CVE-2019-13531 (GCVE-0-2019-13531)

Vulnerability from cvelistv5 – Published: 2019-11-08 19:46 – Updated: 2025-05-22 18:37
VLAI
Title
Medtronic Valleylab FT10 and LS10 Improper Authentication
Summary
In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.
CWE
  • CWE-287 - Improper Authentication
Assigner
icscert CNA under the icscert root
CNA scorecard B 75/100 over 337 records in the last 180 days details
Impacted products
Vendor Product Version CPE status
Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) Affected: 0 , ≤ 2.1.0 (custom)
Affected: 0 , ≤ 2.0.3 (custom)
guessed Create a notification for this product.
Medtronic Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) Affected: 0 , ≤ 1.20.2 (custom)
guessed Create a notification for this product.
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2024-08-04T23:57:39.240Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "tags": [
              "x_refsource_MISC",
              "x_transferred"
            ],
            "url": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01"
          }
        ],
        "title": "CVE Program Container"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "Valleylab FT10 Energy Platform (VLFT10GEN)",
          "vendor": "Medtronic",
          "versions": [
            {
              "lessThanOrEqual": "2.1.0",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            },
            {
              "lessThanOrEqual": "2.0.3",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unaffected",
          "product": "Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States)",
          "vendor": "Medtronic",
          "versions": [
            {
              "lessThanOrEqual": "1.20.2",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Medtronic reported these vulnerabilities to CISA."
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eIn Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.\u003c/p\u003e"
            }
          ],
          "value": "In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator."
        }
      ],
      "metrics": [
        {
          "cvssV3_1": {
            "attackComplexity": "HIGH",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "LOW",
            "baseScore": 4.8,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "NONE",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-287",
              "description": "CWE-287 Improper Authentication",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2025-05-22T18:37:04.526Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/valleylab-generator-rfid-vulnerabilities.html"
        },
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-01"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eA software patch is available now for the affected Valleylab platforms. If you suspect you are in possession of an instrument that is not FDA approved or cleared to be used with Medtronic Valleylab FT10 or LS10, please contact Medtronic or your medical device supplier. If you have concerns about FDA clearance or approval of current or future instruments, please contact your medical device supplier. Please contact \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/covidien/en-us/support/software.html\"\u003eMedtronic\u003c/a\u003e\u0026nbsp;to obtain the software patch.\u003c/p\u003e\u003cp\u003eMedtronic has released additional patient focused information at the following location:\u003c/p\u003e\u003cp\u003e\u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e\n\n\u003cbr\u003e"
            }
          ],
          "value": "A software patch is available now for the affected Valleylab platforms. If you suspect you are in possession of an instrument that is not FDA approved or cleared to be used with Medtronic Valleylab FT10 or LS10, please contact Medtronic or your medical device supplier. If you have concerns about FDA clearance or approval of current or future instruments, please contact your medical device supplier. Please contact  https://www.medtronic.com/security"
        }
      ],
      "source": {
        "advisory": "ICSMA-19-311-01",
        "discovery": "INTERNAL"
      },
      "title": "Medtronic Valleylab FT10 and LS10 Improper Authentication",
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      },
      "x_legacyV4Record": {
        "CVE_data_meta": {
          "ASSIGNER": "ics-cert@hq.dhs.gov",
          "ID": "CVE-2019-13531",
          "STATE": "PUBLIC"
        },
        "affects": {
          "vendor": {
            "vendor_data": [
              {
                "product": {
                  "product_data": [
                    {
                      "product_name": "Valleylab FT10 Energy Platform (VLFT10GEN)",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "version 2.1.0 and lower"
                          },
                          {
                            "version_value": "version 2.0.3 and lower"
                          }
                        ]
                      }
                    },
                    {
                      "product_name": "Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States)",
                      "version": {
                        "version_data": [
                          {
                            "version_value": "version 1.20.2 and lower"
                          }
                        ]
                      }
                    }
                  ]
                },
                "vendor_name": "Medtronic"
              }
            ]
          }
        },
        "data_format": "MITRE",
        "data_type": "CVE",
        "data_version": "4.0",
        "description": {
          "description_data": [
            {
              "lang": "eng",
              "value": "In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator."
            }
          ]
        },
        "problemtype": {
          "problemtype_data": [
            {
              "description": [
                {
                  "lang": "eng",
                  "value": "IMPROPER AUTHENTICATION CWE-287"
                }
              ]
            }
          ]
        },
        "references": {
          "reference_data": [
            {
              "name": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01",
              "refsource": "MISC",
              "url": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01"
            }
          ]
        }
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2019-13531",
    "datePublished": "2019-11-08T19:46:45.000Z",
    "dateReserved": "2019-07-11T00:00:00.000Z",
    "dateUpdated": "2025-05-22T18:37:04.526Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1",
  "vulnerability-lookup:meta": {
    "epss": {
      "cve": "CVE-2019-13531",
      "date": "2026-11-17",
      "epss": "0.00388",
      "percentile": "0.33438"
    },
    "nvd": {
      "cve": {
        "configurations": [
          {
            "nodes": [
              {
                "cpeMatch": [
                  {
                    "criteria": "cpe:2.3:o:medtronic:valleylab_ft10_energy_platform_firmware:2.0.3:*:*:*:*:*:*:*",
                    "matchCriteriaId": "89DAFB01-536A-4966-8EAC-5D6CED32CC63",
                    "vulnerable": true
                  },
                  {
                    "criteria": "cpe:2.3:o:medtronic:valleylab_ft10_energy_platform_firmware:2.1.0:*:*:*:*:*:*:*",
                    "matchCriteriaId": "31DF6D35-968D-4E8D-A8F5-578B4B9C2BE9",
                    "vulnerable": true
                  }
                ],
                "negate": false,
                "operator": "OR"
              },
              {
                "cpeMatch": [
                  {
                    "criteria": "cpe:2.3:h:medtronic:valleylab_ft10_energy_platform:-:*:*:*:*:*:*:*",
                    "matchCriteriaId": "164230CB-E2BF-447F-8537-C9401FA0CC09",
                    "vulnerable": false
                  }
                ],
                "negate": false,
                "operator": "OR"
              }
            ],
            "operator": "AND"
          },
          {
            "nodes": [
              {
                "cpeMatch": [
                  {
                    "criteria": "cpe:2.3:o:medtronic:valleylab_ls10_energy_platform_firmware:*:*:*:*:*:*:*:*",
                    "matchCriteriaId": "EEAE74E9-49AC-4B77-8831-11EA09B204CB",
                    "versionEndIncluding": "1.20.2",
                    "vulnerable": true
                  }
                ],
                "negate": false,
                "operator": "OR"
              },
              {
                "cpeMatch": [
                  {
                    "criteria": "cpe:2.3:h:medtronic:valleylab_ls10_energy_platform:-:*:*:*:*:*:*:*",
                    "matchCriteriaId": "82294F1E-0BD0-4F55-A732-09B68AA37624",
                    "vulnerable": false
                  }
                ],
                "negate": false,
                "operator": "OR"
              }
            ],
            "operator": "AND"
          }
        ],
        "cveTags": [],
        "descriptions": [
          {
            "lang": "en",
            "value": "In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN\u2014not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator."
          },
          {
            "lang": "es",
            "value": "En Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) versi\u00f3n 2.1.0 y versiones anteriores y versi\u00f3n 2.0.3 y versiones atneriores, y Valleylab LS10 Energy Platform (VLLS10GEN \u2014no disponible en los Estados Unidos) versi\u00f3n 1.20.2 y versiones anteriores, se utiliz\u00f3 el mecanismo de seguridad RFID para la autenticaci\u00f3n entre la plataforma de energ\u00eda FT10 / LS10 y los instrumentos se pueden omitir, lo que permite que los instrumentos no aut\u00e9nticos se conecten al generador."
          }
        ],
        "id": "CVE-2019-13531",
        "lastModified": "2025-05-22T19:15:22.673",
        "metrics": {
          "cvssMetricV2": [
            {
              "acInsufInfo": false,
              "baseSeverity": "LOW",
              "cvssData": {
                "accessComplexity": "LOW",
                "accessVector": "LOCAL",
                "authentication": "NONE",
                "availabilityImpact": "NONE",
                "baseScore": 2.1,
                "confidentialityImpact": "NONE",
                "integrityImpact": "PARTIAL",
                "vectorString": "AV:L/AC:L/Au:N/C:N/I:P/A:N",
                "version": "2.0"
              },
              "exploitabilityScore": 3.9,
              "impactScore": 2.9,
              "obtainAllPrivilege": false,
              "obtainOtherPrivilege": false,
              "obtainUserPrivilege": false,
              "source": "nvd@nist.gov",
              "type": "Primary",
              "userInteractionRequired": false
            }
          ],
          "cvssMetricV31": [
            {
              "cvssData": {
                "attackComplexity": "HIGH",
                "attackVector": "PHYSICAL",
                "availabilityImpact": "LOW",
                "baseScore": 4.8,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:L",
                "version": "3.1"
              },
              "exploitabilityScore": 0.5,
              "impactScore": 4.2,
              "source": "ics-cert@hq.dhs.gov",
              "type": "Secondary"
            },
            {
              "cvssData": {
                "attackComplexity": "LOW",
                "attackVector": "PHYSICAL",
                "availabilityImpact": "NONE",
                "baseScore": 4.6,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N",
                "version": "3.1"
              },
              "exploitabilityScore": 0.9,
              "impactScore": 3.6,
              "source": "nvd@nist.gov",
              "type": "Primary"
            }
          ]
        },
        "published": "2019-11-08T20:15:10.570",
        "references": [
          {
            "source": "ics-cert@hq.dhs.gov",
            "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/valleylab-generator-rfid-vulnerabilities.html"
          },
          {
            "source": "ics-cert@hq.dhs.gov",
            "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-01"
          },
          {
            "source": "af854a3a-2127-422b-91ae-364da2661108",
            "tags": [
              "Third Party Advisory",
              "US Government Resource"
            ],
            "url": "https://www.us-cert.gov/ics/advisories/icsma-19-311-01"
          }
        ],
        "sourceIdentifier": "ics-cert@hq.dhs.gov",
        "vulnStatus": "Modified",
        "weaknesses": [
          {
            "description": [
              {
                "lang": "en",
                "value": "CWE-287"
              }
            ],
            "source": "ics-cert@hq.dhs.gov",
            "type": "Secondary"
          },
          {
            "description": [
              {
                "lang": "en",
                "value": "NVD-CWE-noinfo"
              }
            ],
            "source": "nvd@nist.gov",
            "type": "Primary"
          }
        ]
      }
    }
  }
}


Solutions

CNA · icscert
A software patch is available now for the affected Valleylab platforms. If you suspect you are in possession of an instrument that is not FDA approved or cleared to be used with Medtronic Valleylab FT10 or LS10, please contact Medtronic or your medical device supplier. If you have concerns about FDA clearance or approval of current or future instruments, please contact your medical device supplier. Please contact https://www.medtronic.com/security

Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Drag nodes to arrange them. Double-click a related vulnerability to open it, or a source to visit its link. The Physics button on the left tunes the layout and the simulation; the legend at the bottom left hides or shows nodes by kind and by sighting type.

Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.

Sightings

Author Source Type Date Other

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Vulnerability timeline

Sightings and recorded dates, grouped by day · UTC

JSON
How durations are calculated

Following the VTEM framework (Black Hat USA 2015), time to patch starts when the vendor or a coordinator was informed, total time to patch at discovery, and disclosure to patch at public disclosure; vendor response time runs from the report to its acknowledgement. Time to exploit and time unpatched start at the earliest discovery or report, falling back to public disclosure. Unpatched time ends at the first recorded patch, or the calculation time when no patch is recorded. Exploitation to patch shows the interval between those two events; a published proof of concept is a public exploit, not exploitation.

Each result shows its source dates. A negative interval means the end event came first; when date-only inputs leave the order open, the result says so. Date-only inputs are approximate. The earliest evidence wins: KEV report dates and sightings are labelled as such. Records describing another vulnerability are listed but never calculated from, and a PoC alone does not establish exploitation. Missing patch evidence does not prove that no patch exists. Filters below change the event list only.

Events included in the timeline

Loading timeline…


    Loading…

    Loading…

    Loading…

    Related by attack behaviour

    Vulnerabilities whose description is nearest to this one in the vector space of the CIRCL/vulnerability-attack-technique-biencoder model. This is a similarity search over the bi-encoder space (plain cosine), not a classification, and it has no measured accuracy.