CVE-2019-25720 (GCVE-0-2019-25720)
Vulnerability from cvelistv5 – Published: 2026-06-03 16:56 – Updated: 2026-06-03 17:44- CWE-1286 - Improper Validation of Syntactic Correctness of Input
| URL | Tags |
|---|---|
| https://static.draeger.com/security/download/2019… | vendor-advisory |
| https://www.vulncheck.com/advisories/dr-ger-sc-mo… | third-party-advisory |
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2019-25720",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-06-03T17:44:08.218138Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T17:44:47.456Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unknown",
"product": "SC 6002XL",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC 6002XL",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "SC6802XL",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC6802XL",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "SC 7000",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC 7000",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "SC8000",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC8000",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "SC90000 XL",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC90000 XL",
"versionType": "custom"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "Jeroen Slobbe and Max Grim"
}
],
"datePublic": "2019-11-21T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eDr\u00e4ger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.\u003c/p\u003e"
}
],
"value": "Dr\u00e4ger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "ADJACENT",
"baseScore": 7.1,
"baseSeverity": "HIGH",
"exploitMaturity": "NOT_DEFINED",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "NONE",
"vulnIntegrityImpact": "NONE",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 6.5,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-1286",
"description": "CWE-1286 Improper Validation of Syntactic Correctness of Input",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T16:56:29.683Z",
"orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"shortName": "VulnCheck"
},
"references": [
{
"tags": [
"vendor-advisory"
],
"url": "https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf"
},
{
"tags": [
"third-party-advisory"
],
"url": "https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-dos-via-malformed-network-packet"
}
],
"source": {
"discovery": "EXTERNAL"
},
"title": "Dr\u00e4ger SC Monitoring Devices DoS via Malformed Network Packet"
}
},
"cveMetadata": {
"assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"assignerShortName": "VulnCheck",
"cveId": "CVE-2019-25720",
"datePublished": "2026-06-03T16:56:29.683Z",
"dateReserved": "2026-06-02T14:22:01.500Z",
"dateUpdated": "2026-06-03T17:44:47.456Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2",
"vulnerability-lookup:meta": {
"epss": {
"cve": "CVE-2019-25720",
"date": "2026-11-17",
"epss": "0.00199",
"percentile": "0.09849"
},
"nvd": {
"cve": {
"affected": [
{
"affectedData": [
{
"defaultStatus": "unknown",
"product": "SC 6002XL",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC 6002XL",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "SC6802XL",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC6802XL",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "SC 7000",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC 7000",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "SC8000",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC8000",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "SC90000 XL",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC90000 XL",
"versionType": "custom"
}
]
}
],
"source": "disclosure@vulncheck.com"
}
],
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "Dr\u00e4ger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity."
},
{
"lang": "es",
"value": "Los dispositivos de monitorizaci\u00f3n Dr\u00e4ger SC (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contienen una vulnerabilidad de denegaci\u00f3n de servicio en todas las versiones de software que permite a atacantes no autenticados reiniciar el monitor enviando un paquete de red malformado. Los atacantes pueden enviar repetidamente dichos paquetes malformados para interrumpir la monitorizaci\u00f3n del paciente hasta que el dispositivo vuelve a la configuraci\u00f3n predeterminada y pierde la conectividad de red."
}
],
"id": "CVE-2019-25720",
"lastModified": "2026-07-22T19:10:00.120",
"metrics": {
"cvssMetricV31": [
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 6.5,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
},
"exploitabilityScore": 2.8,
"impactScore": 3.6,
"source": "disclosure@vulncheck.com",
"type": "Secondary"
}
],
"cvssMetricV40": [
{
"cvssData": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "ADJACENT",
"availabilityRequirement": "NOT_DEFINED",
"baseScore": 7.1,
"baseSeverity": "HIGH",
"confidentialityRequirement": "NOT_DEFINED",
"exploitMaturity": "NOT_DEFINED",
"integrityRequirement": "NOT_DEFINED",
"modifiedAttackComplexity": "NOT_DEFINED",
"modifiedAttackRequirements": "NOT_DEFINED",
"modifiedAttackVector": "NOT_DEFINED",
"modifiedPrivilegesRequired": "NOT_DEFINED",
"modifiedSubAvailabilityImpact": "NOT_DEFINED",
"modifiedSubConfidentialityImpact": "NOT_DEFINED",
"modifiedSubIntegrityImpact": "NOT_DEFINED",
"modifiedUserInteraction": "NOT_DEFINED",
"modifiedVulnAvailabilityImpact": "NOT_DEFINED",
"modifiedVulnConfidentialityImpact": "NOT_DEFINED",
"modifiedVulnIntegrityImpact": "NOT_DEFINED",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "NONE",
"vulnIntegrityImpact": "NONE",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"source": "disclosure@vulncheck.com",
"type": "Secondary"
}
],
"ssvcV203": [
{
"source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"ssvcData": {
"id": "CVE-2019-25720",
"options": [
{
"exploitation": "none"
},
{
"automatable": "no"
},
{
"technicalImpact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-06-03T17:44:08.218138Z",
"version": "2.0.3"
}
}
]
},
"published": "2026-06-03T18:16:15.910",
"references": [
{
"source": "disclosure@vulncheck.com",
"url": "https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf"
},
{
"source": "disclosure@vulncheck.com",
"url": "https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-dos-via-malformed-network-packet"
}
],
"sourceIdentifier": "disclosure@vulncheck.com",
"vulnStatus": "Awaiting Analysis",
"weaknesses": [
{
"description": [
{
"lang": "en",
"value": "CWE-1286"
}
],
"source": "disclosure@vulncheck.com",
"type": "Secondary"
}
]
}
},
"vulnrichment": {
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2019-25720",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-06-03T17:44:08.218138Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T17:44:28.963Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unknown",
"product": "SC 6002XL",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC 6002XL",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "SC6802XL",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC6802XL",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "SC 7000",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC 7000",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "SC8000",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC8000",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "SC90000 XL",
"vendor": "Dr\u00e4ger",
"versions": [
{
"status": "affected",
"version": "SC90000 XL",
"versionType": "custom"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "Jeroen Slobbe and Max Grim"
}
],
"datePublic": "2019-11-21T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eDr\u00e4ger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.\u003c/p\u003e"
}
],
"value": "Dr\u00e4ger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain a denial-of-service vulnerability in all software versions that allows unauthenticated attackers to reboot the monitor by sending a malformed network packet. Attackers can repeatedly send such malformed packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "ADJACENT",
"baseScore": 7.1,
"baseSeverity": "HIGH",
"exploitMaturity": "NOT_DEFINED",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "NONE",
"vulnIntegrityImpact": "NONE",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 6.5,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-1286",
"description": "CWE-1286 Improper Validation of Syntactic Correctness of Input",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T16:56:29.683Z",
"orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"shortName": "VulnCheck"
},
"references": [
{
"tags": [
"vendor-advisory"
],
"url": "https://static.draeger.com/security/download/2019-11-27-Draeger-SC7000-SC9000-security-advisory-update-v1-5.pdf"
},
{
"tags": [
"third-party-advisory"
],
"url": "https://www.vulncheck.com/advisories/dr-ger-sc-monitoring-devices-dos-via-malformed-network-packet"
}
],
"source": {
"discovery": "EXTERNAL"
},
"title": "Dr\u00e4ger SC Monitoring Devices DoS via Malformed Network Packet"
}
},
"cveMetadata": {
"assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"assignerShortName": "VulnCheck",
"cveId": "CVE-2019-25720",
"datePublished": "2026-06-03T16:56:29.683Z",
"dateReserved": "2026-06-02T14:22:01.500Z",
"dateUpdated": "2026-06-03T17:44:47.456Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
}
}
Drag nodes to arrange them. Double-click a related vulnerability to open it, or a source to visit its link. The Physics button on the left tunes the layout and the simulation; the legend at the bottom left hides or shows nodes by kind and by sighting type.
Sightings
| Author | Source | Type | Date | Other |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Vulnerability timeline
Sightings and recorded dates, grouped by day · UTC
How durations are calculated
Following the VTEM framework (Black Hat USA 2015), time to patch starts when the vendor or a coordinator was informed, total time to patch at discovery, and disclosure to patch at public disclosure; vendor response time runs from the report to its acknowledgement. Time to exploit and time unpatched start at the earliest discovery or report, falling back to public disclosure. Unpatched time ends at the first recorded patch, or the calculation time when no patch is recorded. Exploitation to patch shows the interval between those two events; a published proof of concept is a public exploit, not exploitation.
Each result shows its source dates. A negative interval means the end event came first; when date-only inputs leave the order open, the result says so. Date-only inputs are approximate. The earliest evidence wins: KEV report dates and sightings are labelled as such. Records describing another vulnerability are listed but never calculated from, and a PoC alone does not establish exploitation. Missing patch evidence does not prove that no patch exists. Filters below change the event list only.
Loading timeline…
The approach is described in our paper Mapping CVEs to MITRE ATT&CK Techniques: A Curated Gold-Set Classifier and the Limits of LLM-Assisted Label Expansion.
Browse all ATT&CK techniques and the vulnerabilities related to each.
Related by attack behaviour
Vulnerabilities whose description is nearest to this one in the vector space of the CIRCL/vulnerability-attack-technique-biencoder model. This is a similarity search over the bi-encoder space (plain cosine), not a classification, and it has no measured accuracy.