    CVE-2018-10626 (GCVE-0-2018-10626)

    Vulnerability from nvd – Published: 2018-08-10 18:00 – Updated: 2026-05-19 14:32
    VLAI
    Title
    Medtronic MyCareLink 24950 Patient Monitor Insufficient Verification of Data Authenticity
    Summary
    Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network.
    CWE
    • CWE-345 - Insufficient Verification of Data Authenticity
    Assigner
    Impacted products
    Date Public
    2018-08-07 06:00
    Credits
    Billy Rios, Jesse Young, and Jonathan Butts of Whitescope LLC reported these vulnerabilities to CISA.
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-05T07:46:45.928Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "name": "105042",
                "tags": [
                  "vdb-entry",
                  "x_refsource_BID",
                  "x_transferred"
                ],
                "url": "http://www.securityfocus.com/bid/105042"
              },
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "24950 MyCareLink Monitor",
              "vendor": "Medtronic",
              "versions": [
                {
                  "status": "affected",
                  "version": "All versions"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "24952 MyCareLink Monitor",
              "vendor": "Medtronic",
              "versions": [
                {
                  "status": "affected",
                  "version": "All versions"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Billy Rios, Jesse Young, and Jonathan Butts of Whitescope LLC reported these vulnerabilities to CISA."
            }
          ],
          "datePublic": "2018-08-07T06:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003e\u003cspan\u003eMedtronic MyCareLink Patient Monitor\u2019s update service\u0026nbsp;\u003c/span\u003edoes not sufficiently verify the authenticity of the data uploaded. An \nattacker who obtains per-product credentials from the monitor and paired\n implantable cardiac device information can potentially upload invalid \ndata to the Medtronic CareLink network.\n\n\u003c/p\u003e"
                }
              ],
              "value": "Medtronic MyCareLink Patient Monitor\u2019s update service\u00a0does not sufficiently verify the authenticity of the data uploaded. An \nattacker who obtains per-product credentials from the monitor and paired\n implantable cardiac device information can potentially upload invalid \ndata to the Medtronic CareLink network."
            }
          ],
          "metrics": [
            {
              "cvssV3_1": {
                "attackComplexity": "HIGH",
                "attackVector": "ADJACENT_NETWORK",
                "availabilityImpact": "NONE",
                "baseScore": 4.4,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "LOW",
                "integrityImpact": "LOW",
                "privilegesRequired": "LOW",
                "scope": "CHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:N",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-345",
                  "description": "CWE-345 Insufficient Verification of Data Authenticity",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-05-19T14:32:21.570Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "name": "105042",
              "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-8-7-18.html"
            },
            {
              "tags": [
                "vdb-entry"
              ],
              "url": "http://www.securityfocus.com/bid/105042"
            },
            {
              "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-219-01"
            },
            {
              "url": "https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2018/icsma-18-219-01.json"
            }
          ],
          "source": {
            "advisory": "ICSMA-18-219-01",
            "discovery": "EXTERNAL"
          },
          "title": "Medtronic MyCareLink 24950 Patient Monitor Insufficient Verification of Data Authenticity",
          "workarounds": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "Medtronic has made server-side updates to address this insufficient \nverification vulnerability. Medtronic is \nimplementing additional server-side mitigations to enhance data \nintegrity and authenticity."
                }
              ],
              "value": "Medtronic has made server-side updates to address this insufficient \nverification vulnerability. Medtronic is \nimplementing additional server-side mitigations to enhance data \nintegrity and authenticity."
            },
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eMedtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, users should:\u003c/p\u003e\n\u003cul\u003e\u003cli\u003eMaintain good physical control over the home monitor.\u003c/li\u003e\u003cli\u003eOnly use home monitors obtained directly from their healthcare \nprovider or a Medtronic representative to ensure integrity of the \nsystem.\u003c/li\u003e\u003c/ul\u003e"
                }
              ],
              "value": "Medtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, users should:\n\n\n  *  Maintain good physical control over the home monitor.\n  *  Only use home monitors obtained directly from their healthcare \nprovider or a Medtronic representative to ensure integrity of the \nsystem."
            },
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eMedtronic has released additional patient focused information, at the following location:\u003c/p\u003e\n\u003cp\u003e\u003ca href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e"
                }
              ],
              "value": "Medtronic has released additional patient focused information, at the following location:\n\n\n\n\n https://www.medtronic.com/security"
            },
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "Users should follow CISA\u0027s guidance in the following areas:\u003cbr\u003e\n* \u003ca href=\"https://www.cisa.gov/news-events/news/securing-internet-things-iot\"\u003eSecuring the Internet of Things\u003c/a\u003e\u003cbr\u003e\n* \u003ca href=\"https://www.cisa.gov/news-events/news/home-network-security\"\u003eHome Network Security\u003c/a\u003e"
                }
              ],
              "value": "Users should follow CISA\u0027s guidance in the following areas:\n\n*  Securing the Internet of Things https://www.cisa.gov/news-events/news/securing-internet-things-iot \n\n*  Home Network Security https://www.cisa.gov/news-events/news/home-network-security"
            }
          ],
          "x_generator": {
            "engine": "Vulnogram 0.2.0"
          },
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "DATE_PUBLIC": "2018-08-07T00:00:00",
              "ID": "CVE-2018-10626",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Medtronic MyCareLink 24950, 24952 Patient Monitor",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "All versions"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "ICS-CERT"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "A vulnerability was discovered in all versions of Medtronic MyCareLink 24950 and 24952 Patient Monitor. The affected product\u0027s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "INSUFFICIENT VERIFICATION OF DATA AUTHENTICITY CWE-345"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "105042",
                  "refsource": "BID",
                  "url": "http://www.securityfocus.com/bid/105042"
                },
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2018-10626",
        "datePublished": "2018-08-10T18:00:00.000Z",
        "dateReserved": "2018-05-01T00:00:00.000Z",
        "dateUpdated": "2026-05-19T14:32:21.570Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2018-10622 (GCVE-0-2018-10622)

    Vulnerability from nvd – Published: 2018-08-10 18:00 – Updated: 2026-06-22 20:42
    VLAI
    Title
    Medtronic MyCareLink Patient Monitor Network Credential Weakness
    Summary
    Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication.
    CWE
    Assigner
    Impacted products
    Vendor Product Version
    Medtronic 24950 MyCareLink Monitor Affected: 0 , < February 25, 2026 (custom)
    Create a notification for this product.
    Medtronic 24952 MyCareLink Monitor Affected: 0 , < February 25, 2026 (custom)
    Create a notification for this product.
    Date Public
    2026-05-07 06:00
    Credits
    Billy Rios, Jesse Young, and Jonathan Butts of Whitescope LLC reported these vulnerabilities to CISA.
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-05T07:46:47.036Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "name": "105042",
                "tags": [
                  "vdb-entry",
                  "x_refsource_BID",
                  "x_transferred"
                ],
                "url": "http://www.securityfocus.com/bid/105042"
              },
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "24950 MyCareLink Monitor",
              "vendor": "Medtronic",
              "versions": [
                {
                  "lessThan": "February 25, 2026",
                  "status": "affected",
                  "version": "0",
                  "versionType": "custom"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "24952 MyCareLink Monitor",
              "vendor": "Medtronic",
              "versions": [
                {
                  "lessThan": "February 25, 2026",
                  "status": "affected",
                  "version": "0",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Billy Rios, Jesse Young, and Jonathan Butts of Whitescope LLC reported these vulnerabilities to CISA."
            }
          ],
          "datePublic": "2026-05-07T06:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cdiv\u003e\u003cdiv\u003eMedtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication.\u003c/div\u003e\u003c/div\u003e"
                }
              ],
              "value": "Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication."
            }
          ],
          "metrics": [
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "PHYSICAL",
                "availabilityImpact": "NONE",
                "baseScore": 5.2,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "LOW",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:N",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-313",
                  "description": "CWE-313",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-22T20:42:16.327Z",
            "orgId": "3c337f33-b2bd-4529-9792-f48cdb2838b4",
            "shortName": "Medtronic"
          },
          "references": [
            {
              "name": "105042",
              "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-8-7-18.html"
            },
            {
              "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-219-01"
            },
            {
              "tags": [
                "vdb-entry"
              ],
              "url": "http://www.securityfocus.com/bid/105042"
            },
            {
              "url": "https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2018/icsma-18-219-01.json"
            }
          ],
          "source": {
            "advisory": "ICSMA-18-219-01",
            "discovery": "EXTERNAL"
          },
          "title": "Medtronic MyCareLink Patient Monitor Network Credential Weakness",
          "workarounds": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eMedtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, users should:\u003c/p\u003e\n\u003cul\u003e\u003cli\u003eMaintain good physical control over the home monitor.\u003c/li\u003e\u003cli\u003eOnly use home monitors obtained directly from their healthcare \nprovider or a Medtronic representative to ensure integrity of the \nsystem.\u003c/li\u003e\u003c/ul\u003e"
                }
              ],
              "value": "Medtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, users should:\n\n\n  *  Maintain good physical control over the home monitor.\n  *  Only use home monitors obtained directly from their healthcare \nprovider or a Medtronic representative to ensure integrity of the \nsystem."
            },
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eMedtronic has released additional patient focused information, at the following location:\u003c/p\u003e\n\u003cp\u003e\u003ca href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e"
                }
              ],
              "value": "Medtronic has released additional patient focused information, at the following location:\n\n\n\n\n https://www.medtronic.com/security"
            },
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "Users should follow CISA\u0027s guidance in the following areas:\u003cbr\u003e\n* \u003ca href=\"https://www.cisa.gov/news-events/news/securing-internet-things-iot\"\u003eSecuring the Internet of Things\u003c/a\u003e\u003cbr\u003e\n* \u003ca href=\"https://www.cisa.gov/news-events/news/home-network-security\"\u003eHome Network Security\u003c/a\u003e"
                }
              ],
              "value": "Users should follow CISA\u0027s guidance in the following areas:\n\n*  Securing the Internet of Things https://www.cisa.gov/news-events/news/securing-internet-things-iot \n\n*  Home Network Security https://www.cisa.gov/news-events/news/home-network-security"
            }
          ],
          "x_generator": {
            "engine": "Vulnogram 0.2.0"
          },
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "DATE_PUBLIC": "2018-08-07T00:00:00",
              "ID": "CVE-2018-10626",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Medtronic MyCareLink 24950, 24952 Patient Monitor",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "All versions"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "ICS-CERT"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "A vulnerability was discovered in all versions of Medtronic MyCareLink 24950 and 24952 Patient Monitor. The affected product\u0027s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "INSUFFICIENT VERIFICATION OF DATA AUTHENTICITY CWE-345"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "105042",
                  "refsource": "BID",
                  "url": "http://www.securityfocus.com/bid/105042"
                },
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2018-10622",
        "datePublished": "2018-08-10T18:00:00.000Z",
        "dateReserved": "2018-05-01T00:00:00.000Z",
        "dateUpdated": "2026-06-22T20:42:16.327Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2018-10626 (GCVE-0-2018-10626)

    Vulnerability from cvelistv5 – Published: 2018-08-10 18:00 – Updated: 2026-05-19 14:32
    VLAI
    Title
    Medtronic MyCareLink 24950 Patient Monitor Insufficient Verification of Data Authenticity
    Summary
    Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network.
    CWE
    • CWE-345 - Insufficient Verification of Data Authenticity
    Assigner
    Impacted products
    Date Public
    2018-08-07 06:00
    Credits
    Billy Rios, Jesse Young, and Jonathan Butts of Whitescope LLC reported these vulnerabilities to CISA.
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-05T07:46:45.928Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "name": "105042",
                "tags": [
                  "vdb-entry",
                  "x_refsource_BID",
                  "x_transferred"
                ],
                "url": "http://www.securityfocus.com/bid/105042"
              },
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "24950 MyCareLink Monitor",
              "vendor": "Medtronic",
              "versions": [
                {
                  "status": "affected",
                  "version": "All versions"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "24952 MyCareLink Monitor",
              "vendor": "Medtronic",
              "versions": [
                {
                  "status": "affected",
                  "version": "All versions"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Billy Rios, Jesse Young, and Jonathan Butts of Whitescope LLC reported these vulnerabilities to CISA."
            }
          ],
          "datePublic": "2018-08-07T06:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003e\u003cspan\u003eMedtronic MyCareLink Patient Monitor\u2019s update service\u0026nbsp;\u003c/span\u003edoes not sufficiently verify the authenticity of the data uploaded. An \nattacker who obtains per-product credentials from the monitor and paired\n implantable cardiac device information can potentially upload invalid \ndata to the Medtronic CareLink network.\n\n\u003c/p\u003e"
                }
              ],
              "value": "Medtronic MyCareLink Patient Monitor\u2019s update service\u00a0does not sufficiently verify the authenticity of the data uploaded. An \nattacker who obtains per-product credentials from the monitor and paired\n implantable cardiac device information can potentially upload invalid \ndata to the Medtronic CareLink network."
            }
          ],
          "metrics": [
            {
              "cvssV3_1": {
                "attackComplexity": "HIGH",
                "attackVector": "ADJACENT_NETWORK",
                "availabilityImpact": "NONE",
                "baseScore": 4.4,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "LOW",
                "integrityImpact": "LOW",
                "privilegesRequired": "LOW",
                "scope": "CHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:N",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-345",
                  "description": "CWE-345 Insufficient Verification of Data Authenticity",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-05-19T14:32:21.570Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "name": "105042",
              "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-8-7-18.html"
            },
            {
              "tags": [
                "vdb-entry"
              ],
              "url": "http://www.securityfocus.com/bid/105042"
            },
            {
              "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-219-01"
            },
            {
              "url": "https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2018/icsma-18-219-01.json"
            }
          ],
          "source": {
            "advisory": "ICSMA-18-219-01",
            "discovery": "EXTERNAL"
          },
          "title": "Medtronic MyCareLink 24950 Patient Monitor Insufficient Verification of Data Authenticity",
          "workarounds": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "Medtronic has made server-side updates to address this insufficient \nverification vulnerability. Medtronic is \nimplementing additional server-side mitigations to enhance data \nintegrity and authenticity."
                }
              ],
              "value": "Medtronic has made server-side updates to address this insufficient \nverification vulnerability. Medtronic is \nimplementing additional server-side mitigations to enhance data \nintegrity and authenticity."
            },
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eMedtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, users should:\u003c/p\u003e\n\u003cul\u003e\u003cli\u003eMaintain good physical control over the home monitor.\u003c/li\u003e\u003cli\u003eOnly use home monitors obtained directly from their healthcare \nprovider or a Medtronic representative to ensure integrity of the \nsystem.\u003c/li\u003e\u003c/ul\u003e"
                }
              ],
              "value": "Medtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, users should:\n\n\n  *  Maintain good physical control over the home monitor.\n  *  Only use home monitors obtained directly from their healthcare \nprovider or a Medtronic representative to ensure integrity of the \nsystem."
            },
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eMedtronic has released additional patient focused information, at the following location:\u003c/p\u003e\n\u003cp\u003e\u003ca href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e"
                }
              ],
              "value": "Medtronic has released additional patient focused information, at the following location:\n\n\n\n\n https://www.medtronic.com/security"
            },
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "Users should follow CISA\u0027s guidance in the following areas:\u003cbr\u003e\n* \u003ca href=\"https://www.cisa.gov/news-events/news/securing-internet-things-iot\"\u003eSecuring the Internet of Things\u003c/a\u003e\u003cbr\u003e\n* \u003ca href=\"https://www.cisa.gov/news-events/news/home-network-security\"\u003eHome Network Security\u003c/a\u003e"
                }
              ],
              "value": "Users should follow CISA\u0027s guidance in the following areas:\n\n*  Securing the Internet of Things https://www.cisa.gov/news-events/news/securing-internet-things-iot \n\n*  Home Network Security https://www.cisa.gov/news-events/news/home-network-security"
            }
          ],
          "x_generator": {
            "engine": "Vulnogram 0.2.0"
          },
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "DATE_PUBLIC": "2018-08-07T00:00:00",
              "ID": "CVE-2018-10626",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Medtronic MyCareLink 24950, 24952 Patient Monitor",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "All versions"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "ICS-CERT"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "A vulnerability was discovered in all versions of Medtronic MyCareLink 24950 and 24952 Patient Monitor. The affected product\u0027s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "INSUFFICIENT VERIFICATION OF DATA AUTHENTICITY CWE-345"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "105042",
                  "refsource": "BID",
                  "url": "http://www.securityfocus.com/bid/105042"
                },
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2018-10626",
        "datePublished": "2018-08-10T18:00:00.000Z",
        "dateReserved": "2018-05-01T00:00:00.000Z",
        "dateUpdated": "2026-05-19T14:32:21.570Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2018-10622 (GCVE-0-2018-10622)

    Vulnerability from cvelistv5 – Published: 2018-08-10 18:00 – Updated: 2026-06-22 20:42
    VLAI
    Title
    Medtronic MyCareLink Patient Monitor Network Credential Weakness
    Summary
    Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication.
    CWE
    Assigner
    Impacted products
    Vendor Product Version
    Medtronic 24950 MyCareLink Monitor Affected: 0 , < February 25, 2026 (custom)
    Create a notification for this product.
    Medtronic 24952 MyCareLink Monitor Affected: 0 , < February 25, 2026 (custom)
    Create a notification for this product.
    Date Public
    2026-05-07 06:00
    Credits
    Billy Rios, Jesse Young, and Jonathan Butts of Whitescope LLC reported these vulnerabilities to CISA.
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-05T07:46:47.036Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "name": "105042",
                "tags": [
                  "vdb-entry",
                  "x_refsource_BID",
                  "x_transferred"
                ],
                "url": "http://www.securityfocus.com/bid/105042"
              },
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "24950 MyCareLink Monitor",
              "vendor": "Medtronic",
              "versions": [
                {
                  "lessThan": "February 25, 2026",
                  "status": "affected",
                  "version": "0",
                  "versionType": "custom"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "24952 MyCareLink Monitor",
              "vendor": "Medtronic",
              "versions": [
                {
                  "lessThan": "February 25, 2026",
                  "status": "affected",
                  "version": "0",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Billy Rios, Jesse Young, and Jonathan Butts of Whitescope LLC reported these vulnerabilities to CISA."
            }
          ],
          "datePublic": "2026-05-07T06:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cdiv\u003e\u003cdiv\u003eMedtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication.\u003c/div\u003e\u003c/div\u003e"
                }
              ],
              "value": "Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication."
            }
          ],
          "metrics": [
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "PHYSICAL",
                "availabilityImpact": "NONE",
                "baseScore": 5.2,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "LOW",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:N",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-313",
                  "description": "CWE-313",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-22T20:42:16.327Z",
            "orgId": "3c337f33-b2bd-4529-9792-f48cdb2838b4",
            "shortName": "Medtronic"
          },
          "references": [
            {
              "name": "105042",
              "url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-8-7-18.html"
            },
            {
              "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-219-01"
            },
            {
              "tags": [
                "vdb-entry"
              ],
              "url": "http://www.securityfocus.com/bid/105042"
            },
            {
              "url": "https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2018/icsma-18-219-01.json"
            }
          ],
          "source": {
            "advisory": "ICSMA-18-219-01",
            "discovery": "EXTERNAL"
          },
          "title": "Medtronic MyCareLink Patient Monitor Network Credential Weakness",
          "workarounds": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eMedtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, users should:\u003c/p\u003e\n\u003cul\u003e\u003cli\u003eMaintain good physical control over the home monitor.\u003c/li\u003e\u003cli\u003eOnly use home monitors obtained directly from their healthcare \nprovider or a Medtronic representative to ensure integrity of the \nsystem.\u003c/li\u003e\u003c/ul\u003e"
                }
              ],
              "value": "Medtronic recommends users take additional defensive measures to minimize the risk of exploitation. Specifically, users should:\n\n\n  *  Maintain good physical control over the home monitor.\n  *  Only use home monitors obtained directly from their healthcare \nprovider or a Medtronic representative to ensure integrity of the \nsystem."
            },
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eMedtronic has released additional patient focused information, at the following location:\u003c/p\u003e\n\u003cp\u003e\u003ca href=\"https://www.medtronic.com/security\"\u003ehttps://www.medtronic.com/security\u003c/a\u003e\u003c/p\u003e"
                }
              ],
              "value": "Medtronic has released additional patient focused information, at the following location:\n\n\n\n\n https://www.medtronic.com/security"
            },
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "Users should follow CISA\u0027s guidance in the following areas:\u003cbr\u003e\n* \u003ca href=\"https://www.cisa.gov/news-events/news/securing-internet-things-iot\"\u003eSecuring the Internet of Things\u003c/a\u003e\u003cbr\u003e\n* \u003ca href=\"https://www.cisa.gov/news-events/news/home-network-security\"\u003eHome Network Security\u003c/a\u003e"
                }
              ],
              "value": "Users should follow CISA\u0027s guidance in the following areas:\n\n*  Securing the Internet of Things https://www.cisa.gov/news-events/news/securing-internet-things-iot \n\n*  Home Network Security https://www.cisa.gov/news-events/news/home-network-security"
            }
          ],
          "x_generator": {
            "engine": "Vulnogram 0.2.0"
          },
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "DATE_PUBLIC": "2018-08-07T00:00:00",
              "ID": "CVE-2018-10626",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Medtronic MyCareLink 24950, 24952 Patient Monitor",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "All versions"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "ICS-CERT"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "A vulnerability was discovered in all versions of Medtronic MyCareLink 24950 and 24952 Patient Monitor. The affected product\u0027s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "INSUFFICIENT VERIFICATION OF DATA AUTHENTICITY CWE-345"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "105042",
                  "refsource": "BID",
                  "url": "http://www.securityfocus.com/bid/105042"
                },
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2018-10622",
        "datePublished": "2018-08-10T18:00:00.000Z",
        "dateReserved": "2018-05-01T00:00:00.000Z",
        "dateUpdated": "2026-06-22T20:42:16.327Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }