThanOrEqual": "3.1",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "source": "ics-cert@hq.dhs.gov"
    }
  ],
  "configurations": [
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:a:thermofisher:abi_prism_310_data_collection_software:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "052676E2-92C8-480C-B163-D3C4AEBA6ECF",
              "versionEndIncluding": "3.1",
              "vulnerable": true
            },
            {
              "criteria": "cpe:2.3:a:thermofisher:abi_prism_3100\\/3100-avant_data_collection_software:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "24B89BAB-EFA0-48D1-9E12-621A3AD95438",
              "versionEndIncluding": "2.0",
              "vulnerable": true
            },
            {
              "criteria": "cpe:2.3:a:thermofisher:applied_biosystems_3130_series_data_collection_software:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "2DA06A19-0C69-43AC-91C4-CBCCEB3E1525",
              "versionEndIncluding": "4.1",
              "vulnerable": true
            },
            {
              "criteria": "cpe:2.3:a:thermofisher:applied_biosystems_3500\\/3500xl_series_data_collection_software:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "A594FA04-80D2-44F5-B0E3-BB8A9600D68A",
              "versionEndExcluding": "4.0.3",
              "vulnerable": true
            },
            {
              "criteria": "cpe:2.3:a:thermofisher:applied_biosystems_3730\\/3730xl_series_data_collection_software:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "946E2337-8B1B-49F4-9597-A38C1CCC7A30",
              "versionEndExcluding": "5.0.3",
              "vulnerable": true
            },
            {
              "criteria": "cpe:2.3:a:thermofisher:applied_biosystems_seqstudio_flex_series_instrument_software:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "A4A8182F-41FD-4271-9120-844764414F85",
              "versionEndExcluding": "1.2.1",
              "vulnerable": true
            },
            {
              "criteria": "cpe:2.3:a:thermofisher:applied_biosystems_seqstudio_genetic_analyzer_data_collection_software:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "2C85423A-B2C6-4D68-8B0A-394B5E8DB3AA",
              "versionEndExcluding": "1.2.6",
              "vulnerable": true
            },
            {
              "criteria": "cpe:2.3:a:thermofisher:genemapper_id-x:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "E32BD188-9C4D-428D-8C64-154709AFE4AA",
              "versionEndExcluding": "1.7.4",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ]
    }
  ],
  "cveTags": [],
  "descriptions": [
    {
      "lang": "en",
      "value": "The affected\n\nThermo Fisher Applied Biosystems Genetic Analyzers are\u00a0vulnerable because .fsa/.hid output files can be edited. An attacker could tamper with these files, altering DNA data and resulting in inaccurate DNA test outcomes."
    }
  ],
  "id": "CVE-2026-17583",
  "lastModified": "2026-08-26T14:50:43.400",
  "metrics": {
    "cvssMetricV31": [
      {
        "cvssData": {
          "attackComplexity": "LOW",
          "attackVector": "LOCAL",
          "availabilityImpact": "HIGH",
          "baseScore": 8.4,
          "baseSeverity": "HIGH",
          "confidentialityImpact": "HIGH",
          "integrityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "scope": "UNCHANGED",
          "userInteraction": "NONE",
          "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
          "version": "3.1"
        },
        "exploitabilityScore": 2.5,
        "impactScore": 5.9,
        "source": "ics-cert@hq.dhs.gov",
        "type": "Secondary"
      }
    ],
    "cvssMetricV40": [
      {
        "cvssData": {
          "Automatable": "NOT_DEFINED",
          "Recovery": "NOT_DEFINED",
          "Safety": "NOT_DEFINED",
          "attackComplexity": "LOW",
          "attackRequirements": "NONE",
          "attackVector": "LOCAL",
          "availabilityRequirement": "NOT_DEFINED",
          "baseScore": 8.3,
          "baseSeverity": "HIGH",
          "confidentialityRequirement": "NOT_DEFINED",
          "exploitMaturity": "NOT_DEFINED",
          "integrityRequirement": "NOT_DEFINED",
          "modifiedAttackComplexity": "NOT_DEFINED",
          "modifiedAttackRequirements": "NOT_DEFINED",
          "modifiedAttackVector": "NOT_DEFINED",
          "modifiedPrivilegesRequired": "NOT_DEFINED",
          "modifiedSubAvailabilityImpact": "NOT_DEFINED",
          "modifiedSubConfidentialityImpact": "NOT_DEFINED",
          "modifiedSubIntegrityImpact": "NOT_DEFINED",
          "modifiedUserInteraction": "NOT_DEFINED",
          "modifiedVulnAvailabilityImpact": "NOT_DEFINED",
          "modifiedVulnConfidentialityImpact": "NOT_DEFINED",
          "modifiedVulnIntegrityImpact": "NOT_DEFINED",
          "privilegesRequired": "NONE",
          "providerUrgency": "NOT_DEFINED",
          "subAvailabilityImpact": "NONE",
          "subConfidentialityImpact": "NONE",
          "subIntegrityImpact": "HIGH",
          "userInteraction": "NONE",
          "valueDensity": "NOT_DEFINED",
          "vectorString": "CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:N/VI:H/VA:H/SC:N/SI:H/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X",
          "version": "4.0",
          "vulnAvailabilityImpact": "HIGH",
          "vulnConfidentialityImpact": "NONE",
          "vulnIntegrityImpact": "HIGH",
          "vulnerabilityResponseEffort": "NOT_DEFINED"
        },
        "source": "ics-cert@hq.dhs.gov",
        "type": "Secondary"
      }
    ],
    "ssvcV203": [
      {
        "source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
        "ssvcData": {
          "id": "CVE-2026-17583",
          "options": [
            {
              "exploitation": "none"
            },
            {
              "automatable": "no"
            },
            {
              "technicalImpact": "total"
            }
          ],
          "role": "CISA Coordinator",
          "timestamp": "2026-08-06T13:34:47.788856Z",
          "version": "2.0.3"
        }
      }
    ]
  },
  "published": "2026-08-05T21:16:57.393",
  "references": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "tags": [
        "Patch",
        "Vendor Advisory"
      ],
      "url": "https://documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/fsa_hid_bulletin.pdf"
    },
    {
      "source": "ics-cert@hq.dhs.gov",
      "tags": [
        "Patch",
        "Third Party Advisory",
        "US Government Resource"
      ],
      "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-216-01"
    },
    {
      "source": "ics-cert@hq.dhs.gov",
      "tags": [
        "Third Party Advisory"
      ],
      "url": "https://www.cve.org/CVERecord?id=CVE-2026-17583"
    }
  ],
  "sourceIdentifier": "ics-cert@hq.dhs.gov",
  "vulnStatus": "Analyzed",
  "weaknesses": [
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-353"
        }
      ],
      "source": "ics-cert@hq.dhs.gov",
      "type": "Secondary"
    }
  ]
}



Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.

Sightings

Author Source Type Date Other

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.

Loading…

Loading…

Loading…

Related by attack behaviour

Vulnerabilities whose description is nearest to this one in the vector space of the CIRCL/vulnerability-attack-technique-biencoder model. This is a similarity search over the bi-encoder space (plain cosine), not a classification, and it has no measured accuracy.


Loading…