CVE-2018-4846 (GCVE-0-2018-4846)
Vulnerability from nvd – Published: 2018-06-26 18:00 – Updated: 2024-09-17 01:01
VLAI?
Summary
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
Severity ?
No CVSS data available.
CWE
- CWE-798 - Use of Hard-coded Credentials
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Siemens AG | RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systems |
Affected:
RAPIDLab 1200 systems RAPIDPoint 400 systems / RAPIDPoint 500 systems All versions _without_ use of Siemens Healthineers Informatics products RAPIDLab 1200 Series All versions < V3.3 _with_Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions >= V3.0 _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems V2.4.X _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions =< V2.3 _with_ Siemens Healthineers Informatics products RAPIDPoint 400 systems All versions_with_ Siemens Healthineers Informatics products
|
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-05T05:18:26.455Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_CONFIRM",
"x_transferred"
],
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systems",
"vendor": "Siemens AG",
"versions": [
{
"status": "affected",
"version": "RAPIDLab 1200 systems RAPIDPoint 400 systems / RAPIDPoint 500 systems All versions _without_ use of Siemens Healthineers Informatics products RAPIDLab 1200 Series All versions \u003c V3.3 _with_Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems V2.4.X _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products RAPIDPoint 400 systems All versions_with_ Siemens Healthineers Informatics products"
}
]
}
],
"datePublic": "2018-06-15T00:00:00",
"descriptions": [
{
"lang": "en",
"value": "A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions \u003c V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue."
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-798",
"description": "CWE-798: Use of Hard-coded Credentials",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2018-06-26T17:57:01",
"orgId": "cec7a2ec-15b4-4faf-bd53-b40f371f3a77",
"shortName": "siemens"
},
"references": [
{
"tags": [
"x_refsource_CONFIRM"
],
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf"
}
],
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "productcert@siemens.com",
"DATE_PUBLIC": "2018-06-15T00:00:00",
"ID": "CVE-2018-4846",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systems",
"version": {
"version_data": [
{
"version_value": "RAPIDLab 1200 systems RAPIDPoint 400 systems / RAPIDPoint 500 systems All versions _without_ use of Siemens Healthineers Informatics products RAPIDLab 1200 Series All versions \u003c V3.3 _with_Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems V2.4.X _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products RAPIDPoint 400 systems All versions_with_ Siemens Healthineers Informatics products"
}
]
}
}
]
},
"vendor_name": "Siemens AG"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions \u003c V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-798: Use of Hard-coded Credentials"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf",
"refsource": "CONFIRM",
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf"
}
]
}
}
}
},
"cveMetadata": {
"assignerOrgId": "cec7a2ec-15b4-4faf-bd53-b40f371f3a77",
"assignerShortName": "siemens",
"cveId": "CVE-2018-4846",
"datePublished": "2018-06-26T18:00:00Z",
"dateReserved": "2018-01-02T00:00:00",
"dateUpdated": "2024-09-17T01:01:45.175Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2018-4845 (GCVE-0-2018-4845)
Vulnerability from nvd – Published: 2018-06-26 18:00 – Updated: 2024-09-16 18:55
VLAI?
Summary
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). Remote attackers with either local or remote credentialed access to the "Remote View" feature might be able to elevate their privileges, compromising confidentiality, integrity, and availability of the system. No special skills or user interaction are required to perform this attack. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
Severity ?
No CVSS data available.
CWE
- CWE-284 - Improper Access Control
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Siemens AG | RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems |
Affected:
RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems : All versions _without_ use of Siemens Healthineers Informatics products
Affected: RAPIDLab 1200 Series : All versions < V3.3 _with_ Siemens Healthineers Informatics products Affected: RAPIDPoint 500 systems : All versions >= V3.0 _with_ Siemens Healthineers Informatics products Affected: RAPIDPoint 500 systems : V2.4.X _with_ Siemens Healthineers Informatics products Affected: RAPIDPoint 500 systems : All versions =< V2.3 _with_ Siemens Healthineers Informatics products Affected: RAPIDPoint 400 systems : All versions_with_ Siemens Healthineers Informatics products |
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-05T05:18:26.591Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_CONFIRM",
"x_transferred"
],
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems",
"vendor": "Siemens AG",
"versions": [
{
"status": "affected",
"version": "RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems : All versions _without_ use of Siemens Healthineers Informatics products"
},
{
"status": "affected",
"version": "RAPIDLab 1200 Series : All versions \u003c V3.3 _with_ Siemens Healthineers Informatics products"
},
{
"status": "affected",
"version": "RAPIDPoint 500 systems : All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products"
},
{
"status": "affected",
"version": "RAPIDPoint 500 systems : V2.4.X _with_ Siemens Healthineers Informatics products"
},
{
"status": "affected",
"version": "RAPIDPoint 500 systems : All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products"
},
{
"status": "affected",
"version": "RAPIDPoint 400 systems : All versions_with_ Siemens Healthineers Informatics products"
}
]
}
],
"datePublic": "2018-06-15T00:00:00",
"descriptions": [
{
"lang": "en",
"value": "A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions \u003c V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). Remote attackers with either local or remote credentialed access to the \"Remote View\" feature might be able to elevate their privileges, compromising confidentiality, integrity, and availability of the system. No special skills or user interaction are required to perform this attack. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue."
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-284",
"description": "CWE-284: Improper Access Control",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2018-06-26T17:57:01",
"orgId": "cec7a2ec-15b4-4faf-bd53-b40f371f3a77",
"shortName": "siemens"
},
"references": [
{
"tags": [
"x_refsource_CONFIRM"
],
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf"
}
],
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "productcert@siemens.com",
"DATE_PUBLIC": "2018-06-15T00:00:00",
"ID": "CVE-2018-4845",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems",
"version": {
"version_data": [
{
"version_value": "RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems : All versions _without_ use of Siemens Healthineers Informatics products"
},
{
"version_value": "RAPIDLab 1200 Series : All versions \u003c V3.3 _with_ Siemens Healthineers Informatics products"
},
{
"version_value": "RAPIDPoint 500 systems : All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products"
},
{
"version_value": "RAPIDPoint 500 systems : V2.4.X _with_ Siemens Healthineers Informatics products"
},
{
"version_value": "RAPIDPoint 500 systems : All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products"
},
{
"version_value": "RAPIDPoint 400 systems : All versions_with_ Siemens Healthineers Informatics products"
}
]
}
}
]
},
"vendor_name": "Siemens AG"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions \u003c V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). Remote attackers with either local or remote credentialed access to the \"Remote View\" feature might be able to elevate their privileges, compromising confidentiality, integrity, and availability of the system. No special skills or user interaction are required to perform this attack. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-284: Improper Access Control"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf",
"refsource": "CONFIRM",
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf"
}
]
}
}
}
},
"cveMetadata": {
"assignerOrgId": "cec7a2ec-15b4-4faf-bd53-b40f371f3a77",
"assignerShortName": "siemens",
"cveId": "CVE-2018-4845",
"datePublished": "2018-06-26T18:00:00Z",
"dateReserved": "2018-01-02T00:00:00",
"dateUpdated": "2024-09-16T18:55:43.476Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2018-4845 (GCVE-0-2018-4845)
Vulnerability from cvelistv5 – Published: 2018-06-26 18:00 – Updated: 2024-09-16 18:55
VLAI?
Summary
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). Remote attackers with either local or remote credentialed access to the "Remote View" feature might be able to elevate their privileges, compromising confidentiality, integrity, and availability of the system. No special skills or user interaction are required to perform this attack. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
Severity ?
No CVSS data available.
CWE
- CWE-284 - Improper Access Control
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Siemens AG | RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems |
Affected:
RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems : All versions _without_ use of Siemens Healthineers Informatics products
Affected: RAPIDLab 1200 Series : All versions < V3.3 _with_ Siemens Healthineers Informatics products Affected: RAPIDPoint 500 systems : All versions >= V3.0 _with_ Siemens Healthineers Informatics products Affected: RAPIDPoint 500 systems : V2.4.X _with_ Siemens Healthineers Informatics products Affected: RAPIDPoint 500 systems : All versions =< V2.3 _with_ Siemens Healthineers Informatics products Affected: RAPIDPoint 400 systems : All versions_with_ Siemens Healthineers Informatics products |
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-05T05:18:26.591Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_CONFIRM",
"x_transferred"
],
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems",
"vendor": "Siemens AG",
"versions": [
{
"status": "affected",
"version": "RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems : All versions _without_ use of Siemens Healthineers Informatics products"
},
{
"status": "affected",
"version": "RAPIDLab 1200 Series : All versions \u003c V3.3 _with_ Siemens Healthineers Informatics products"
},
{
"status": "affected",
"version": "RAPIDPoint 500 systems : All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products"
},
{
"status": "affected",
"version": "RAPIDPoint 500 systems : V2.4.X _with_ Siemens Healthineers Informatics products"
},
{
"status": "affected",
"version": "RAPIDPoint 500 systems : All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products"
},
{
"status": "affected",
"version": "RAPIDPoint 400 systems : All versions_with_ Siemens Healthineers Informatics products"
}
]
}
],
"datePublic": "2018-06-15T00:00:00",
"descriptions": [
{
"lang": "en",
"value": "A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions \u003c V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). Remote attackers with either local or remote credentialed access to the \"Remote View\" feature might be able to elevate their privileges, compromising confidentiality, integrity, and availability of the system. No special skills or user interaction are required to perform this attack. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue."
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-284",
"description": "CWE-284: Improper Access Control",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2018-06-26T17:57:01",
"orgId": "cec7a2ec-15b4-4faf-bd53-b40f371f3a77",
"shortName": "siemens"
},
"references": [
{
"tags": [
"x_refsource_CONFIRM"
],
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf"
}
],
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "productcert@siemens.com",
"DATE_PUBLIC": "2018-06-15T00:00:00",
"ID": "CVE-2018-4845",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems",
"version": {
"version_data": [
{
"version_value": "RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems : All versions _without_ use of Siemens Healthineers Informatics products"
},
{
"version_value": "RAPIDLab 1200 Series : All versions \u003c V3.3 _with_ Siemens Healthineers Informatics products"
},
{
"version_value": "RAPIDPoint 500 systems : All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products"
},
{
"version_value": "RAPIDPoint 500 systems : V2.4.X _with_ Siemens Healthineers Informatics products"
},
{
"version_value": "RAPIDPoint 500 systems : All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products"
},
{
"version_value": "RAPIDPoint 400 systems : All versions_with_ Siemens Healthineers Informatics products"
}
]
}
}
]
},
"vendor_name": "Siemens AG"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions \u003c V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). Remote attackers with either local or remote credentialed access to the \"Remote View\" feature might be able to elevate their privileges, compromising confidentiality, integrity, and availability of the system. No special skills or user interaction are required to perform this attack. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-284: Improper Access Control"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf",
"refsource": "CONFIRM",
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf"
}
]
}
}
}
},
"cveMetadata": {
"assignerOrgId": "cec7a2ec-15b4-4faf-bd53-b40f371f3a77",
"assignerShortName": "siemens",
"cveId": "CVE-2018-4845",
"datePublished": "2018-06-26T18:00:00Z",
"dateReserved": "2018-01-02T00:00:00",
"dateUpdated": "2024-09-16T18:55:43.476Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}
CVE-2018-4846 (GCVE-0-2018-4846)
Vulnerability from cvelistv5 – Published: 2018-06-26 18:00 – Updated: 2024-09-17 01:01
VLAI?
Summary
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
Severity ?
No CVSS data available.
CWE
- CWE-798 - Use of Hard-coded Credentials
Assigner
References
| URL | Tags | ||||
|---|---|---|---|---|---|
|
|||||
Impacted products
| Vendor | Product | Version | ||
|---|---|---|---|---|
| Siemens AG | RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systems |
Affected:
RAPIDLab 1200 systems RAPIDPoint 400 systems / RAPIDPoint 500 systems All versions _without_ use of Siemens Healthineers Informatics products RAPIDLab 1200 Series All versions < V3.3 _with_Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions >= V3.0 _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems V2.4.X _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions =< V2.3 _with_ Siemens Healthineers Informatics products RAPIDPoint 400 systems All versions_with_ Siemens Healthineers Informatics products
|
{
"containers": {
"adp": [
{
"providerMetadata": {
"dateUpdated": "2024-08-05T05:18:26.455Z",
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE"
},
"references": [
{
"tags": [
"x_refsource_CONFIRM",
"x_transferred"
],
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf"
}
],
"title": "CVE Program Container"
}
],
"cna": {
"affected": [
{
"product": "RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systems",
"vendor": "Siemens AG",
"versions": [
{
"status": "affected",
"version": "RAPIDLab 1200 systems RAPIDPoint 400 systems / RAPIDPoint 500 systems All versions _without_ use of Siemens Healthineers Informatics products RAPIDLab 1200 Series All versions \u003c V3.3 _with_Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems V2.4.X _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products RAPIDPoint 400 systems All versions_with_ Siemens Healthineers Informatics products"
}
]
}
],
"datePublic": "2018-06-15T00:00:00",
"descriptions": [
{
"lang": "en",
"value": "A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions \u003c V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue."
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-798",
"description": "CWE-798: Use of Hard-coded Credentials",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2018-06-26T17:57:01",
"orgId": "cec7a2ec-15b4-4faf-bd53-b40f371f3a77",
"shortName": "siemens"
},
"references": [
{
"tags": [
"x_refsource_CONFIRM"
],
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf"
}
],
"x_legacyV4Record": {
"CVE_data_meta": {
"ASSIGNER": "productcert@siemens.com",
"DATE_PUBLIC": "2018-06-15T00:00:00",
"ID": "CVE-2018-4846",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systems",
"version": {
"version_data": [
{
"version_value": "RAPIDLab 1200 systems RAPIDPoint 400 systems / RAPIDPoint 500 systems All versions _without_ use of Siemens Healthineers Informatics products RAPIDLab 1200 Series All versions \u003c V3.3 _with_Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems V2.4.X _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products RAPIDPoint 400 systems All versions_with_ Siemens Healthineers Informatics products"
}
]
}
}
]
},
"vendor_name": "Siemens AG"
}
]
}
},
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions \u003c V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions \u003e= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =\u003c V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-798: Use of Hard-coded Credentials"
}
]
}
]
},
"references": {
"reference_data": [
{
"name": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf",
"refsource": "CONFIRM",
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf"
}
]
}
}
}
},
"cveMetadata": {
"assignerOrgId": "cec7a2ec-15b4-4faf-bd53-b40f371f3a77",
"assignerShortName": "siemens",
"cveId": "CVE-2018-4846",
"datePublished": "2018-06-26T18:00:00Z",
"dateReserved": "2018-01-02T00:00:00",
"dateUpdated": "2024-09-17T01:01:45.175Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.1"
}