VAR-201807-0333
Vulnerability from variot - Updated: 2025-05-23 23:07Medtronic 2090 CareLink Programmer
uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications. Medtronic 2090 CareLink Programmer Contains an information disclosure vulnerability.Information is obtained, information is altered, and service operation is disrupted (DoS) There is a possibility of being put into a state. Medtronic2090CareLinkProgrammer is a set of portable computer products from Medtronic Corporation of the United States. This product is used to manage and program cardiac devices in the medical industry
Show details on source website{
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/VARIoTentry#",
"affected_products": {
"@id": "https://www.variotdbs.pl/ref/affected_products"
},
"configurations": {
"@id": "https://www.variotdbs.pl/ref/configurations"
},
"credits": {
"@id": "https://www.variotdbs.pl/ref/credits"
},
"cvss": {
"@id": "https://www.variotdbs.pl/ref/cvss/"
},
"description": {
"@id": "https://www.variotdbs.pl/ref/description/"
},
"exploit_availability": {
"@id": "https://www.variotdbs.pl/ref/exploit_availability/"
},
"external_ids": {
"@id": "https://www.variotdbs.pl/ref/external_ids/"
},
"iot": {
"@id": "https://www.variotdbs.pl/ref/iot/"
},
"iot_taxonomy": {
"@id": "https://www.variotdbs.pl/ref/iot_taxonomy/"
},
"patch": {
"@id": "https://www.variotdbs.pl/ref/patch/"
},
"problemtype_data": {
"@id": "https://www.variotdbs.pl/ref/problemtype_data/"
},
"references": {
"@id": "https://www.variotdbs.pl/ref/references/"
},
"sources": {
"@id": "https://www.variotdbs.pl/ref/sources/"
},
"sources_release_date": {
"@id": "https://www.variotdbs.pl/ref/sources_release_date/"
},
"sources_update_date": {
"@id": "https://www.variotdbs.pl/ref/sources_update_date/"
},
"threat_type": {
"@id": "https://www.variotdbs.pl/ref/threat_type/"
},
"title": {
"@id": "https://www.variotdbs.pl/ref/title/"
},
"type": {
"@id": "https://www.variotdbs.pl/ref/type/"
}
},
"@id": "https://www.variotdbs.pl/vuln/VAR-201807-0333",
"affected_products": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/affected_products#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
},
"@id": "https://www.variotdbs.pl/ref/sources"
}
},
"data": [
{
"model": "2090 carelink programmer",
"scope": null,
"trust": 1.4,
"vendor": "medtronic",
"version": null
},
{
"model": "2090 carelink programmer",
"scope": "eq",
"trust": 1.0,
"vendor": "medtronic",
"version": "*"
},
{
"model": "carelink programmer",
"scope": "eq",
"trust": 0.6,
"vendor": "medtronic",
"version": "2090"
}
],
"sources": [
{
"db": "CNVD",
"id": "CNVD-2018-12557"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-008035"
},
{
"db": "CNNVD",
"id": "CNNVD-201807-212"
},
{
"db": "NVD",
"id": "CVE-2018-10596"
}
]
},
"configurations": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/configurations#",
"children": {
"@container": "@list"
},
"cpe_match": {
"@container": "@list"
},
"data": {
"@container": "@list"
},
"nodes": {
"@container": "@list"
}
},
"data": [
{
"CVE_data_version": "4.0",
"nodes": [
{
"cpe_match": [
{
"cpe22Uri": "cpe:/o:medtronic:2090_carelink_programmer_firmware",
"vulnerable": true
}
],
"operator": "OR"
}
]
}
],
"sources": [
{
"db": "JVNDB",
"id": "JVNDB-2018-008035"
}
]
},
"cve": "CVE-2018-10596",
"cvss": {
"@context": {
"cvssV2": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/cvss/cvssV2#"
},
"@id": "https://www.variotdbs.pl/ref/cvss/cvssV2"
},
"cvssV3": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/cvss/cvssV3#"
},
"@id": "https://www.variotdbs.pl/ref/cvss/cvssV3/"
},
"severity": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/cvss/severity#"
},
"@id": "https://www.variotdbs.pl/ref/cvss/severity"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
},
"@id": "https://www.variotdbs.pl/ref/sources"
}
},
"data": [
{
"cvssV2": [
{
"accessComplexity": "LOW",
"accessVector": "ADJACENT_NETWORK",
"authentication": "SINGLE",
"author": "nvd@nist.gov",
"availabilityImpact": "PARTIAL",
"baseScore": 5.2,
"confidentialityImpact": "PARTIAL",
"exploitabilityScore": 5.1,
"id": "CVE-2018-10596",
"impactScore": 6.4,
"integrityImpact": "PARTIAL",
"severity": "MEDIUM",
"trust": 1.8,
"vectorString": "AV:A/AC:L/Au:S/C:P/I:P/A:P",
"version": "2.0"
},
{
"accessComplexity": "HIGH",
"accessVector": "ADJACENT_NETWORK",
"authentication": "NONE",
"author": "CNVD",
"availabilityImpact": "COMPLETE",
"baseScore": 6.8,
"confidentialityImpact": "COMPLETE",
"exploitabilityScore": 3.2,
"id": "CNVD-2018-12557",
"impactScore": 10.0,
"integrityImpact": "COMPLETE",
"severity": "MEDIUM",
"trust": 0.6,
"vectorString": "AV:A/AC:H/Au:N/C:C/I:C/A:C",
"version": "2.0"
},
{
"accessComplexity": "LOW",
"accessVector": "ADJACENT_NETWORK",
"authentication": "SINGLE",
"author": "VULHUB",
"availabilityImpact": "PARTIAL",
"baseScore": 5.2,
"confidentialityImpact": "PARTIAL",
"exploitabilityScore": 5.1,
"id": "VHN-120371",
"impactScore": 6.4,
"integrityImpact": "PARTIAL",
"severity": "MEDIUM",
"trust": 0.1,
"vectorString": "AV:A/AC:L/AU:S/C:P/I:P/A:P",
"version": "2.0"
}
],
"cvssV3": [
{
"attackComplexity": "LOW",
"attackVector": "ADJACENT",
"author": "nvd@nist.gov",
"availabilityImpact": "HIGH",
"baseScore": 8.0,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"exploitabilityScore": 2.1,
"id": "CVE-2018-10596",
"impactScore": 5.9,
"integrityImpact": "HIGH",
"privilegesRequired": "LOW",
"scope": "UNCHANGED",
"trust": 1.8,
"userInteraction": "NONE",
"vectorString": "CVSS:3.0/AV:A/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H",
"version": "3.0"
},
{
"attackComplexity": "HIGH",
"attackVector": "NETWORK",
"author": "ics-cert@hq.dhs.gov",
"availabilityImpact": "HIGH",
"baseScore": 7.1,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"exploitabilityScore": 1.2,
"id": "CVE-2018-10596",
"impactScore": 5.9,
"integrityImpact": "HIGH",
"privilegesRequired": "LOW",
"scope": "UNCHANGED",
"trust": 1.0,
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H",
"version": "3.1"
}
],
"severity": [
{
"author": "nvd@nist.gov",
"id": "CVE-2018-10596",
"trust": 1.0,
"value": "HIGH"
},
{
"author": "ics-cert@hq.dhs.gov",
"id": "CVE-2018-10596",
"trust": 1.0,
"value": "HIGH"
},
{
"author": "NVD",
"id": "CVE-2018-10596",
"trust": 0.8,
"value": "High"
},
{
"author": "CNVD",
"id": "CNVD-2018-12557",
"trust": 0.6,
"value": "MEDIUM"
},
{
"author": "CNNVD",
"id": "CNNVD-201807-212",
"trust": 0.6,
"value": "HIGH"
},
{
"author": "VULHUB",
"id": "VHN-120371",
"trust": 0.1,
"value": "MEDIUM"
}
]
}
],
"sources": [
{
"db": "CNVD",
"id": "CNVD-2018-12557"
},
{
"db": "VULHUB",
"id": "VHN-120371"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-008035"
},
{
"db": "CNNVD",
"id": "CNNVD-201807-212"
},
{
"db": "NVD",
"id": "CVE-2018-10596"
},
{
"db": "NVD",
"id": "CVE-2018-10596"
}
]
},
"description": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/description#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "Medtronic 2090 CareLink Programmer \n\nuses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications. Medtronic 2090 CareLink Programmer Contains an information disclosure vulnerability.Information is obtained, information is altered, and service operation is disrupted (DoS) There is a possibility of being put into a state. Medtronic2090CareLinkProgrammer is a set of portable computer products from Medtronic Corporation of the United States. This product is used to manage and program cardiac devices in the medical industry",
"sources": [
{
"db": "NVD",
"id": "CVE-2018-10596"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-008035"
},
{
"db": "CNVD",
"id": "CNVD-2018-12557"
},
{
"db": "VULHUB",
"id": "VHN-120371"
}
],
"trust": 2.25
},
"external_ids": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/external_ids#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"db": "NVD",
"id": "CVE-2018-10596",
"trust": 3.1
},
{
"db": "ICS CERT",
"id": "ICSMA-18-058-01",
"trust": 3.1
},
{
"db": "JVNDB",
"id": "JVNDB-2018-008035",
"trust": 0.8
},
{
"db": "CNNVD",
"id": "CNNVD-201807-212",
"trust": 0.7
},
{
"db": "CNVD",
"id": "CNVD-2018-12557",
"trust": 0.6
},
{
"db": "NSFOCUS",
"id": "47493",
"trust": 0.6
},
{
"db": "VULHUB",
"id": "VHN-120371",
"trust": 0.1
}
],
"sources": [
{
"db": "CNVD",
"id": "CNVD-2018-12557"
},
{
"db": "VULHUB",
"id": "VHN-120371"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-008035"
},
{
"db": "CNNVD",
"id": "CNNVD-201807-212"
},
{
"db": "NVD",
"id": "CVE-2018-10596"
}
]
},
"id": "VAR-201807-0333",
"iot": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/iot#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": true,
"sources": [
{
"db": "CNVD",
"id": "CNVD-2018-12557"
},
{
"db": "VULHUB",
"id": "VHN-120371"
}
],
"trust": 1.5
},
"iot_taxonomy": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/iot_taxonomy#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"category": [
"Network device"
],
"sub_category": null,
"trust": 0.6
}
],
"sources": [
{
"db": "CNVD",
"id": "CNVD-2018-12557"
}
]
},
"last_update_date": "2025-05-23T23:07:14.887000Z",
"patch": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/patch#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"title": "Top Page",
"trust": 0.8,
"url": "http://www.medtronic.com/us-en/index.html"
}
],
"sources": [
{
"db": "JVNDB",
"id": "JVNDB-2018-008035"
}
]
},
"problemtype_data": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/problemtype_data#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"problemtype": "CWE-200",
"trust": 1.9
},
{
"problemtype": "CWE-923",
"trust": 1.0
}
],
"sources": [
{
"db": "VULHUB",
"id": "VHN-120371"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-008035"
},
{
"db": "NVD",
"id": "CVE-2018-10596"
}
]
},
"references": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/references#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"trust": 3.1,
"url": "https://ics-cert.us-cert.gov/advisories/icsma-18-058-01"
},
{
"trust": 1.0,
"url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-2090-29901.html"
},
{
"trust": 1.0,
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01"
},
{
"trust": 0.8,
"url": "https://cve.mitre.org/cgi-bin/cvename.cgi?name=cve-2018-10596"
},
{
"trust": 0.8,
"url": "https://nvd.nist.gov/vuln/detail/cve-2018-10596"
},
{
"trust": 0.6,
"url": "https://www.us-cert.gov/ics/advisories/icsma-18-058-01"
},
{
"trust": 0.6,
"url": "http://www.nsfocus.net/vulndb/47493"
}
],
"sources": [
{
"db": "CNVD",
"id": "CNVD-2018-12557"
},
{
"db": "VULHUB",
"id": "VHN-120371"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-008035"
},
{
"db": "CNNVD",
"id": "CNNVD-201807-212"
},
{
"db": "NVD",
"id": "CVE-2018-10596"
}
]
},
"sources": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#",
"data": {
"@container": "@list"
}
},
"data": [
{
"db": "CNVD",
"id": "CNVD-2018-12557"
},
{
"db": "VULHUB",
"id": "VHN-120371"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-008035"
},
{
"db": "CNNVD",
"id": "CNNVD-201807-212"
},
{
"db": "NVD",
"id": "CVE-2018-10596"
}
]
},
"sources_release_date": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources_release_date#",
"data": {
"@container": "@list"
}
},
"data": [
{
"date": "2018-07-05T00:00:00",
"db": "CNVD",
"id": "CNVD-2018-12557"
},
{
"date": "2018-07-03T00:00:00",
"db": "VULHUB",
"id": "VHN-120371"
},
{
"date": "2018-10-05T00:00:00",
"db": "JVNDB",
"id": "JVNDB-2018-008035"
},
{
"date": "2018-07-04T00:00:00",
"db": "CNNVD",
"id": "CNNVD-201807-212"
},
{
"date": "2018-07-03T01:29:00.487000",
"db": "NVD",
"id": "CVE-2018-10596"
}
]
},
"sources_update_date": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources_update_date#",
"data": {
"@container": "@list"
}
},
"data": [
{
"date": "2018-07-05T00:00:00",
"db": "CNVD",
"id": "CNVD-2018-12557"
},
{
"date": "2019-10-09T00:00:00",
"db": "VULHUB",
"id": "VHN-120371"
},
{
"date": "2018-10-05T00:00:00",
"db": "JVNDB",
"id": "JVNDB-2018-008035"
},
{
"date": "2020-08-07T00:00:00",
"db": "CNNVD",
"id": "CNNVD-201807-212"
},
{
"date": "2025-05-22T18:15:21.490000",
"db": "NVD",
"id": "CVE-2018-10596"
}
]
},
"threat_type": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/threat_type#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "remote or local",
"sources": [
{
"db": "CNNVD",
"id": "CNNVD-201807-212"
}
],
"trust": 0.6
},
"title": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/title#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "Medtronic 2090 CareLink Programmer Vulnerable to information disclosure",
"sources": [
{
"db": "JVNDB",
"id": "JVNDB-2018-008035"
}
],
"trust": 0.8
},
"type": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/type#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "information disclosure",
"sources": [
{
"db": "CNNVD",
"id": "CNNVD-201807-212"
}
],
"trust": 0.6
}
}
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.