CVE-2019-25720 (GCVE-0-2019-25720)

Vulnerability from cvelistv5 – Published: 2026-06-03 16:56 – Updated: 2026-06-03 17:44
VLAI
Title
Dräger SC Monitoring Devices DoS via Malformed Network Packet
Summary
Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.
SSVC
Exploitation: none Automatable: no Technical Impact: partial
CISA Coordinator (v2.0.3)
CWE
  • CWE-1286 - Improper Validation of Syntactic Correctness of Input
Assigner
References
Impacted products
Date Public
2019-11-21 00:00
Credits
Jeroen Slobbe and Max Grim
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2019-25720",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "partial"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2026-06-03T17:44:08.218138Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2026-06-03T17:44:47.456Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unknown",
          "product": "SC 6002XL",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "status": "affected",
              "version": "SC 6002XL",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unknown",
          "product": "SC6802XL",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "status": "affected",
              "version": "SC6802XL",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unknown",
          "product": "SC 7000",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "status": "affected",
              "version": "SC 7000",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unknown",
          "product": "SC8000",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "status": "affected",
              "version": "SC8000",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unknown",
          "product": "SC90000 XL",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "status": "affected",
              "version": "SC90000 XL",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Jeroen Slobbe and Max Grim"
        }
      ],
      "datePublic": "2019-11-21T00:00:00.000Z",
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eDr\u00e4ger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.\u003c/p\u003e"
            }
          ],
          "value": "Dr\u00e4ger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity."
        }
      ],
      "metrics": [
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "LOW",
            "attackRequirements": "NONE",
            "attackVector": "ADJACENT",
            "baseScore": 7.1,
            "baseSeverity": "HIGH",
            "exploitMaturity": "NOT_DEFINED",
            "privilegesRequired": "NONE",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "HIGH",
            "vulnConfidentialityImpact": "NONE",
            "vulnIntegrityImpact": "NONE",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        },
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "ADJACENT_NETWORK",
            "availabilityImpact": "HIGH",
            "baseScore": 6.5,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "NONE",
            "integrityImpact": "NONE",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-1286",
              "description": "CWE-1286 Improper Validation of Syntactic Correctness of Input",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2026-06-03T16:56:29.683Z",
        "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "shortName": "VulnCheck"
      },
      "references": [
        {
          "tags": [
            "vendor-advisory"
          ],
          "url": "https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf"
        },
        {
          "tags": [
            "third-party-advisory"
          ],
          "url": "https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-dos-via-malformed-network-packet"
        }
      ],
      "source": {
        "discovery": "EXTERNAL"
      },
      "title": "Dr\u00e4ger SC Monitoring Devices DoS via Malformed Network Packet"
    }
  },
  "cveMetadata": {
    "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
    "assignerShortName": "VulnCheck",
    "cveId": "CVE-2019-25720",
    "datePublished": "2026-06-03T16:56:29.683Z",
    "dateReserved": "2026-06-02T14:22:01.500Z",
    "dateUpdated": "2026-06-03T17:44:47.456Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.2",
  "vulnerability-lookup:meta": {
    "epss": {
      "cve": "CVE-2019-25720",
      "date": "2026-08-04",
      "epss": "0.00199",
      "percentile": "0.09972"
    },
    "nvd": "{\"cve\":{\"id\":\"CVE-2019-25720\",\"sourceIdentifier\":\"disclosure@vulncheck.com\",\"published\":\"2026-06-03T18:16:15.910\",\"lastModified\":\"2026-07-22T19:10:00.120\",\"vulnStatus\":\"Awaiting Analysis\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"Dr\u00e4ger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.\"},{\"lang\":\"es\",\"value\":\"Los dispositivos de monitorizaci\u00f3n Dr\u00e4ger SC (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contienen una vulnerabilidad de denegaci\u00f3n de servicio en todas las versiones de software que permite a atacantes no autenticados reiniciar el monitor enviando un paquete de red malformado. Los atacantes pueden enviar repetidamente dichos paquetes malformados para interrumpir la monitorizaci\u00f3n del paciente hasta que el dispositivo vuelve a la configuraci\u00f3n predeterminada y pierde la conectividad de red.\"}],\"affected\":[{\"source\":\"disclosure@vulncheck.com\",\"affectedData\":[{\"vendor\":\"Dr\u00e4ger\",\"product\":\"SC 6002XL\",\"defaultStatus\":\"unknown\",\"versions\":[{\"version\":\"SC 6002XL\",\"versionType\":\"custom\",\"status\":\"affected\"}]},{\"vendor\":\"Dr\u00e4ger\",\"product\":\"SC6802XL\",\"defaultStatus\":\"unknown\",\"versions\":[{\"version\":\"SC6802XL\",\"versionType\":\"custom\",\"status\":\"affected\"}]},{\"vendor\":\"Dr\u00e4ger\",\"product\":\"SC 7000\",\"defaultStatus\":\"unknown\",\"versions\":[{\"version\":\"SC 7000\",\"versionType\":\"custom\",\"status\":\"affected\"}]},{\"vendor\":\"Dr\u00e4ger\",\"product\":\"SC8000\",\"defaultStatus\":\"unknown\",\"versions\":[{\"version\":\"SC8000\",\"versionType\":\"custom\",\"status\":\"affected\"}]},{\"vendor\":\"Dr\u00e4ger\",\"product\":\"SC90000 XL\",\"defaultStatus\":\"unknown\",\"versions\":[{\"version\":\"SC90000 XL\",\"versionType\":\"custom\",\"status\":\"affected\"}]}]}],\"metrics\":{\"cvssMetricV40\":[{\"source\":\"disclosure@vulncheck.com\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"4.0\",\"vectorString\":\"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X\",\"baseScore\":7.1,\"baseSeverity\":\"HIGH\",\"attackVector\":\"ADJACENT\",\"attackComplexity\":\"LOW\",\"attackRequirements\":\"NONE\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"vulnConfidentialityImpact\":\"NONE\",\"vulnIntegrityImpact\":\"NONE\",\"vulnAvailabilityImpact\":\"HIGH\",\"subConfidentialityImpact\":\"NONE\",\"subIntegrityImpact\":\"NONE\",\"subAvailabilityImpact\":\"NONE\",\"exploitMaturity\":\"NOT_DEFINED\",\"confidentialityRequirement\":\"NOT_DEFINED\",\"integrityRequirement\":\"NOT_DEFINED\",\"availabilityRequirement\":\"NOT_DEFINED\",\"modifiedAttackVector\":\"NOT_DEFINED\",\"modifiedAttackComplexity\":\"NOT_DEFINED\",\"modifiedAttackRequirements\":\"NOT_DEFINED\",\"modifiedPrivilegesRequired\":\"NOT_DEFINED\",\"modifiedUserInteraction\":\"NOT_DEFINED\",\"modifiedVulnConfidentialityImpact\":\"NOT_DEFINED\",\"modifiedVulnIntegrityImpact\":\"NOT_DEFINED\",\"modifiedVulnAvailabilityImpact\":\"NOT_DEFINED\",\"modifiedSubConfidentialityImpact\":\"NOT_DEFINED\",\"modifiedSubIntegrityImpact\":\"NOT_DEFINED\",\"modifiedSubAvailabilityImpact\":\"NOT_DEFINED\",\"Safety\":\"NOT_DEFINED\",\"Automatable\":\"NOT_DEFINED\",\"Recovery\":\"NOT_DEFINED\",\"valueDensity\":\"NOT_DEFINED\",\"vulnerabilityResponseEffort\":\"NOT_DEFINED\",\"providerUrgency\":\"NOT_DEFINED\"}}],\"cvssMetricV31\":[{\"source\":\"disclosure@vulncheck.com\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H\",\"baseScore\":6.5,\"baseSeverity\":\"MEDIUM\",\"attackVector\":\"ADJACENT_NETWORK\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"NONE\",\"integrityImpact\":\"NONE\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":2.8,\"impactScore\":3.6}],\"ssvcV203\":[{\"source\":\"134c704f-9b21-4f2e-91b3-4a467353bcc0\",\"ssvcData\":{\"timestamp\":\"2026-06-03T17:44:08.218138Z\",\"id\":\"CVE-2019-25720\",\"options\":[{\"exploitation\":\"none\"},{\"automatable\":\"no\"},{\"technicalImpact\":\"partial\"}],\"role\":\"CISA Coordinator\",\"version\":\"2.0.3\"}}]},\"weaknesses\":[{\"source\":\"disclosure@vulncheck.com\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-1286\"}]}],\"references\":[{\"url\":\"https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf\",\"source\":\"disclosure@vulncheck.com\"},{\"url\":\"https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-dos-via-malformed-network-packet\",\"source\":\"disclosure@vulncheck.com\"}]}}",
    "vulnrichment": {
      "containers": "{\"adp\": [{\"title\": \"CISA ADP Vulnrichment\", \"metrics\": [{\"other\": {\"type\": \"ssvc\", \"content\": {\"id\": \"CVE-2019-25720\", \"role\": \"CISA Coordinator\", \"options\": [{\"Exploitation\": \"none\"}, {\"Automatable\": \"no\"}, {\"Technical Impact\": \"partial\"}], \"version\": \"2.0.3\", \"timestamp\": \"2026-06-03T17:44:08.218138Z\"}}}], \"providerMetadata\": {\"orgId\": \"134c704f-9b21-4f2e-91b3-4a467353bcc0\", \"shortName\": \"CISA-ADP\", \"dateUpdated\": \"2026-06-03T17:44:28.963Z\"}}], \"cna\": {\"title\": \"Dr\\u00e4ger SC Monitoring Devices DoS via Malformed Network Packet\", \"source\": {\"discovery\": \"EXTERNAL\"}, \"credits\": [{\"lang\": \"en\", \"type\": \"finder\", \"value\": \"Jeroen Slobbe and Max Grim\"}], \"metrics\": [{\"format\": \"CVSS\", \"cvssV4_0\": {\"Safety\": \"NOT_DEFINED\", \"version\": \"4.0\", \"Recovery\": \"NOT_DEFINED\", \"baseScore\": 7.1, \"Automatable\": \"NOT_DEFINED\", \"attackVector\": \"ADJACENT\", \"baseSeverity\": \"HIGH\", \"valueDensity\": \"NOT_DEFINED\", \"vectorString\": \"CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N\", \"exploitMaturity\": \"NOT_DEFINED\", \"providerUrgency\": \"NOT_DEFINED\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"attackRequirements\": \"NONE\", \"privilegesRequired\": \"NONE\", \"subIntegrityImpact\": \"NONE\", \"vulnIntegrityImpact\": \"NONE\", \"subAvailabilityImpact\": \"NONE\", \"vulnAvailabilityImpact\": \"HIGH\", \"subConfidentialityImpact\": \"NONE\", \"vulnConfidentialityImpact\": \"NONE\", \"vulnerabilityResponseEffort\": \"NOT_DEFINED\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}, {\"format\": \"CVSS\", \"cvssV3_1\": {\"scope\": \"UNCHANGED\", \"version\": \"3.1\", \"baseScore\": 6.5, \"attackVector\": \"ADJACENT_NETWORK\", \"baseSeverity\": \"MEDIUM\", \"vectorString\": \"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H\", \"integrityImpact\": \"NONE\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"availabilityImpact\": \"HIGH\", \"privilegesRequired\": \"NONE\", \"confidentialityImpact\": \"NONE\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}], \"affected\": [{\"vendor\": \"Dr\\u00e4ger\", \"product\": \"SC 6002XL\", \"versions\": [{\"status\": \"affected\", \"version\": \"SC 6002XL\", \"versionType\": \"custom\"}], \"defaultStatus\": \"unknown\"}, {\"vendor\": \"Dr\\u00e4ger\", \"product\": \"SC6802XL\", \"versions\": [{\"status\": \"affected\", \"version\": \"SC6802XL\", \"versionType\": \"custom\"}], \"defaultStatus\": \"unknown\"}, {\"vendor\": \"Dr\\u00e4ger\", \"product\": \"SC 7000\", \"versions\": [{\"status\": \"affected\", \"version\": \"SC 7000\", \"versionType\": \"custom\"}], \"defaultStatus\": \"unknown\"}, {\"vendor\": \"Dr\\u00e4ger\", \"product\": \"SC8000\", \"versions\": [{\"status\": \"affected\", \"version\": \"SC8000\", \"versionType\": \"custom\"}], \"defaultStatus\": \"unknown\"}, {\"vendor\": \"Dr\\u00e4ger\", \"product\": \"SC90000 XL\", \"versions\": [{\"status\": \"affected\", \"version\": \"SC90000 XL\", \"versionType\": \"custom\"}], \"defaultStatus\": \"unknown\"}], \"datePublic\": \"2019-11-21T00:00:00.000Z\", \"references\": [{\"url\": \"https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf\", \"tags\": [\"vendor-advisory\"]}, {\"url\": \"https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-dos-via-malformed-network-packet\", \"tags\": [\"third-party-advisory\"]}], \"descriptions\": [{\"lang\": \"en\", \"value\": \"Dr\\u00e4ger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"\u003cp\u003eDr\\u00e4ger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.\u003c/p\u003e\", \"base64\": false}]}], \"problemTypes\": [{\"descriptions\": [{\"lang\": \"en\", \"type\": \"CWE\", \"cweId\": \"CWE-1286\", \"description\": \"CWE-1286 Improper Validation of Syntactic Correctness of Input\"}]}], \"providerMetadata\": {\"orgId\": \"83251b91-4cc7-4094-a5c7-464a1b83ea10\", \"shortName\": \"VulnCheck\", \"dateUpdated\": \"2026-06-03T16:56:29.683Z\"}}}",
      "cveMetadata": "{\"cveId\": \"CVE-2019-25720\", \"state\": \"PUBLISHED\", \"dateUpdated\": \"2026-06-03T17:44:47.456Z\", \"dateReserved\": \"2026-06-02T14:22:01.500Z\", \"assignerOrgId\": \"83251b91-4cc7-4094-a5c7-464a1b83ea10\", \"datePublished\": \"2026-06-03T16:56:29.683Z\", \"assignerShortName\": \"VulnCheck\"}",
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }
  }
}



Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.

Sightings

Author Source Type Date Other

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.

Loading…

Detection rules are retrieved from Rulezet.

Loading…

Loading…