Search

Find a vulnerability

Search criteria

    2 vulnerabilities by Eppendorf

    CVE-2026-7251 (GCVE-0-2026-7251)

    Vulnerability from nvd – Published: 2026-05-26 17:06 – Updated: 2026-06-04 21:10
    VLAI
    Title
    Eppendorf BioFlo 320 Use of hard-coded password
    Summary
    Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
    SSVC
    Exploitation: none Automatable: yes Technical Impact: total
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-259 - Use of hard-coded password
    Assigner
    Impacted products
    Vendor Product Version
    Eppendorf BioFlo 320 Affected: All (custom)
    Create a notification for this product.
    Date Public
    2026-05-26 17:00
    Credits
    BIO-ISAC reported this vulnerability to CISA.
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2026-7251",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "yes"
                      },
                      {
                        "Technical Impact": "total"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-05-26T18:32:05.373317Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-05-26T18:39:20.583Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "BioFlo 320",
              "vendor": "Eppendorf",
              "versions": [
                {
                  "status": "affected",
                  "version": "All",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "BIO-ISAC reported this vulnerability to CISA."
            }
          ],
          "datePublic": "2026-05-26T17:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "Eppendorf BioFlo 320\u0026nbsp;is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted."
                }
              ],
              "value": "Eppendorf BioFlo 320\u00a0is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "NETWORK",
                "baseScore": 9.3,
                "baseSeverity": "CRITICAL",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "HIGH",
                "vulnConfidentialityImpact": "HIGH",
                "vulnIntegrityImpact": "HIGH",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "NETWORK",
                "availabilityImpact": "HIGH",
                "baseScore": 9.8,
                "baseSeverity": "CRITICAL",
                "confidentialityImpact": "HIGH",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-259",
                  "description": "CWE-259 Use of hard-coded password",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-04T21:10:51.687Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "url": "https://www.eppendorf.com/software-downloads"
            },
            {
              "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-146-01"
            },
            {
              "url": "https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-146-01.json"
            }
          ],
          "solutions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eEppendorf has released a software update that permanently removes VNC access from the controller. Users should download and apply this update from: https://www.eppendorf.com/software-downloads.\u003cbr\u003e\u003ca href=\"https://www.eppendorf.com/software-downloads\"\u003ehttps://www.eppendorf.com/software-downloads\u003c/a\u003e\u003c/p\u003e\u003cp\u003eAll affected BioFlo 320 systems always shipped with Virtual Network Computing (VNC) disabled by default, and VNC can only be enabled locally at the tower. Eppendorf has removed VNC configuration information from all current documentation, so it no longer appears in BioFlo 320 Operating Manuals.\u003c/p\u003e\u003cp\u003eEppendorf recommends user do the following:\u003cbr\u003e- Verify that VNC is disabled on the controller\u003cbr\u003e- Enable security so that only Admin and Supervisor roles can change VNC settings\u003cbr\u003e- Install Version 5.0 Software as soon as possible\u003c/p\u003e"
                }
              ],
              "value": "Eppendorf has released a software update that permanently removes VNC access from the controller. Users should download and apply this update from: https://www.eppendorf.com/software-downloads.\n https://www.eppendorf.com/software-downloads \n\n\n\nAll affected BioFlo 320 systems always shipped with Virtual Network Computing (VNC) disabled by default, and VNC can only be enabled locally at the tower. Eppendorf has removed VNC configuration information from all current documentation, so it no longer appears in BioFlo 320 Operating Manuals.\n\n\n\nEppendorf recommends user do the following:\n- Verify that VNC is disabled on the controller\n- Enable security so that only Admin and Supervisor roles can change VNC settings\n- Install Version 5.0 Software as soon as possible"
            }
          ],
          "source": {
            "advisory": "ICSMA-26-146-01",
            "discovery": "EXTERNAL"
          },
          "title": "Eppendorf BioFlo 320 Use of hard-coded password",
          "x_generator": {
            "engine": "Vulnogram 1.0.2"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2026-7251",
        "datePublished": "2026-05-26T17:06:46.563Z",
        "dateReserved": "2026-04-27T18:37:39.380Z",
        "dateUpdated": "2026-06-04T21:10:51.687Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2026-7251 (GCVE-0-2026-7251)

    Vulnerability from cvelistv5 – Published: 2026-05-26 17:06 – Updated: 2026-06-04 21:10
    VLAI
    Title
    Eppendorf BioFlo 320 Use of hard-coded password
    Summary
    Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
    SSVC
    Exploitation: none Automatable: yes Technical Impact: total
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-259 - Use of hard-coded password
    Assigner
    Impacted products
    Vendor Product Version
    Eppendorf BioFlo 320 Affected: All (custom)
    Create a notification for this product.
    Date Public
    2026-05-26 17:00
    Credits
    BIO-ISAC reported this vulnerability to CISA.
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2026-7251",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "yes"
                      },
                      {
                        "Technical Impact": "total"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-05-26T18:32:05.373317Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-05-26T18:39:20.583Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "BioFlo 320",
              "vendor": "Eppendorf",
              "versions": [
                {
                  "status": "affected",
                  "version": "All",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "BIO-ISAC reported this vulnerability to CISA."
            }
          ],
          "datePublic": "2026-05-26T17:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "Eppendorf BioFlo 320\u0026nbsp;is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted."
                }
              ],
              "value": "Eppendorf BioFlo 320\u00a0is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "NETWORK",
                "baseScore": 9.3,
                "baseSeverity": "CRITICAL",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "HIGH",
                "vulnConfidentialityImpact": "HIGH",
                "vulnIntegrityImpact": "HIGH",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "NETWORK",
                "availabilityImpact": "HIGH",
                "baseScore": 9.8,
                "baseSeverity": "CRITICAL",
                "confidentialityImpact": "HIGH",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-259",
                  "description": "CWE-259 Use of hard-coded password",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-04T21:10:51.687Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "url": "https://www.eppendorf.com/software-downloads"
            },
            {
              "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-146-01"
            },
            {
              "url": "https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-146-01.json"
            }
          ],
          "solutions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eEppendorf has released a software update that permanently removes VNC access from the controller. Users should download and apply this update from: https://www.eppendorf.com/software-downloads.\u003cbr\u003e\u003ca href=\"https://www.eppendorf.com/software-downloads\"\u003ehttps://www.eppendorf.com/software-downloads\u003c/a\u003e\u003c/p\u003e\u003cp\u003eAll affected BioFlo 320 systems always shipped with Virtual Network Computing (VNC) disabled by default, and VNC can only be enabled locally at the tower. Eppendorf has removed VNC configuration information from all current documentation, so it no longer appears in BioFlo 320 Operating Manuals.\u003c/p\u003e\u003cp\u003eEppendorf recommends user do the following:\u003cbr\u003e- Verify that VNC is disabled on the controller\u003cbr\u003e- Enable security so that only Admin and Supervisor roles can change VNC settings\u003cbr\u003e- Install Version 5.0 Software as soon as possible\u003c/p\u003e"
                }
              ],
              "value": "Eppendorf has released a software update that permanently removes VNC access from the controller. Users should download and apply this update from: https://www.eppendorf.com/software-downloads.\n https://www.eppendorf.com/software-downloads \n\n\n\nAll affected BioFlo 320 systems always shipped with Virtual Network Computing (VNC) disabled by default, and VNC can only be enabled locally at the tower. Eppendorf has removed VNC configuration information from all current documentation, so it no longer appears in BioFlo 320 Operating Manuals.\n\n\n\nEppendorf recommends user do the following:\n- Verify that VNC is disabled on the controller\n- Enable security so that only Admin and Supervisor roles can change VNC settings\n- Install Version 5.0 Software as soon as possible"
            }
          ],
          "source": {
            "advisory": "ICSMA-26-146-01",
            "discovery": "EXTERNAL"
          },
          "title": "Eppendorf BioFlo 320 Use of hard-coded password",
          "x_generator": {
            "engine": "Vulnogram 1.0.2"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2026-7251",
        "datePublished": "2026-05-26T17:06:46.563Z",
        "dateReserved": "2026-04-27T18:37:39.380Z",
        "dateUpdated": "2026-06-04T21:10:51.687Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }