Action not permitted
Modal body text goes here.
Modal Title
Modal Body
CVE-2018-8844 (GCVE-0-2018-8844)
Vulnerability from cvelistv5 – Published: 2018-09-26 19:00 – Updated: 2024-09-16 16:22- CWE-352 - CROSS-SITE REQUEST FORGERY (CSRF) CWE-352
| URL | Tags | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
|
|||||||||||
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Philips | e-Alert Unit (non-medical device) |
Affected:
R2.1 and prior
|
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-05T07:10:45.881Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_CONFIRM",
"x_transferred"
],
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"name": "105194",
"tags": [
"vdb-entry",
"x_refsource_BID",
"x_transferred"
],
"url": "http://www.securityfocus.com/bid/105194"
},
{
"tags": [
"x_refsource_MISC",
"x_transferred"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "e-Alert Unit (non-medical device)",
"vendor": "Philips",
"versions": [
{
"status": "affected",
"version": "R2.1 and prior"
}
]
}
],
"datePublic": "2018-08-31T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"value": "Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The web application does not, or cannot, sufficiently verify whether a well-formed, valid, consistent request was intentionally provided by the user who submitted the request."
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-352",
"description": "CROSS-SITE REQUEST FORGERY (CSRF) CWE-352",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2018-09-27T09:57:01.000Z",
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert"
},
"references": [
{
"tags": [
"x_refsource_CONFIRM"
],
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"name": "105194",
"tags": [
"vdb-entry",
"x_refsource_BID"
],
"url": "http://www.securityfocus.com/bid/105194"
},
{
"tags": [
"x_refsource_MISC"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01"
}
],
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2018-08-31T00:00:00",
"ID": "CVE-2018-8844",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "e-Alert Unit (non-medical device)",
"version": {
"version_data": [
{
"version_value": "R2.1 and prior"
}
]
}
}
]
},
"vendor_name": "Philips"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The web application does not, or cannot, sufficiently verify whether a well-formed, valid, consistent request was intentionally provided by the user who submitted the request."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CROSS-SITE REQUEST FORGERY (CSRF) CWE-352"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.usa.philips.com/healthcare/about/customer-support/product-security",
"refsource": "CONFIRM",
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"name": "105194",
"refsource": "BID",
"url": "http://www.securityfocus.com/bid/105194"
},
{
"name": "https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01",
"refsource": "MISC",
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01"
}
]
}
}
}
},
"cveMetadata": {
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"assignerShortName": "icscert",
"cveId": "CVE-2018-8844",
"datePublished": "2018-09-26T19:00:00.000Z",
"dateReserved": "2018-03-20T00:00:00.000Z",
"dateUpdated": "2024-09-16T16:22:34.701Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1",
"vulnerability-lookup:meta": {
"epss": {
"cve": "CVE-2018-8844",
"date": "2026-04-24",
"epss": "0.00149",
"percentile": "0.35113"
},
"nvd": "{\"cve\":{\"id\":\"CVE-2018-8844\",\"sourceIdentifier\":\"ics-cert@hq.dhs.gov\",\"published\":\"2018-09-26T19:29:01.037\",\"lastModified\":\"2024-11-21T04:14:26.177\",\"vulnStatus\":\"Modified\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The web application does not, or cannot, sufficiently verify whether a well-formed, valid, consistent request was intentionally provided by the user who submitted the request.\"},{\"lang\":\"es\",\"value\":\"Philips e-Alert Unit (dispositivo no m\u00e9dico), versiones R2.1 y anteriores. La aplicaci\u00f3n no verifica (o no puede verificar) lo suficiente si una petici\u00f3n consistente, v\u00e1lida y bien formada ha sido intencionadamente proporcionada por el usuario que envi\u00f3 la petici\u00f3n.\"}],\"metrics\":{\"cvssMetricV30\":[{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"3.0\",\"vectorString\":\"CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H\",\"baseScore\":8.8,\"baseSeverity\":\"HIGH\",\"attackVector\":\"NETWORK\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"REQUIRED\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":2.8,\"impactScore\":5.9}],\"cvssMetricV2\":[{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"cvssData\":{\"version\":\"2.0\",\"vectorString\":\"AV:N/AC:M/Au:N/C:P/I:P/A:P\",\"baseScore\":6.8,\"accessVector\":\"NETWORK\",\"accessComplexity\":\"MEDIUM\",\"authentication\":\"NONE\",\"confidentialityImpact\":\"PARTIAL\",\"integrityImpact\":\"PARTIAL\",\"availabilityImpact\":\"PARTIAL\"},\"baseSeverity\":\"MEDIUM\",\"exploitabilityScore\":8.6,\"impactScore\":6.4,\"acInsufInfo\":false,\"obtainAllPrivilege\":false,\"obtainUserPrivilege\":false,\"obtainOtherPrivilege\":false,\"userInteractionRequired\":true}]},\"weaknesses\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-352\"}]},{\"source\":\"nvd@nist.gov\",\"type\":\"Primary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-352\"}]}],\"configurations\":[{\"nodes\":[{\"operator\":\"OR\",\"negate\":false,\"cpeMatch\":[{\"vulnerable\":true,\"criteria\":\"cpe:2.3:o:philips:e-alert_firmware:*:*:*:*:*:*:*:*\",\"versionEndIncluding\":\"r2.1\",\"matchCriteriaId\":\"A20F1AAF-D947-4713-AB4E-12364592DBA3\"}]}]}],\"references\":[{\"url\":\"http://www.securityfocus.com/bid/105194\",\"source\":\"ics-cert@hq.dhs.gov\",\"tags\":[\"Third Party Advisory\",\"VDB Entry\"]},{\"url\":\"https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01\",\"source\":\"ics-cert@hq.dhs.gov\",\"tags\":[\"Mitigation\",\"Third Party Advisory\",\"US Government Resource\"]},{\"url\":\"https://www.usa.philips.com/healthcare/about/customer-support/product-security\",\"source\":\"ics-cert@hq.dhs.gov\",\"tags\":[\"Vendor Advisory\"]},{\"url\":\"http://www.securityfocus.com/bid/105194\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\",\"tags\":[\"Third Party Advisory\",\"VDB Entry\"]},{\"url\":\"https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\",\"tags\":[\"Mitigation\",\"Third Party Advisory\",\"US Government Resource\"]},{\"url\":\"https://www.usa.philips.com/healthcare/about/customer-support/product-security\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\",\"tags\":[\"Vendor Advisory\"]}]}}"
}
}
CNVD-2019-28280
Vulnerability from cnvd - Published: 2019-08-21目前厂商已发布升级补丁以修复漏洞,补丁获取链接: https://www.usa.philips.com/healthcare/about/customer-support/product-security
| Name | Philips e-Alert <=R2.1 |
|---|
{
"cves": {
"cve": {
"cveNumber": "CVE-2018-8844",
"cveUrl": "http://cve.mitre.org/cgi-bin/cvename.cgi?name=CVE-2018-8844"
}
},
"description": "Philips e-Alert\u662f\u8377\u5170\u98de\u5229\u6d66\uff08Philips\uff09\u516c\u53f8\u7684\u4e00\u6b3e\u9488\u5bf9MRI\u7cfb\u7edf\u7684\u7535\u5b50\u8b66\u62a5\u89e3\u51b3\u65b9\u6848\uff0c\u5b83\u4e3b\u8981\u7528\u4e8e\u76d1\u63a7MRI\u7cfb\u7edf\u6027\u80fd\u5e76\u53d1\u51fa\u8b66\u62a5\u3002\n\nPhilips e-Alert R2.1\u53ca\u4e4b\u524d\u7248\u672c\u4e2d\u5b58\u5728\u8de8\u7ad9\u8bf7\u6c42\u4f2a\u9020\u6f0f\u6d1e\u3002\u653b\u51fb\u8005\u53ef\u5229\u7528\u8be5\u6f0f\u6d1e\u6267\u884c\u672a\u6388\u6743\u7684\u64cd\u4f5c\u3002",
"discovererName": "Philips",
"formalWay": "\u76ee\u524d\u5382\u5546\u5df2\u53d1\u5e03\u5347\u7ea7\u8865\u4e01\u4ee5\u4fee\u590d\u6f0f\u6d1e\uff0c\u8865\u4e01\u83b7\u53d6\u94fe\u63a5:\r\nhttps://www.usa.philips.com/healthcare/about/customer-support/product-security",
"isEvent": "\u901a\u7528\u8f6f\u786c\u4ef6\u6f0f\u6d1e",
"number": "CNVD-2019-28280",
"openTime": "2019-08-21",
"patchDescription": "Philips e-Alert\u662f\u8377\u5170\u98de\u5229\u6d66\uff08Philips\uff09\u516c\u53f8\u7684\u4e00\u6b3e\u9488\u5bf9MRI\u7cfb\u7edf\u7684\u7535\u5b50\u8b66\u62a5\u89e3\u51b3\u65b9\u6848\uff0c\u5b83\u4e3b\u8981\u7528\u4e8e\u76d1\u63a7MRI\u7cfb\u7edf\u6027\u80fd\u5e76\u53d1\u51fa\u8b66\u62a5\u3002\r\n\r\nPhilips e-Alert R2.1\u53ca\u4e4b\u524d\u7248\u672c\u4e2d\u5b58\u5728\u8de8\u7ad9\u8bf7\u6c42\u4f2a\u9020\u6f0f\u6d1e\u3002\u653b\u51fb\u8005\u53ef\u5229\u7528\u8be5\u6f0f\u6d1e\u6267\u884c\u672a\u6388\u6743\u7684\u64cd\u4f5c\u3002\u76ee\u524d\uff0c\u4f9b\u5e94\u5546\u53d1\u5e03\u4e86\u5b89\u5168\u516c\u544a\u53ca\u76f8\u5173\u8865\u4e01\u4fe1\u606f\uff0c\u4fee\u590d\u4e86\u6b64\u6f0f\u6d1e\u3002",
"patchName": "Philips e-Alert\u8de8\u7ad9\u8bf7\u6c42\u4f2a\u9020\u6f0f\u6d1e\u7684\u8865\u4e01",
"products": {
"product": "Philips e-Alert \u003c=R2.1"
},
"referenceLink": "https://www.usa.philips.com/healthcare/about/customer-support/product-security",
"serverity": "\u4e2d",
"submitTime": "2018-09-05",
"title": "Philips e-Alert\u8de8\u7ad9\u8bf7\u6c42\u4f2a\u9020\u6f0f\u6d1e"
}
VAR-201809-1166
Vulnerability from variot - Updated: 2024-11-23 22:26Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The web application does not, or cannot, sufficiently verify whether a well-formed, valid, consistent request was intentionally provided by the user who submitted the request. Philips e-Alert Unit Contains a cross-site request forgery vulnerability.Information is obtained, information is altered, and service operation is disrupted (DoS) There is a possibility of being put into a state. Philips e-Alert is prone to the following security vulnerabilities: 1. An input-validation vulnerability 2. A cross-site scripting vulnerability 3. Multiple information-disclosure vulnerabilities 4. An insecure default permissions vulnerability 5. A cross-site request-forgery vulnerability 6. A session-fixation vulnerability 7. A denial-of-service vulnerability 8. A security-bypass vulnerability Attackers may exploit these issues to gain unauthorized access to the affected device, or to bypass certain security restrictions to perform unauthorized actions, to compromise the application to access or modify data and to exploit vulnerabilities in the underlying database, to execute arbitrary script code in the browser of an unsuspecting user in the context of the affected site or to execute arbitrary code within the context of the affected device. e-Alert R2.1 and prior are vulnerable. Philips e-Alert is an electronic alert solution for MRI systems from Philips, the Netherlands. It is mainly used to monitor the performance of MRI systems and issue alerts. There is a cross-site request forgery vulnerability in Philips e-Alert R2.1 and earlier versions. A remote attacker could exploit this vulnerability to perform unauthorized operations
Show details on source website{
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/VARIoTentry#",
"affected_products": {
"@id": "https://www.variotdbs.pl/ref/affected_products"
},
"configurations": {
"@id": "https://www.variotdbs.pl/ref/configurations"
},
"credits": {
"@id": "https://www.variotdbs.pl/ref/credits"
},
"cvss": {
"@id": "https://www.variotdbs.pl/ref/cvss/"
},
"description": {
"@id": "https://www.variotdbs.pl/ref/description/"
},
"exploit_availability": {
"@id": "https://www.variotdbs.pl/ref/exploit_availability/"
},
"external_ids": {
"@id": "https://www.variotdbs.pl/ref/external_ids/"
},
"iot": {
"@id": "https://www.variotdbs.pl/ref/iot/"
},
"iot_taxonomy": {
"@id": "https://www.variotdbs.pl/ref/iot_taxonomy/"
},
"patch": {
"@id": "https://www.variotdbs.pl/ref/patch/"
},
"problemtype_data": {
"@id": "https://www.variotdbs.pl/ref/problemtype_data/"
},
"references": {
"@id": "https://www.variotdbs.pl/ref/references/"
},
"sources": {
"@id": "https://www.variotdbs.pl/ref/sources/"
},
"sources_release_date": {
"@id": "https://www.variotdbs.pl/ref/sources_release_date/"
},
"sources_update_date": {
"@id": "https://www.variotdbs.pl/ref/sources_update_date/"
},
"threat_type": {
"@id": "https://www.variotdbs.pl/ref/threat_type/"
},
"title": {
"@id": "https://www.variotdbs.pl/ref/title/"
},
"type": {
"@id": "https://www.variotdbs.pl/ref/type/"
}
},
"@id": "https://www.variotdbs.pl/vuln/VAR-201809-1166",
"affected_products": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/affected_products#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
},
"@id": "https://www.variotdbs.pl/ref/sources"
}
},
"data": [
{
"model": "e-alert",
"scope": "lte",
"trust": 1.8,
"vendor": "philips",
"version": "r2.1"
},
{
"model": "e-alert",
"scope": "eq",
"trust": 0.6,
"vendor": "philips",
"version": "r2.1"
},
{
"model": "e-alert r2.1",
"scope": null,
"trust": 0.3,
"vendor": "philips",
"version": null
},
{
"model": "e-alert r2",
"scope": null,
"trust": 0.3,
"vendor": "philips",
"version": null
}
],
"sources": [
{
"db": "BID",
"id": "105194"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-010694"
},
{
"db": "CNNVD",
"id": "CNNVD-201809-115"
},
{
"db": "NVD",
"id": "CVE-2018-8844"
}
]
},
"configurations": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/configurations#",
"children": {
"@container": "@list"
},
"cpe_match": {
"@container": "@list"
},
"data": {
"@container": "@list"
},
"nodes": {
"@container": "@list"
}
},
"data": [
{
"CVE_data_version": "4.0",
"nodes": [
{
"cpe_match": [
{
"cpe22Uri": "cpe:/o:philips:e-alert_firmware",
"vulnerable": true
}
],
"operator": "OR"
}
]
}
],
"sources": [
{
"db": "JVNDB",
"id": "JVNDB-2018-010694"
}
]
},
"credits": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/credits#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "The vendor reported this issue.",
"sources": [
{
"db": "BID",
"id": "105194"
}
],
"trust": 0.3
},
"cve": "CVE-2018-8844",
"cvss": {
"@context": {
"cvssV2": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/cvss/cvssV2#"
},
"@id": "https://www.variotdbs.pl/ref/cvss/cvssV2"
},
"cvssV3": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/cvss/cvssV3#"
},
"@id": "https://www.variotdbs.pl/ref/cvss/cvssV3/"
},
"severity": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/cvss/severity#"
},
"@id": "https://www.variotdbs.pl/ref/cvss/severity"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
},
"@id": "https://www.variotdbs.pl/ref/sources"
}
},
"data": [
{
"cvssV2": [
{
"accessComplexity": "MEDIUM",
"accessVector": "NETWORK",
"authentication": "NONE",
"author": "nvd@nist.gov",
"availabilityImpact": "PARTIAL",
"baseScore": 6.8,
"confidentialityImpact": "PARTIAL",
"exploitabilityScore": 8.6,
"id": "CVE-2018-8844",
"impactScore": 6.4,
"integrityImpact": "PARTIAL",
"severity": "MEDIUM",
"trust": 1.8,
"vectorString": "AV:N/AC:M/Au:N/C:P/I:P/A:P",
"version": "2.0"
},
{
"accessComplexity": "MEDIUM",
"accessVector": "NETWORK",
"authentication": "NONE",
"author": "VULHUB",
"availabilityImpact": "PARTIAL",
"baseScore": 6.8,
"confidentialityImpact": "PARTIAL",
"exploitabilityScore": 8.6,
"id": "VHN-138876",
"impactScore": 6.4,
"integrityImpact": "PARTIAL",
"severity": "MEDIUM",
"trust": 0.1,
"vectorString": "AV:N/AC:M/AU:N/C:P/I:P/A:P",
"version": "2.0"
}
],
"cvssV3": [
{
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"author": "nvd@nist.gov",
"availabilityImpact": "HIGH",
"baseScore": 8.8,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"exploitabilityScore": 2.8,
"id": "CVE-2018-8844",
"impactScore": 5.9,
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"trust": 1.8,
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H",
"version": "3.0"
}
],
"severity": [
{
"author": "nvd@nist.gov",
"id": "CVE-2018-8844",
"trust": 1.0,
"value": "HIGH"
},
{
"author": "NVD",
"id": "CVE-2018-8844",
"trust": 0.8,
"value": "High"
},
{
"author": "CNNVD",
"id": "CNNVD-201809-115",
"trust": 0.6,
"value": "HIGH"
},
{
"author": "VULHUB",
"id": "VHN-138876",
"trust": 0.1,
"value": "MEDIUM"
}
]
}
],
"sources": [
{
"db": "VULHUB",
"id": "VHN-138876"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-010694"
},
{
"db": "CNNVD",
"id": "CNNVD-201809-115"
},
{
"db": "NVD",
"id": "CVE-2018-8844"
}
]
},
"description": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/description#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The web application does not, or cannot, sufficiently verify whether a well-formed, valid, consistent request was intentionally provided by the user who submitted the request. Philips e-Alert Unit Contains a cross-site request forgery vulnerability.Information is obtained, information is altered, and service operation is disrupted (DoS) There is a possibility of being put into a state. Philips e-Alert is prone to the following security vulnerabilities:\n1. An input-validation vulnerability\n2. A cross-site scripting vulnerability\n3. Multiple information-disclosure vulnerabilities\n4. An insecure default permissions vulnerability\n5. A cross-site request-forgery vulnerability\n6. A session-fixation vulnerability\n7. A denial-of-service vulnerability\n8. A security-bypass vulnerability\nAttackers may exploit these issues to gain unauthorized access to the affected device, or to bypass certain security restrictions to perform unauthorized actions, to compromise the application to access or modify data and to exploit vulnerabilities in the underlying database, to execute arbitrary script code in the browser of an unsuspecting user in the context of the affected site or to execute arbitrary code within the context of the affected device. \ne-Alert R2.1 and prior are vulnerable. Philips e-Alert is an electronic alert solution for MRI systems from Philips, the Netherlands. It is mainly used to monitor the performance of MRI systems and issue alerts. There is a cross-site request forgery vulnerability in Philips e-Alert R2.1 and earlier versions. A remote attacker could exploit this vulnerability to perform unauthorized operations",
"sources": [
{
"db": "NVD",
"id": "CVE-2018-8844"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-010694"
},
{
"db": "BID",
"id": "105194"
},
{
"db": "VULHUB",
"id": "VHN-138876"
}
],
"trust": 1.98
},
"external_ids": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/external_ids#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"db": "ICS CERT",
"id": "ICSA-18-242-01",
"trust": 2.8
},
{
"db": "NVD",
"id": "CVE-2018-8844",
"trust": 2.8
},
{
"db": "BID",
"id": "105194",
"trust": 2.0
},
{
"db": "JVNDB",
"id": "JVNDB-2018-010694",
"trust": 0.8
},
{
"db": "CNNVD",
"id": "CNNVD-201809-115",
"trust": 0.7
},
{
"db": "VULHUB",
"id": "VHN-138876",
"trust": 0.1
}
],
"sources": [
{
"db": "VULHUB",
"id": "VHN-138876"
},
{
"db": "BID",
"id": "105194"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-010694"
},
{
"db": "CNNVD",
"id": "CNNVD-201809-115"
},
{
"db": "NVD",
"id": "CVE-2018-8844"
}
]
},
"id": "VAR-201809-1166",
"iot": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/iot#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": true,
"sources": [
{
"db": "VULHUB",
"id": "VHN-138876"
}
],
"trust": 0.01
},
"last_update_date": "2024-11-23T22:26:12.985000Z",
"patch": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/patch#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"title": "Philips e-Alert Unit Vulnerabilities (30-AUG-2018)",
"trust": 0.8,
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"title": "Philips e-Alert Fixes for cross-site request forgery vulnerabilities",
"trust": 0.6,
"url": "http://www.cnnvd.org.cn/web/xxk/bdxqById.tag?id=84476"
}
],
"sources": [
{
"db": "JVNDB",
"id": "JVNDB-2018-010694"
},
{
"db": "CNNVD",
"id": "CNNVD-201809-115"
}
]
},
"problemtype_data": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/problemtype_data#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"problemtype": "CWE-352",
"trust": 1.9
}
],
"sources": [
{
"db": "VULHUB",
"id": "VHN-138876"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-010694"
},
{
"db": "NVD",
"id": "CVE-2018-8844"
}
]
},
"references": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/references#",
"data": {
"@container": "@list"
},
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": [
{
"trust": 2.8,
"url": "https://ics-cert.us-cert.gov/advisories/icsa-18-242-01"
},
{
"trust": 1.7,
"url": "http://www.securityfocus.com/bid/105194"
},
{
"trust": 1.7,
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"trust": 0.8,
"url": "https://cve.mitre.org/cgi-bin/cvename.cgi?name=cve-2018-8844"
},
{
"trust": 0.8,
"url": "https://nvd.nist.gov/vuln/detail/cve-2018-8844"
}
],
"sources": [
{
"db": "VULHUB",
"id": "VHN-138876"
},
{
"db": "BID",
"id": "105194"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-010694"
},
{
"db": "CNNVD",
"id": "CNNVD-201809-115"
},
{
"db": "NVD",
"id": "CVE-2018-8844"
}
]
},
"sources": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#",
"data": {
"@container": "@list"
}
},
"data": [
{
"db": "VULHUB",
"id": "VHN-138876"
},
{
"db": "BID",
"id": "105194"
},
{
"db": "JVNDB",
"id": "JVNDB-2018-010694"
},
{
"db": "CNNVD",
"id": "CNNVD-201809-115"
},
{
"db": "NVD",
"id": "CVE-2018-8844"
}
]
},
"sources_release_date": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources_release_date#",
"data": {
"@container": "@list"
}
},
"data": [
{
"date": "2018-09-26T00:00:00",
"db": "VULHUB",
"id": "VHN-138876"
},
{
"date": "2018-08-30T00:00:00",
"db": "BID",
"id": "105194"
},
{
"date": "2018-12-20T00:00:00",
"db": "JVNDB",
"id": "JVNDB-2018-010694"
},
{
"date": "2018-09-04T00:00:00",
"db": "CNNVD",
"id": "CNNVD-201809-115"
},
{
"date": "2018-09-26T19:29:01.037000",
"db": "NVD",
"id": "CVE-2018-8844"
}
]
},
"sources_update_date": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources_update_date#",
"data": {
"@container": "@list"
}
},
"data": [
{
"date": "2019-10-09T00:00:00",
"db": "VULHUB",
"id": "VHN-138876"
},
{
"date": "2018-08-30T00:00:00",
"db": "BID",
"id": "105194"
},
{
"date": "2018-12-20T00:00:00",
"db": "JVNDB",
"id": "JVNDB-2018-010694"
},
{
"date": "2019-10-17T00:00:00",
"db": "CNNVD",
"id": "CNNVD-201809-115"
},
{
"date": "2024-11-21T04:14:26.177000",
"db": "NVD",
"id": "CVE-2018-8844"
}
]
},
"threat_type": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/threat_type#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "remote",
"sources": [
{
"db": "CNNVD",
"id": "CNNVD-201809-115"
}
],
"trust": 0.6
},
"title": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/title#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "Philips e-Alert Unit Vulnerable to cross-site request forgery",
"sources": [
{
"db": "JVNDB",
"id": "JVNDB-2018-010694"
}
],
"trust": 0.8
},
"type": {
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/type#",
"sources": {
"@container": "@list",
"@context": {
"@vocab": "https://www.variotdbs.pl/ref/sources#"
}
}
},
"data": "cross-site request forgery",
"sources": [
{
"db": "CNNVD",
"id": "CNNVD-201809-115"
}
],
"trust": 0.6
}
}
ICSA-18-242-01
Vulnerability from csaf_cisa - Published: 2018-08-30 00:00 - Updated: 2018-08-30 00:00{
"document": {
"acknowledgments": [
{
"organization": "Philips",
"summary": "reporting these vulnerabilities to NCCIC"
}
],
"category": "csaf_security_advisory",
"csaf_version": "2.0",
"distribution": {
"text": "Disclosure is not limited",
"tlp": {
"label": "WHITE",
"url": "https://us-cert.cisa.gov/tlp/"
}
},
"lang": "en-US",
"notes": [
{
"category": "general",
"text": "This CSAF advisory was extracted from unstructured data and may contain inaccuracies. If you notice any errors, please reach out to the designated contact at CISA CSAF: central@cisa.dhs.gov",
"title": "CISA Disclaimer"
},
{
"category": "legal_disclaimer",
"text": "All information products included in https://us-cert.cisa.gov/ics are provided \"as is\" for informational purposes only. The Department of Homeland Security (DHS) does not provide any warranties of any kind regarding any information contained within. DHS does not endorse any commercial product or service, referenced in this product or otherwise. Further dissemination of this product is governed by the Traffic Light Protocol (TLP) marking in the header. For more information about TLP, see https://us-cert.cisa.gov/tlp/.",
"title": "Legal Notice"
},
{
"category": "summary",
"text": "Successful exploitation from an attacker within the same subnet may impact or compromise user contact details, unit integrity, and/or unit availability. The vulnerabilities may allow attackers to provide unexpected input into the application, execute arbitrary code, display unit information, or potentially cause e-Alert to crash.",
"title": "Risk evaluation"
},
{
"category": "other",
"text": "Healthcare and Public Health",
"title": "Critical infrastructure sectors"
},
{
"category": "other",
"text": "Worldwide",
"title": "Countries/areas deployed"
},
{
"category": "other",
"text": "Netherlands",
"title": "Company headquarters location"
},
{
"category": "general",
"text": "NCCIC recommends users take defensive measures to minimize the risk of exploitation of this vulnerability. Specifically, users should:",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.NCCIC also provides a section for control systems security recommended practices on the ICS-CERT web page. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.",
"title": "Recommended Practices"
},
{
"category": "general",
"text": "Additional mitigation guidance and recommended practices are publicly available on the ICS-CERT website in the Technical Information Paper, ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation Strategies.Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to NCCIC for tracking and correlation against other incidents.",
"title": "Recommended Practices"
},
{
"category": "other",
"text": "No known public exploits specifically target these vulnerabilities.",
"title": "Exploitability"
}
],
"publisher": {
"category": "coordinator",
"contact_details": "Email: CISAservicedesk@cisa.dhs.gov;\n Toll Free: 1-888-282-0870",
"name": "CISA",
"namespace": "https://www.cisa.gov/"
},
"references": [
{
"category": "self",
"summary": "ICS Advisory ICSA-18-242-01 JSON",
"url": "https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2018/icsa-18-242-01.json"
},
{
"category": "self",
"summary": "ICS Advisory ICSA-18-242-01 Web Version",
"url": "https://www.cisa.gov/news-events/ics-advisories/icsa-18-242-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/uscert/ics/alerts/ICS-ALERT-10-301-01"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/uscert/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf"
},
{
"category": "external",
"summary": "Recommended Practices",
"url": "https://www.cisa.gov/uscert/ics/tips/ICS-TIP-12-146-01B"
}
],
"title": "Philips e-Alert Unit",
"tracking": {
"current_release_date": "2018-08-30T00:00:00.000000Z",
"generator": {
"engine": {
"name": "CISA CSAF Generator",
"version": "1.0.0"
}
},
"id": "ICSA-18-242-01",
"initial_release_date": "2018-08-30T00:00:00.000000Z",
"revision_history": [
{
"date": "2018-08-30T00:00:00.000000Z",
"legacy_version": "Initial",
"number": "1",
"summary": "ICSA-18-242-01 Philips e-Alert Unit"
},
{
"date": "2018-08-30T00:00:00.000000Z",
"legacy_version": "A",
"number": "2",
"summary": "ICSA-18-242-01 Philips e-Alert Unit (Update A)"
}
],
"status": "final",
"version": "2"
}
},
"product_tree": {
"branches": [
{
"branches": [
{
"branches": [
{
"category": "product_version_range",
"name": "\u003c= R2.1",
"product": {
"name": "Philips e-Alert Unit (non-medical device): Version R2.1 and prior",
"product_id": "CSAFPID-0001"
}
}
],
"category": "product_name",
"name": "Philips e-Alert Unit (non-medical device)"
}
],
"category": "vendor",
"name": "Philips"
}
]
},
"vulnerabilities": [
{
"cve": "CVE-2018-8850",
"cwe": {
"id": "CWE-20",
"name": "Improper Input Validation"
},
"notes": [
{
"category": "summary",
"text": "The software does not validate input properly, allowing an attacker to craft the input in a form that is not expected by the rest of the application. This would lead to parts of the unit receiving unintended input, which may result in altered control flow, arbitrary control of a resource, or arbitrary code execution.CVE-2018-8850 has been assigned to this vulnerability. A CVSS v3 base score of 7.1 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:H/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8850"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:H/A:N"
}
],
"remediations": [
{
"category": "mitigation",
"details": "Please see the Philips product security website for the latest public security information on this matter and for other Philips products:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.philips.com/productsecurity"
},
{
"category": "mitigation",
"details": "Ensure that network security best practices are implemented, and",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Limit network access to e-Alert in accordance with product documentation.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Users with questions regarding their specific e-Alert installations should contact their local Philips service support team or their regional e-Alert service support. Contact information is available at the following location:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.1,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:H/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2018-8846",
"cwe": {
"id": "CWE-79",
"name": "Improper Neutralization of Input During Web Page Generation (\u0027Cross-site Scripting\u0027)"
},
"notes": [
{
"category": "summary",
"text": "The software does not neutralize or incorrectly neutralizes user-controllable input before it is placed in output that is used as a web page that is then served to other users.CVE-2018-8846 has been assigned to this vulnerability. A CVSS v3 base score of 7.1 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:H/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8846"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:H/A:N"
}
],
"remediations": [
{
"category": "mitigation",
"details": "Please see the Philips product security website for the latest public security information on this matter and for other Philips products:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.philips.com/productsecurity"
},
{
"category": "mitigation",
"details": "Ensure that network security best practices are implemented, and",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Limit network access to e-Alert in accordance with product documentation.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Users with questions regarding their specific e-Alert installations should contact their local Philips service support team or their regional e-Alert service support. Contact information is available at the following location:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.1,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:H/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2018-14803",
"cwe": {
"id": "CWE-200",
"name": "Exposure of Sensitive Information to an Unauthorized Actor"
},
"notes": [
{
"category": "summary",
"text": "The Philips e-Alert contains a banner disclosure vulnerability that could allow attackers to obtain extraneous product information, such as OS and software components, via the HTTP response header that is normally not available to the attacker, but might be useful information in an attack.CVE-2018-14803has been assigned to this vulnerability. A CVSS v3 base score of 5.3 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-14803"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N"
}
],
"remediations": [
{
"category": "mitigation",
"details": "Please see the Philips product security website for the latest public security information on this matter and for other Philips products:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.philips.com/productsecurity"
},
{
"category": "mitigation",
"details": "Ensure that network security best practices are implemented, and",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Limit network access to e-Alert in accordance with product documentation.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Users with questions regarding their specific e-Alert installations should contact their local Philips service support team or their regional e-Alert service support. Contact information is available at the following location:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 5.3,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2018-8848",
"cwe": {
"id": "CWE-276",
"name": "Incorrect Default Permissions"
},
"notes": [
{
"category": "summary",
"text": "The software, upon installation, sets incorrect permissions for an object that exposes it to an unintended actor.CVE-2018-8848 has been assigned to this vulnerability. A CVSS v3 base score of 7.1 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:H/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8848"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:H/A:N"
}
],
"remediations": [
{
"category": "mitigation",
"details": "Please see the Philips product security website for the latest public security information on this matter and for other Philips products:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.philips.com/productsecurity"
},
{
"category": "mitigation",
"details": "Ensure that network security best practices are implemented, and",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Limit network access to e-Alert in accordance with product documentation.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Users with questions regarding their specific e-Alert installations should contact their local Philips service support team or their regional e-Alert service support. Contact information is available at the following location:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.1,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:H/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2018-8842",
"cwe": {
"id": "CWE-319",
"name": "Cleartext Transmission of Sensitive Information"
},
"notes": [
{
"category": "summary",
"text": "The software transmits sensitive or security-critical data in cleartext in a communication channel that can be sniffed by unauthorized actors. The Philips e-Alert communication channel is not encrypted which could therefore lead to disclosure of personal contact information and application login credentials from within the same subnet.CVE-2018-8842 has been assigned to this vulnerability. A CVSS v3 base score of 7.5 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8842"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N"
}
],
"remediations": [
{
"category": "mitigation",
"details": "Please see the Philips product security website for the latest public security information on this matter and for other Philips products:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.philips.com/productsecurity"
},
{
"category": "mitigation",
"details": "Ensure that network security best practices are implemented, and",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Limit network access to e-Alert in accordance with product documentation.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Users with questions regarding their specific e-Alert installations should contact their local Philips service support team or their regional e-Alert service support. Contact information is available at the following location:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.5,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2018-8844",
"cwe": {
"id": "CWE-352",
"name": "Cross-Site Request Forgery (CSRF)"
},
"notes": [
{
"category": "summary",
"text": "The web application does not, or cannot, sufficiently verify whether a well-formed, valid, consistent request was intentionally provided by the user who submitted the request.CVE-2018-8844 has been assigned to this vulnerability. A CVSS v3 base score of 6.8 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:N/UI:R/S:U/C:N/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8844"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:H/PR:N/UI:R/S:U/C:N/I:H/A:H"
}
],
"remediations": [
{
"category": "mitigation",
"details": "Please see the Philips product security website for the latest public security information on this matter and for other Philips products:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.philips.com/productsecurity"
},
{
"category": "mitigation",
"details": "Ensure that network security best practices are implemented, and",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Limit network access to e-Alert in accordance with product documentation.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Users with questions regarding their specific e-Alert installations should contact their local Philips service support team or their regional e-Alert service support. Contact information is available at the following location:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.8,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:N/AC:H/PR:N/UI:R/S:U/C:N/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2018-8852",
"cwe": {
"id": "CWE-384",
"name": "Session Fixation"
},
"notes": [
{
"category": "summary",
"text": "When authenticating a user or otherwise establishing a new user session, the software gives an attacker the opportunity to steal authenticated sessions without invalidating any existing session identifier.CVE-2018-8852 has been assigned to this vulnerability. A CVSS v3 base score of 6.4 has been calculated; the CVSS vector string is (AV:N/AC:H/PR:H/UI:R/S:U/C:H/I:H/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8852"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:H/PR:H/UI:R/S:U/C:H/I:H/A:H"
}
],
"remediations": [
{
"category": "mitigation",
"details": "Please see the Philips product security website for the latest public security information on this matter and for other Philips products:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.philips.com/productsecurity"
},
{
"category": "mitigation",
"details": "Ensure that network security best practices are implemented, and",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Limit network access to e-Alert in accordance with product documentation.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Users with questions regarding their specific e-Alert installations should contact their local Philips service support team or their regional e-Alert service support. Contact information is available at the following location:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 6.4,
"baseSeverity": "MEDIUM",
"vectorString": "CVSS:3.0/AV:N/AC:H/PR:H/UI:R/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
},
{
"cve": "CVE-2018-8854",
"cwe": {
"id": "CWE-400",
"name": "Uncontrolled Resource Consumption"
},
"notes": [
{
"category": "summary",
"text": "The software does not properly restrict the size or amount of resources requested or influenced by an actor, which can be used to consume more resources than intended.CVE-2018-8854 has been assigned to this vulnerability. A CVSS v3 base score of 7.5 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).",
"title": "Vulnerability Summary"
}
],
"product_status": {
"known_affected": [
"CSAFPID-0001"
]
},
"references": [
{
"category": "external",
"summary": "web.nvd.nist.gov",
"url": "http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8854"
},
{
"category": "external",
"summary": "www.first.org",
"url": "https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H"
}
],
"remediations": [
{
"category": "mitigation",
"details": "Please see the Philips product security website for the latest public security information on this matter and for other Philips products:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.philips.com/productsecurity"
},
{
"category": "mitigation",
"details": "Ensure that network security best practices are implemented, and",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Limit network access to e-Alert in accordance with product documentation.",
"product_ids": [
"CSAFPID-0001"
]
},
{
"category": "mitigation",
"details": "Users with questions regarding their specific e-Alert installations should contact their local Philips service support team or their regional e-Alert service support. Contact information is available at the following location:",
"product_ids": [
"CSAFPID-0001"
],
"url": "https://www.usa.philips.com/healthcare/solutions/customer-service-solutions"
}
],
"scores": [
{
"cvss_v3": {
"baseScore": 7.5,
"baseSeverity": "HIGH",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.0"
},
"products": [
"CSAFPID-0001"
]
}
]
}
]
}
GSD-2018-8844
Vulnerability from gsd - Updated: 2023-12-13 01:22{
"GSD": {
"alias": "CVE-2018-8844",
"description": "Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The web application does not, or cannot, sufficiently verify whether a well-formed, valid, consistent request was intentionally provided by the user who submitted the request.",
"id": "GSD-2018-8844"
},
"gsd": {
"metadata": {
"exploitCode": "unknown",
"remediation": "unknown",
"reportConfidence": "confirmed",
"type": "vulnerability"
},
"osvSchema": {
"aliases": [
"CVE-2018-8844"
],
"details": "Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The web application does not, or cannot, sufficiently verify whether a well-formed, valid, consistent request was intentionally provided by the user who submitted the request.",
"id": "GSD-2018-8844",
"modified": "2023-12-13T01:22:34.715953Z",
"schema_version": "1.4.0"
}
},
"namespaces": {
"cve.org": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2018-08-31T00:00:00",
"ID": "CVE-2018-8844",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "e-Alert Unit (non-medical device)",
"version": {
"version_data": [
{
"version_value": "R2.1 and prior"
}
]
}
}
]
},
"vendor_name": "Philips"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The web application does not, or cannot, sufficiently verify whether a well-formed, valid, consistent request was intentionally provided by the user who submitted the request."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CROSS-SITE REQUEST FORGERY (CSRF) CWE-352"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.usa.philips.com/healthcare/about/customer-support/product-security",
"refsource": "CONFIRM",
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"name": "105194",
"refsource": "BID",
"url": "http://www.securityfocus.com/bid/105194"
},
{
"name": "https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01",
"refsource": "MISC",
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01"
}
]
}
},
"nvd.nist.gov": {
"configurations": {
"CVE_data_version": "4.0",
"nodes": [
{
"children": [],
"cpe_match": [
{
"cpe23Uri": "cpe:2.3:o:philips:e-alert_firmware:*:*:*:*:*:*:*:*",
"cpe_name": [],
"versionEndIncluding": "r2.1",
"vulnerable": true
}
],
"operator": "OR"
}
]
},
"cve": {
"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"ID": "CVE-2018-8844"
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "en",
"value": "Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The web application does not, or cannot, sufficiently verify whether a well-formed, valid, consistent request was intentionally provided by the user who submitted the request."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "en",
"value": "CWE-352"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://www.usa.philips.com/healthcare/about/customer-support/product-security",
"refsource": "CONFIRM",
"tags": [
"Vendor Advisory"
],
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"name": "https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01",
"refsource": "MISC",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01"
},
{
"name": "105194",
"refsource": "BID",
"tags": [
"Third Party Advisory",
"VDB Entry"
],
"url": "http://www.securityfocus.com/bid/105194"
}
]
}
},
"impact": {
"baseMetricV2": {
"cvssV2": {
"accessComplexity": "MEDIUM",
"accessVector": "NETWORK",
"authentication": "NONE",
"availabilityImpact": "PARTIAL",
"baseScore": 6.8,
"confidentialityImpact": "PARTIAL",
"integrityImpact": "PARTIAL",
"vectorString": "AV:N/AC:M/Au:N/C:P/I:P/A:P",
"version": "2.0"
},
"exploitabilityScore": 8.6,
"impactScore": 6.4,
"obtainAllPrivilege": false,
"obtainOtherPrivilege": false,
"obtainUserPrivilege": false,
"severity": "MEDIUM",
"userInteractionRequired": true
},
"baseMetricV3": {
"cvssV3": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 8.8,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"exploitabilityScore": 2.8,
"impactScore": 5.9
}
},
"lastModifiedDate": "2019-10-09T23:42Z",
"publishedDate": "2018-09-26T19:29Z"
}
}
}
FKIE_CVE-2018-8844
Vulnerability from fkie_nvd - Published: 2018-09-26 19:29 - Updated: 2024-11-21 04:14| URL | Tags | ||
|---|---|---|---|
| ics-cert@hq.dhs.gov | http://www.securityfocus.com/bid/105194 | Third Party Advisory, VDB Entry | |
| ics-cert@hq.dhs.gov | https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01 | Mitigation, Third Party Advisory, US Government Resource | |
| ics-cert@hq.dhs.gov | https://www.usa.philips.com/healthcare/about/customer-support/product-security | Vendor Advisory | |
| af854a3a-2127-422b-91ae-364da2661108 | http://www.securityfocus.com/bid/105194 | Third Party Advisory, VDB Entry | |
| af854a3a-2127-422b-91ae-364da2661108 | https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01 | Mitigation, Third Party Advisory, US Government Resource | |
| af854a3a-2127-422b-91ae-364da2661108 | https://www.usa.philips.com/healthcare/about/customer-support/product-security | Vendor Advisory |
| Vendor | Product | Version | |
|---|---|---|---|
| philips | e-alert_firmware | * |
{
"configurations": [
{
"nodes": [
{
"cpeMatch": [
{
"criteria": "cpe:2.3:o:philips:e-alert_firmware:*:*:*:*:*:*:*:*",
"matchCriteriaId": "A20F1AAF-D947-4713-AB4E-12364592DBA3",
"versionEndIncluding": "r2.1",
"vulnerable": true
}
],
"negate": false,
"operator": "OR"
}
]
}
],
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The web application does not, or cannot, sufficiently verify whether a well-formed, valid, consistent request was intentionally provided by the user who submitted the request."
},
{
"lang": "es",
"value": "Philips e-Alert Unit (dispositivo no m\u00e9dico), versiones R2.1 y anteriores. La aplicaci\u00f3n no verifica (o no puede verificar) lo suficiente si una petici\u00f3n consistente, v\u00e1lida y bien formada ha sido intencionadamente proporcionada por el usuario que envi\u00f3 la petici\u00f3n."
}
],
"id": "CVE-2018-8844",
"lastModified": "2024-11-21T04:14:26.177",
"metrics": {
"cvssMetricV2": [
{
"acInsufInfo": false,
"baseSeverity": "MEDIUM",
"cvssData": {
"accessComplexity": "MEDIUM",
"accessVector": "NETWORK",
"authentication": "NONE",
"availabilityImpact": "PARTIAL",
"baseScore": 6.8,
"confidentialityImpact": "PARTIAL",
"integrityImpact": "PARTIAL",
"vectorString": "AV:N/AC:M/Au:N/C:P/I:P/A:P",
"version": "2.0"
},
"exploitabilityScore": 8.6,
"impactScore": 6.4,
"obtainAllPrivilege": false,
"obtainOtherPrivilege": false,
"obtainUserPrivilege": false,
"source": "nvd@nist.gov",
"type": "Primary",
"userInteractionRequired": true
}
],
"cvssMetricV30": [
{
"cvssData": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 8.8,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H",
"version": "3.0"
},
"exploitabilityScore": 2.8,
"impactScore": 5.9,
"source": "nvd@nist.gov",
"type": "Primary"
}
]
},
"published": "2018-09-26T19:29:01.037",
"references": [
{
"source": "ics-cert@hq.dhs.gov",
"tags": [
"Third Party Advisory",
"VDB Entry"
],
"url": "http://www.securityfocus.com/bid/105194"
},
{
"source": "ics-cert@hq.dhs.gov",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01"
},
{
"source": "ics-cert@hq.dhs.gov",
"tags": [
"Vendor Advisory"
],
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"source": "af854a3a-2127-422b-91ae-364da2661108",
"tags": [
"Third Party Advisory",
"VDB Entry"
],
"url": "http://www.securityfocus.com/bid/105194"
},
{
"source": "af854a3a-2127-422b-91ae-364da2661108",
"tags": [
"Mitigation",
"Third Party Advisory",
"US Government Resource"
],
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01"
},
{
"source": "af854a3a-2127-422b-91ae-364da2661108",
"tags": [
"Vendor Advisory"
],
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
}
],
"sourceIdentifier": "ics-cert@hq.dhs.gov",
"vulnStatus": "Modified",
"weaknesses": [
{
"description": [
{
"lang": "en",
"value": "CWE-352"
}
],
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary"
},
{
"description": [
{
"lang": "en",
"value": "CWE-352"
}
],
"source": "nvd@nist.gov",
"type": "Primary"
}
]
}
GHSA-QQ7J-J4J8-WF6V
Vulnerability from github – Published: 2022-05-13 01:31 – Updated: 2022-05-13 01:31Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The web application does not, or cannot, sufficiently verify whether a well-formed, valid, consistent request was intentionally provided by the user who submitted the request.
{
"affected": [],
"aliases": [
"CVE-2018-8844"
],
"database_specific": {
"cwe_ids": [
"CWE-352"
],
"github_reviewed": false,
"github_reviewed_at": null,
"nvd_published_at": "2018-09-26T19:29:00Z",
"severity": "HIGH"
},
"details": "Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The web application does not, or cannot, sufficiently verify whether a well-formed, valid, consistent request was intentionally provided by the user who submitted the request.",
"id": "GHSA-qq7j-j4j8-wf6v",
"modified": "2022-05-13T01:31:49Z",
"published": "2022-05-13T01:31:49Z",
"references": [
{
"type": "ADVISORY",
"url": "https://nvd.nist.gov/vuln/detail/CVE-2018-8844"
},
{
"type": "WEB",
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01"
},
{
"type": "WEB",
"url": "https://www.usa.philips.com/healthcare/about/customer-support/product-security"
},
{
"type": "WEB",
"url": "http://www.securityfocus.com/bid/105194"
}
],
"schema_version": "1.4.0",
"severity": [
{
"score": "CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H",
"type": "CVSS_V3"
}
]
}
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.