    FKIE_CVE-2022-41627

    Vulnerability from fkie_nvd - Published: 2022-10-27 21:15 - Updated: 2026-06-17 05:03
    Summary
    The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.

    {
      "affected": [
        {
          "affectedData": [
            {
              "defaultStatus": "unaffected",
              "product": "KardiaMobile",
              "vendor": "AliveCor",
              "versions": [
                {
                  "status": "affected",
                  "version": "All"
                }
              ]
            }
          ],
          "source": "ics-cert@hq.dhs.gov"
        }
      ],
      "configurations": [
        {
          "nodes": [
            {
              "cpeMatch": [
                {
                  "criteria": "cpe:2.3:o:alivecor:kardiamobile_firmware:-:*:*:*:*:*:*:*",
                  "matchCriteriaId": "6EADECC1-EF47-43B5-9213-CA92945DE4A8",
                  "vulnerable": true
                }
              ],
              "negate": false,
              "operator": "OR"
            },
            {
              "cpeMatch": [
                {
                  "criteria": "cpe:2.3:h:alivecor:kardiamobile:-:*:*:*:*:*:*:*",
                  "matchCriteriaId": "C5ED3887-643F-430C-B2A1-CCEDBE71F2B6",
                  "vulnerable": false
                }
              ],
              "negate": false,
              "operator": "OR"
            }
          ],
          "operator": "AND"
        },
        {
          "nodes": [
            {
              "cpeMatch": [
                {
                  "criteria": "cpe:2.3:o:alivecor:kardiamobile_6l_firmware:-:*:*:*:*:*:*:*",
                  "matchCriteriaId": "37A5B811-EFDF-4497-9E4C-D2D663C5A15B",
                  "vulnerable": true
                }
              ],
              "negate": false,
              "operator": "OR"
            },
            {
              "cpeMatch": [
                {
                  "criteria": "cpe:2.3:h:alivecor:kardiamobile_6l:-:*:*:*:*:*:*:*",
                  "matchCriteriaId": "EB03F4F8-687B-493D-BBEA-553831EBBA43",
                  "vulnerable": false
                }
              ],
              "negate": false,
              "operator": "OR"
            }
          ],
          "operator": "AND"
        },
        {
          "nodes": [
            {
              "cpeMatch": [
                {
                  "criteria": "cpe:2.3:o:alivecor:kardiamobile_card_firmware:-:*:*:*:*:*:*:*",
                  "matchCriteriaId": "7FF05B50-32FB-4BCE-9C84-BDB46CDAC1D8",
                  "vulnerable": true
                }
              ],
              "negate": false,
              "operator": "OR"
            },
            {
              "cpeMatch": [
                {
                  "criteria": "cpe:2.3:h:alivecor:kardiamobile_card:-:*:*:*:*:*:*:*",
                  "matchCriteriaId": "18DF9CF0-6D64-474C-A65C-5003893A5C79",
                  "vulnerable": false
                }
              ],
              "negate": false,
              "operator": "OR"
            }
          ],
          "operator": "AND"
        }
      ],
      "cveTags": [],
      "descriptions": [
        {
          "lang": "en",
          "value": "\nThe physical IoT device of the AliveCor\u0027s KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone\u2019s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.\n\n"
        },
        {
          "lang": "es",
          "value": "El dispositivo f\u00edsico IoT de AliveCor\u0027s KardiaMobile, basado en un tel\u00e9fono inteligente de electrocardiograma personal (EKG), no tiene cifrado para sus protocolos de datos sobre sonido. Explotar esta vulnerabilidad podr\u00eda permitir a un atacante leer los resultados del EKG del paciente o crear una condici\u00f3n de Denegaci\u00f3n de Servicio al emitir sonidos en frecuencias similares a las del dispositivo, interrumpiendo la capacidad del micr\u00f3fono del tel\u00e9fono inteligente para leer los datos con precisi\u00f3n. Para llevar a cabo este ataque, el atacante debe estar cerca (a menos de 5 pies) para captar y emitir ondas sonoras."
        }
      ],
      "id": "CVE-2022-41627",
      "lastModified": "2026-06-17T05:03:32.063",
      "metrics": {
        "cvssMetricV31": [
          {
            "cvssData": {
              "attackComplexity": "HIGH",
              "attackVector": "PHYSICAL",
              "availabilityImpact": "HIGH",
              "baseScore": 4.8,
              "baseSeverity": "MEDIUM",
              "confidentialityImpact": "NONE",
              "integrityImpact": "LOW",
              "privilegesRequired": "NONE",
              "scope": "UNCHANGED",
              "userInteraction": "NONE",
              "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:L/A:H",
              "version": "3.1"
            },
            "exploitabilityScore": 0.5,
            "impactScore": 4.2,
            "source": "ics-cert@hq.dhs.gov",
            "type": "Secondary"
          },
          {
            "cvssData": {
              "attackComplexity": "LOW",
              "attackVector": "ADJACENT_NETWORK",
              "availabilityImpact": "HIGH",
              "baseScore": 7.6,
              "baseSeverity": "HIGH",
              "confidentialityImpact": "LOW",
              "integrityImpact": "LOW",
              "privilegesRequired": "NONE",
              "scope": "UNCHANGED",
              "userInteraction": "NONE",
              "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H",
              "version": "3.1"
            },
            "exploitabilityScore": 2.8,
            "impactScore": 4.7,
            "source": "nvd@nist.gov",
            "type": "Primary"
          }
        ],
        "ssvcV203": [
          {
            "source": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
            "ssvcData": {
              "id": "CVE-2022-41627",
              "options": [
                {
                  "exploitation": "none"
                },
                {
                  "automatable": "no"
                },
                {
                  "technicalImpact": "partial"
                }
              ],
              "role": "CISA Coordinator",
              "timestamp": "2025-04-16T15:49:32.342453Z",
              "version": "2.0.3"
            }
          }
        ]
      },
      "published": "2022-10-27T21:15:15.573",
      "references": [
        {
          "source": "ics-cert@hq.dhs.gov",
          "tags": [
            "Third Party Advisory",
            "US Government Resource"
          ],
          "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-298-01"
        },
        {
          "source": "af854a3a-2127-422b-91ae-364da2661108",
          "tags": [
            "Third Party Advisory",
            "US Government Resource"
          ],
          "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-298-01"
        }
      ],
      "sourceIdentifier": "ics-cert@hq.dhs.gov",
      "vulnStatus": "Modified",
      "weaknesses": [
        {
          "description": [
            {
              "lang": "en",
              "value": "CWE-311"
            }
          ],
          "source": "ics-cert@hq.dhs.gov",
          "type": "Secondary"
        },
        {
          "description": [
            {
              "lang": "en",
              "value": "CWE-319"
            }
          ],
          "source": "nvd@nist.gov",
          "type": "Primary"
        }
      ]
    }

    CVE-2022-41627 (GCVE-0-2022-41627)

    Vulnerability from nvd – Published: 2022-10-27 20:04 – Updated: 2025-04-16 16:08
    VLAI
    Summary
    The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator · CISA-ADP (v2.0.3)
    Decision recorded 2025-04-16 15:49 UTC
    CWE
    • CWE-311 - Missing Encryption of Sensitive Data
    Assigner
    icscert CNA under the icscert root
    CNA scorecard B 75/100 over 347 records in the last 180 days details
    References
    Impacted products
    Date Public
    2022-10-25 18:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-03T12:49:43.451Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "government-resource",
                  "x_transferred"
                ],
                "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-298-01"
              }
            ],
            "title": "CVE Program Container"
          },
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2022-41627",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2025-04-16T15:49:32.342453Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2025-04-16T16:08:30.914Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "KardiaMobile",
              "vendor": "AliveCor",
              "versions": [
                {
                  "status": "affected",
                  "version": "All"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "user": "00000000-0000-4000-9000-000000000000",
              "value": "Carlos Cilleruelo Rodr\u00edguez"
            },
            {
              "lang": "en",
              "type": "finder",
              "user": "00000000-0000-4000-9000-000000000000",
              "value": "Javier Junquera S\u00e1nchez"
            }
          ],
          "datePublic": "2022-10-25T18:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eThe physical IoT device of the AliveCor\u0027s KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone\u2019s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.\u003c/span\u003e\n\n"
                }
              ],
              "value": "\nThe physical IoT device of the AliveCor\u0027s KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone\u2019s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.\n\n"
            }
          ],
          "metrics": [
            {
              "cvssV3_1": {
                "attackComplexity": "HIGH",
                "attackVector": "PHYSICAL",
                "availabilityImpact": "HIGH",
                "baseScore": 4.8,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "LOW",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:L/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-311",
                  "description": "CWE-311 Missing Encryption of Sensitive Data",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2022-11-01T15:57:59.910Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "government-resource"
              ],
              "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-298-01"
            }
          ],
          "source": {
            "discovery": "EXTERNAL"
          },
          "x_generator": {
            "engine": "cveClient/1.0.13"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2022-41627",
        "datePublished": "2022-10-27T20:04:06.498Z",
        "dateReserved": "2022-09-29T14:09:27.495Z",
        "dateUpdated": "2025-04-16T16:08:30.914Z",
        "requesterUserId": "bc31a57b-b1a5-40e2-9263-67c0ae8a3b8a",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2022-41627 (GCVE-0-2022-41627)

    Vulnerability from cvelistv5 – Published: 2022-10-27 20:04 – Updated: 2025-04-16 16:08
    VLAI
    Summary
    The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator · CISA-ADP (v2.0.3)
    Decision recorded 2025-04-16 15:49 UTC
    CWE
    • CWE-311 - Missing Encryption of Sensitive Data
    Assigner
    icscert CNA under the icscert root
    CNA scorecard B 75/100 over 347 records in the last 180 days details
    References
    Impacted products
    Date Public
    2022-10-25 18:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-03T12:49:43.451Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "government-resource",
                  "x_transferred"
                ],
                "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-298-01"
              }
            ],
            "title": "CVE Program Container"
          },
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2022-41627",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2025-04-16T15:49:32.342453Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2025-04-16T16:08:30.914Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "KardiaMobile",
              "vendor": "AliveCor",
              "versions": [
                {
                  "status": "affected",
                  "version": "All"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "user": "00000000-0000-4000-9000-000000000000",
              "value": "Carlos Cilleruelo Rodr\u00edguez"
            },
            {
              "lang": "en",
              "type": "finder",
              "user": "00000000-0000-4000-9000-000000000000",
              "value": "Javier Junquera S\u00e1nchez"
            }
          ],
          "datePublic": "2022-10-25T18:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eThe physical IoT device of the AliveCor\u0027s KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone\u2019s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.\u003c/span\u003e\n\n"
                }
              ],
              "value": "\nThe physical IoT device of the AliveCor\u0027s KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone\u2019s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.\n\n"
            }
          ],
          "metrics": [
            {
              "cvssV3_1": {
                "attackComplexity": "HIGH",
                "attackVector": "PHYSICAL",
                "availabilityImpact": "HIGH",
                "baseScore": 4.8,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "LOW",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:L/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-311",
                  "description": "CWE-311 Missing Encryption of Sensitive Data",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2022-11-01T15:57:59.910Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "government-resource"
              ],
              "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-298-01"
            }
          ],
          "source": {
            "discovery": "EXTERNAL"
          },
          "x_generator": {
            "engine": "cveClient/1.0.13"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2022-41627",
        "datePublished": "2022-10-27T20:04:06.498Z",
        "dateReserved": "2022-09-29T14:09:27.495Z",
        "dateUpdated": "2025-04-16T16:08:30.914Z",
        "requesterUserId": "bc31a57b-b1a5-40e2-9263-67c0ae8a3b8a",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }