    CVE-2019-25718 (GCVE-0-2019-25718)

    Vulnerability from nvd – Published: 2026-06-01 21:41 – Updated: 2026-06-03 20:11
    VLAI
    Title
    Dräger Infinity Explorer C700 Privilege Escalation via Kiosk Mode Bypass
    Summary
    Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.
    SSVC
    Exploitation: none Automatable: no Technical Impact: total
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-451 - User Interface (UI) Misrepresentation of Critical Information
    Assigner
    References
    Impacted products
    Vendor Product Version
    Dräger Infinity Explorer C700 Affected: Infinity Explorer C700 (custom)
    Create a notification for this product.
    Date Public
    2019-01-22 00:00
    Credits
    Marc Ruef and Rocco Gagliardi, scip AG
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2019-25718",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "total"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-02T12:28:44.610787Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-02T12:29:05.596Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Explorer C700",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Explorer C700",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Marc Ruef and Rocco Gagliardi, scip AG"
            }
          ],
          "datePublic": "2019-01-22T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eDr\u00e4ger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.\u003c/p\u003e"
                }
              ],
              "value": "Dr\u00e4ger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "LOCAL",
                "baseScore": 8.6,
                "baseSeverity": "HIGH",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "HIGH",
                "vulnConfidentialityImpact": "HIGH",
                "vulnIntegrityImpact": "HIGH",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "LOCAL",
                "availabilityImpact": "HIGH",
                "baseScore": 8.4,
                "baseSeverity": "HIGH",
                "confidentialityImpact": "HIGH",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-451",
                  "description": "CWE-451 User Interface (UI) Misrepresentation of Critical Information",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-03T20:11:15.447Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security/download/2019-01-22-draeger-infinity-delta-vf10-1-security-advisory.pdf"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/dr-ger-infinity-explorer-c700-privilege-escalation-via-kiosk-mode-bypass"
            }
          ],
          "source": {
            "discovery": "EXTERNAL"
          },
          "title": "Dr\u00e4ger Infinity Explorer C700 Privilege Escalation via Kiosk Mode Bypass",
          "x_generator": {
            "engine": "Vulnogram 1.0.2"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2019-25718",
        "datePublished": "2026-06-01T21:41:34.625Z",
        "dateReserved": "2026-06-01T21:36:41.544Z",
        "dateUpdated": "2026-06-03T20:11:15.447Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2018-19014 (GCVE-0-2018-19014)

    Vulnerability from nvd – Published: 2019-01-28 22:00 – Updated: 2024-09-16 20:31
    VLAI
    Summary
    Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.
    Severity
    No CVSS data available.
    CWE
    • CWE-532 - INFORMATION EXPOSURE THROUGH LOG FILES CWE-532
    Assigner
    References
    Impacted products
    Vendor Product Version
    ICS-CERT Dräger Infinity Delta Affected: Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions.
    Create a notification for this product.
    Date Public
    2019-01-22 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-05T11:23:08.949Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
              },
              {
                "name": "106683",
                "tags": [
                  "vdb-entry",
                  "x_refsource_BID",
                  "x_transferred"
                ],
                "url": "http://www.securityfocus.com/bid/106683"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Dr\u0026#195;\u0026#164;ger Infinity Delta",
              "vendor": "ICS-CERT",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions."
                }
              ]
            }
          ],
          "datePublic": "2019-01-22T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-532",
                  "description": "INFORMATION EXPOSURE THROUGH LOG FILES CWE-532",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-01-29T10:57:01.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
            },
            {
              "name": "106683",
              "tags": [
                "vdb-entry",
                "x_refsource_BID"
              ],
              "url": "http://www.securityfocus.com/bid/106683"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "DATE_PUBLIC": "2019-01-22T00:00:00",
              "ID": "CVE-2018-19014",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Dr\u0026#195;\u0026#164;ger Infinity Delta",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions."
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "ICS-CERT"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "INFORMATION EXPOSURE THROUGH LOG FILES CWE-532"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
                },
                {
                  "name": "106683",
                  "refsource": "BID",
                  "url": "http://www.securityfocus.com/bid/106683"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2018-19014",
        "datePublished": "2019-01-28T22:00:00.000Z",
        "dateReserved": "2018-11-06T00:00:00.000Z",
        "dateUpdated": "2024-09-16T20:31:32.329Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2018-19010 (GCVE-0-2018-19010)

    Vulnerability from nvd – Published: 2019-01-28 22:00 – Updated: 2024-09-16 17:32
    VLAI
    Summary
    Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.
    Severity
    No CVSS data available.
    CWE
    • CWE-20 - IMPROPER INPUT VALIDATION CWE-20
    Assigner
    References
    Impacted products
    Vendor Product Version
    ICS-CERT Dräger Infinity Delta Affected: Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions.
    Create a notification for this product.
    Date Public
    2019-01-22 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-05T11:23:09.031Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
              },
              {
                "name": "106683",
                "tags": [
                  "vdb-entry",
                  "x_refsource_BID",
                  "x_transferred"
                ],
                "url": "http://www.securityfocus.com/bid/106683"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Dr\u0026#195;\u0026#164;ger Infinity Delta",
              "vendor": "ICS-CERT",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions."
                }
              ]
            }
          ],
          "datePublic": "2019-01-22T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-20",
                  "description": "IMPROPER INPUT VALIDATION CWE-20",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-01-29T10:57:01.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
            },
            {
              "name": "106683",
              "tags": [
                "vdb-entry",
                "x_refsource_BID"
              ],
              "url": "http://www.securityfocus.com/bid/106683"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "DATE_PUBLIC": "2019-01-22T00:00:00",
              "ID": "CVE-2018-19010",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Dr\u0026#195;\u0026#164;ger Infinity Delta",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions."
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "ICS-CERT"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "IMPROPER INPUT VALIDATION CWE-20"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
                },
                {
                  "name": "106683",
                  "refsource": "BID",
                  "url": "http://www.securityfocus.com/bid/106683"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2018-19010",
        "datePublished": "2019-01-28T22:00:00.000Z",
        "dateReserved": "2018-11-06T00:00:00.000Z",
        "dateUpdated": "2024-09-16T17:32:40.689Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2018-19012 (GCVE-0-2018-19012)

    Vulnerability from nvd – Published: 2019-01-28 21:00 – Updated: 2024-09-17 04:08
    VLAI
    Summary
    Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Via a specific dialog it is possible to break out of the kiosk mode and reach the underlying operating system. By breaking out of the kiosk mode, an attacker is able to take control of the operating system.
    Severity
    No CVSS data available.
    CWE
    • CWE-269 - IMPROPER PRIVILEGE MANAGEMENT CWE-269
    Assigner
    References
    Impacted products
    Vendor Product Version
    ICS-CERT Dräger Infinity Delta Affected: Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions.
    Create a notification for this product.
    Date Public
    2018-01-22 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-05T11:23:08.579Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
              },
              {
                "name": "106683",
                "tags": [
                  "vdb-entry",
                  "x_refsource_BID",
                  "x_transferred"
                ],
                "url": "http://www.securityfocus.com/bid/106683"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Dr\u0026#195;\u0026#164;ger Infinity Delta",
              "vendor": "ICS-CERT",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions."
                }
              ]
            }
          ],
          "datePublic": "2018-01-22T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Via a specific dialog it is possible to break out of the kiosk mode and reach the underlying operating system. By breaking out of the kiosk mode, an attacker is able to take control of the operating system."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-269",
                  "description": "IMPROPER PRIVILEGE MANAGEMENT CWE-269",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-01-29T10:57:01.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
            },
            {
              "name": "106683",
              "tags": [
                "vdb-entry",
                "x_refsource_BID"
              ],
              "url": "http://www.securityfocus.com/bid/106683"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "DATE_PUBLIC": "2018-01-22T00:00:00",
              "ID": "CVE-2018-19012",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Dr\u0026#195;\u0026#164;ger Infinity Delta",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions."
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "ICS-CERT"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Via a specific dialog it is possible to break out of the kiosk mode and reach the underlying operating system. By breaking out of the kiosk mode, an attacker is able to take control of the operating system."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "IMPROPER PRIVILEGE MANAGEMENT CWE-269"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
                },
                {
                  "name": "106683",
                  "refsource": "BID",
                  "url": "http://www.securityfocus.com/bid/106683"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2018-19012",
        "datePublished": "2019-01-28T21:00:00.000Z",
        "dateReserved": "2018-11-06T00:00:00.000Z",
        "dateUpdated": "2024-09-17T04:08:49.934Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2019-25718 (GCVE-0-2019-25718)

    Vulnerability from cvelistv5 – Published: 2026-06-01 21:41 – Updated: 2026-06-03 20:11
    VLAI
    Title
    Dräger Infinity Explorer C700 Privilege Escalation via Kiosk Mode Bypass
    Summary
    Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.
    SSVC
    Exploitation: none Automatable: no Technical Impact: total
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-451 - User Interface (UI) Misrepresentation of Critical Information
    Assigner
    References
    Impacted products
    Vendor Product Version
    Dräger Infinity Explorer C700 Affected: Infinity Explorer C700 (custom)
    Create a notification for this product.
    Date Public
    2019-01-22 00:00
    Credits
    Marc Ruef and Rocco Gagliardi, scip AG
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2019-25718",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "total"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-02T12:28:44.610787Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-02T12:29:05.596Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "Infinity Explorer C700",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Explorer C700",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Marc Ruef and Rocco Gagliardi, scip AG"
            }
          ],
          "datePublic": "2019-01-22T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eDr\u00e4ger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.\u003c/p\u003e"
                }
              ],
              "value": "Dr\u00e4ger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "LOCAL",
                "baseScore": 8.6,
                "baseSeverity": "HIGH",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "HIGH",
                "vulnConfidentialityImpact": "HIGH",
                "vulnIntegrityImpact": "HIGH",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "LOCAL",
                "availabilityImpact": "HIGH",
                "baseScore": 8.4,
                "baseSeverity": "HIGH",
                "confidentialityImpact": "HIGH",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-451",
                  "description": "CWE-451 User Interface (UI) Misrepresentation of Critical Information",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-03T20:11:15.447Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security/download/2019-01-22-draeger-infinity-delta-vf10-1-security-advisory.pdf"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/dr-ger-infinity-explorer-c700-privilege-escalation-via-kiosk-mode-bypass"
            }
          ],
          "source": {
            "discovery": "EXTERNAL"
          },
          "title": "Dr\u00e4ger Infinity Explorer C700 Privilege Escalation via Kiosk Mode Bypass",
          "x_generator": {
            "engine": "Vulnogram 1.0.2"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2019-25718",
        "datePublished": "2026-06-01T21:41:34.625Z",
        "dateReserved": "2026-06-01T21:36:41.544Z",
        "dateUpdated": "2026-06-03T20:11:15.447Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2018-19014 (GCVE-0-2018-19014)

    Vulnerability from cvelistv5 – Published: 2019-01-28 22:00 – Updated: 2024-09-16 20:31
    VLAI
    Summary
    Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.
    Severity
    No CVSS data available.
    CWE
    • CWE-532 - INFORMATION EXPOSURE THROUGH LOG FILES CWE-532
    Assigner
    References
    Impacted products
    Vendor Product Version
    ICS-CERT Dräger Infinity Delta Affected: Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions.
    Create a notification for this product.
    Date Public
    2019-01-22 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-05T11:23:08.949Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
              },
              {
                "name": "106683",
                "tags": [
                  "vdb-entry",
                  "x_refsource_BID",
                  "x_transferred"
                ],
                "url": "http://www.securityfocus.com/bid/106683"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Dr\u0026#195;\u0026#164;ger Infinity Delta",
              "vendor": "ICS-CERT",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions."
                }
              ]
            }
          ],
          "datePublic": "2019-01-22T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-532",
                  "description": "INFORMATION EXPOSURE THROUGH LOG FILES CWE-532",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-01-29T10:57:01.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
            },
            {
              "name": "106683",
              "tags": [
                "vdb-entry",
                "x_refsource_BID"
              ],
              "url": "http://www.securityfocus.com/bid/106683"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "DATE_PUBLIC": "2019-01-22T00:00:00",
              "ID": "CVE-2018-19014",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Dr\u0026#195;\u0026#164;ger Infinity Delta",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions."
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "ICS-CERT"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "INFORMATION EXPOSURE THROUGH LOG FILES CWE-532"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
                },
                {
                  "name": "106683",
                  "refsource": "BID",
                  "url": "http://www.securityfocus.com/bid/106683"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2018-19014",
        "datePublished": "2019-01-28T22:00:00.000Z",
        "dateReserved": "2018-11-06T00:00:00.000Z",
        "dateUpdated": "2024-09-16T20:31:32.329Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2018-19010 (GCVE-0-2018-19010)

    Vulnerability from cvelistv5 – Published: 2019-01-28 22:00 – Updated: 2024-09-16 17:32
    VLAI
    Summary
    Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.
    Severity
    No CVSS data available.
    CWE
    • CWE-20 - IMPROPER INPUT VALIDATION CWE-20
    Assigner
    References
    Impacted products
    Vendor Product Version
    ICS-CERT Dräger Infinity Delta Affected: Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions.
    Create a notification for this product.
    Date Public
    2019-01-22 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-05T11:23:09.031Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
              },
              {
                "name": "106683",
                "tags": [
                  "vdb-entry",
                  "x_refsource_BID",
                  "x_transferred"
                ],
                "url": "http://www.securityfocus.com/bid/106683"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Dr\u0026#195;\u0026#164;ger Infinity Delta",
              "vendor": "ICS-CERT",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions."
                }
              ]
            }
          ],
          "datePublic": "2019-01-22T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-20",
                  "description": "IMPROPER INPUT VALIDATION CWE-20",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-01-29T10:57:01.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
            },
            {
              "name": "106683",
              "tags": [
                "vdb-entry",
                "x_refsource_BID"
              ],
              "url": "http://www.securityfocus.com/bid/106683"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "DATE_PUBLIC": "2019-01-22T00:00:00",
              "ID": "CVE-2018-19010",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Dr\u0026#195;\u0026#164;ger Infinity Delta",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions."
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "ICS-CERT"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "IMPROPER INPUT VALIDATION CWE-20"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
                },
                {
                  "name": "106683",
                  "refsource": "BID",
                  "url": "http://www.securityfocus.com/bid/106683"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2018-19010",
        "datePublished": "2019-01-28T22:00:00.000Z",
        "dateReserved": "2018-11-06T00:00:00.000Z",
        "dateUpdated": "2024-09-16T17:32:40.689Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2018-19012 (GCVE-0-2018-19012)

    Vulnerability from cvelistv5 – Published: 2019-01-28 21:00 – Updated: 2024-09-17 04:08
    VLAI
    Summary
    Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Via a specific dialog it is possible to break out of the kiosk mode and reach the underlying operating system. By breaking out of the kiosk mode, an attacker is able to take control of the operating system.
    Severity
    No CVSS data available.
    CWE
    • CWE-269 - IMPROPER PRIVILEGE MANAGEMENT CWE-269
    Assigner
    References
    Impacted products
    Vendor Product Version
    ICS-CERT Dräger Infinity Delta Affected: Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions.
    Create a notification for this product.
    Date Public
    2018-01-22 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-05T11:23:08.579Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
              },
              {
                "name": "106683",
                "tags": [
                  "vdb-entry",
                  "x_refsource_BID",
                  "x_transferred"
                ],
                "url": "http://www.securityfocus.com/bid/106683"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Dr\u0026#195;\u0026#164;ger Infinity Delta",
              "vendor": "ICS-CERT",
              "versions": [
                {
                  "status": "affected",
                  "version": "Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions."
                }
              ]
            }
          ],
          "datePublic": "2018-01-22T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Via a specific dialog it is possible to break out of the kiosk mode and reach the underlying operating system. By breaking out of the kiosk mode, an attacker is able to take control of the operating system."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-269",
                  "description": "IMPROPER PRIVILEGE MANAGEMENT CWE-269",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2019-01-29T10:57:01.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
            },
            {
              "name": "106683",
              "tags": [
                "vdb-entry",
                "x_refsource_BID"
              ],
              "url": "http://www.securityfocus.com/bid/106683"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "DATE_PUBLIC": "2018-01-22T00:00:00",
              "ID": "CVE-2018-19012",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Dr\u0026#195;\u0026#164;ger Infinity Delta",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions."
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "ICS-CERT"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Via a specific dialog it is possible to break out of the kiosk mode and reach the underlying operating system. By breaking out of the kiosk mode, an attacker is able to take control of the operating system."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "IMPROPER PRIVILEGE MANAGEMENT CWE-269"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01",
                  "refsource": "MISC",
                  "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-022-01"
                },
                {
                  "name": "106683",
                  "refsource": "BID",
                  "url": "http://www.securityfocus.com/bid/106683"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2018-19012",
        "datePublished": "2019-01-28T21:00:00.000Z",
        "dateReserved": "2018-11-06T00:00:00.000Z",
        "dateUpdated": "2024-09-17T04:08:49.934Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }