Search

Find a vulnerability

Search criteria

    12 vulnerabilities found for em1200_firmware by baxter

    CVE-2020-12032 (GCVE-0-2020-12032)

    Vulnerability from nvd – Published: 2020-06-29 13:53 – Updated: 2024-08-04 11:48
    VLAI
    Summary
    Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. This could allow an attacker with network access to view or modify sensitive data including PHI.
    Severity
    No CVSS data available.
    CWE
    • CWE-311 - MISSING ENCRYPTION OF SENSITIVE DATA CWE-311
    Assigner
    References
    Impacted products
    Vendor Product Version
    n/a Baxter ExactaMix EM 2400 & EM 1200 Affected: ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-04T11:48:57.903Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
              "vendor": "n/a",
              "versions": [
                {
                  "status": "affected",
                  "version": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                }
              ]
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "value": "Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. This could allow an attacker with network access to view or modify sensitive data including PHI."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-311",
                  "description": "MISSING ENCRYPTION OF SENSITIVE DATA CWE-311",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2020-06-29T13:53:23.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2020-12032",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "n/a"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. This could allow an attacker with network access to view or modify sensitive data including PHI."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "MISSING ENCRYPTION OF SENSITIVE DATA CWE-311"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
                  "refsource": "MISC",
                  "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2020-12032",
        "datePublished": "2020-06-29T13:53:23.000Z",
        "dateReserved": "2020-04-21T00:00:00.000Z",
        "dateUpdated": "2024-08-04T11:48:57.903Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2020-12024 (GCVE-0-2020-12024)

    Vulnerability from nvd – Published: 2020-06-29 13:51 – Updated: 2024-08-04 11:48
    VLAI
    Summary
    Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access. Successful exploitation of this vulnerability may allow an attacker with physical access to the system the ability to load an unauthorized payload or unauthorized access to the hard drive by booting a live USB OS. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI.
    Severity
    No CVSS data available.
    CWE
    • CWE-284 - IMPROPER ACCESS CONTROL CWE-284
    Assigner
    References
    Impacted products
    Vendor Product Version
    n/a Baxter ExactaMix EM 2400 & EM 1200 Affected: ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-04T11:48:58.000Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
              "vendor": "n/a",
              "versions": [
                {
                  "status": "affected",
                  "version": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                }
              ]
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "value": "Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access. Successful exploitation of this vulnerability may allow an attacker with physical access to the system the ability to load an unauthorized payload or unauthorized access to the hard drive by booting a live USB OS. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-284",
                  "description": "IMPROPER ACCESS CONTROL CWE-284",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2020-06-29T13:51:31.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2020-12024",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "n/a"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access. Successful exploitation of this vulnerability may allow an attacker with physical access to the system the ability to load an unauthorized payload or unauthorized access to the hard drive by booting a live USB OS. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "IMPROPER ACCESS CONTROL CWE-284"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
                  "refsource": "MISC",
                  "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2020-12024",
        "datePublished": "2020-06-29T13:51:31.000Z",
        "dateReserved": "2020-04-21T00:00:00.000Z",
        "dateUpdated": "2024-08-04T11:48:58.000Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2020-12020 (GCVE-0-2020-12020)

    Vulnerability from nvd – Published: 2020-06-29 13:51 – Updated: 2024-08-04 11:48
    VLAI
    Summary
    Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application startup script. Successful exploitation of this vulnerability may allow an attacker to alter the startup script as the limited-access user.
    Severity
    No CVSS data available.
    CWE
    • CWE-668 - EXPOSURE OF RESOURCE TO WRONG SPHERE CWE-668
    Assigner
    References
    Impacted products
    Vendor Product Version
    n/a Baxter ExactaMix EM 2400 & EM 1200 Affected: ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-04T11:48:57.381Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
              "vendor": "n/a",
              "versions": [
                {
                  "status": "affected",
                  "version": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                }
              ]
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "value": "Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application startup script. Successful exploitation of this vulnerability may allow an attacker to alter the startup script as the limited-access user."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-668",
                  "description": "EXPOSURE OF RESOURCE TO WRONG SPHERE CWE-668",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2020-06-29T13:51:49.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2020-12020",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "n/a"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application startup script. Successful exploitation of this vulnerability may allow an attacker to alter the startup script as the limited-access user."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "EXPOSURE OF RESOURCE TO WRONG SPHERE CWE-668"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
                  "refsource": "MISC",
                  "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2020-12020",
        "datePublished": "2020-06-29T13:51:49.000Z",
        "dateReserved": "2020-04-21T00:00:00.000Z",
        "dateUpdated": "2024-08-04T11:48:57.381Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2020-12016 (GCVE-0-2020-12016)

    Vulnerability from nvd – Published: 2020-06-29 13:54 – Updated: 2024-08-04 11:48
    VLAI
    Summary
    Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. Successful exploitation of this vulnerability may allow an attacker who has gained unauthorized access to system resources, including access to execute software or to view/update files, directories, or system configuration. This could allow an attacker with network access to view sensitive data including PHI.
    Severity
    No CVSS data available.
    CWE
    • CWE-259 - USE OF HARD-CODED PASSWORD CWE-259
    Assigner
    References
    Impacted products
    Vendor Product Version
    n/a Baxter ExactaMix EM 2400 & EM 1200 Affected: ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-04T11:48:57.993Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
              "vendor": "n/a",
              "versions": [
                {
                  "status": "affected",
                  "version": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                }
              ]
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "value": "Baxter ExactaMix EM 2400 \u0026 EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. Successful exploitation of this vulnerability may allow an attacker who has gained unauthorized access to system resources, including access to execute software or to view/update files, directories, or system configuration. This could allow an attacker with network access to view sensitive data including PHI."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-259",
                  "description": "USE OF HARD-CODED PASSWORD CWE-259",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2020-06-29T13:54:50.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2020-12016",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "n/a"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Baxter ExactaMix EM 2400 \u0026 EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. Successful exploitation of this vulnerability may allow an attacker who has gained unauthorized access to system resources, including access to execute software or to view/update files, directories, or system configuration. This could allow an attacker with network access to view sensitive data including PHI."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "USE OF HARD-CODED PASSWORD CWE-259"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
                  "refsource": "MISC",
                  "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2020-12016",
        "datePublished": "2020-06-29T13:54:50.000Z",
        "dateReserved": "2020-04-21T00:00:00.000Z",
        "dateUpdated": "2024-08-04T11:48:57.993Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2020-12012 (GCVE-0-2020-12012)

    Vulnerability from nvd – Published: 2020-06-29 13:54 – Updated: 2024-08-04 11:48
    VLAI
    Summary
    Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. Successful exploitation of this vulnerability may allow an attacker with physical access to gain unauthorized access to view/update system configuration or data. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI.
    Severity
    No CVSS data available.
    CWE
    • CWE-259 - USE OF HARD-CODED PASSWORD CWE-259
    Assigner
    References
    Impacted products
    Vendor Product Version
    n/a Baxter ExactaMix EM 2400 & EM 1200 Affected: ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-04T11:48:57.903Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
              "vendor": "n/a",
              "versions": [
                {
                  "status": "affected",
                  "version": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                }
              ]
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "value": "Baxter ExactaMix EM 2400 \u0026 EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. Successful exploitation of this vulnerability may allow an attacker with physical access to gain unauthorized access to view/update system configuration or data. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-259",
                  "description": "USE OF HARD-CODED PASSWORD CWE-259",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2020-06-29T13:54:53.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2020-12012",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "n/a"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Baxter ExactaMix EM 2400 \u0026 EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. Successful exploitation of this vulnerability may allow an attacker with physical access to gain unauthorized access to view/update system configuration or data. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "USE OF HARD-CODED PASSWORD CWE-259"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
                  "refsource": "MISC",
                  "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2020-12012",
        "datePublished": "2020-06-29T13:54:53.000Z",
        "dateReserved": "2020-04-21T00:00:00.000Z",
        "dateUpdated": "2024-08-04T11:48:57.903Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2020-12008 (GCVE-0-2020-12008)

    Vulnerability from nvd – Published: 2020-06-29 13:53 – Updated: 2024-08-04 11:48
    VLAI
    Summary
    Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system. This could allow an attacker with network access to view sensitive data including PHI.
    Severity
    No CVSS data available.
    CWE
    • CWE-319 - CLEARTEXT TRANSMISSION OF SENSITIVE DATA CWE-319
    Assigner
    References
    Impacted products
    Vendor Product Version
    n/a Baxter ExactaMix EM 2400 & EM 1200 Affected: ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-04T11:48:57.121Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
              "vendor": "n/a",
              "versions": [
                {
                  "status": "affected",
                  "version": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                }
              ]
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "value": "Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system. This could allow an attacker with network access to view sensitive data including PHI."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-319",
                  "description": "CLEARTEXT TRANSMISSION OF SENSITIVE DATA CWE-319",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2020-06-29T13:53:25.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2020-12008",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "n/a"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system. This could allow an attacker with network access to view sensitive data including PHI."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "CLEARTEXT TRANSMISSION OF SENSITIVE DATA CWE-319"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
                  "refsource": "MISC",
                  "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2020-12008",
        "datePublished": "2020-06-29T13:53:25.000Z",
        "dateReserved": "2020-04-21T00:00:00.000Z",
        "dateUpdated": "2024-08-04T11:48:57.121Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2020-12012 (GCVE-0-2020-12012)

    Vulnerability from cvelistv5 – Published: 2020-06-29 13:54 – Updated: 2024-08-04 11:48
    VLAI
    Summary
    Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. Successful exploitation of this vulnerability may allow an attacker with physical access to gain unauthorized access to view/update system configuration or data. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI.
    Severity
    No CVSS data available.
    CWE
    • CWE-259 - USE OF HARD-CODED PASSWORD CWE-259
    Assigner
    References
    Impacted products
    Vendor Product Version
    n/a Baxter ExactaMix EM 2400 & EM 1200 Affected: ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-04T11:48:57.903Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
              "vendor": "n/a",
              "versions": [
                {
                  "status": "affected",
                  "version": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                }
              ]
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "value": "Baxter ExactaMix EM 2400 \u0026 EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. Successful exploitation of this vulnerability may allow an attacker with physical access to gain unauthorized access to view/update system configuration or data. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-259",
                  "description": "USE OF HARD-CODED PASSWORD CWE-259",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2020-06-29T13:54:53.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2020-12012",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "n/a"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Baxter ExactaMix EM 2400 \u0026 EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. Successful exploitation of this vulnerability may allow an attacker with physical access to gain unauthorized access to view/update system configuration or data. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "USE OF HARD-CODED PASSWORD CWE-259"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
                  "refsource": "MISC",
                  "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2020-12012",
        "datePublished": "2020-06-29T13:54:53.000Z",
        "dateReserved": "2020-04-21T00:00:00.000Z",
        "dateUpdated": "2024-08-04T11:48:57.903Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2020-12016 (GCVE-0-2020-12016)

    Vulnerability from cvelistv5 – Published: 2020-06-29 13:54 – Updated: 2024-08-04 11:48
    VLAI
    Summary
    Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. Successful exploitation of this vulnerability may allow an attacker who has gained unauthorized access to system resources, including access to execute software or to view/update files, directories, or system configuration. This could allow an attacker with network access to view sensitive data including PHI.
    Severity
    No CVSS data available.
    CWE
    • CWE-259 - USE OF HARD-CODED PASSWORD CWE-259
    Assigner
    References
    Impacted products
    Vendor Product Version
    n/a Baxter ExactaMix EM 2400 & EM 1200 Affected: ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-04T11:48:57.993Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
              "vendor": "n/a",
              "versions": [
                {
                  "status": "affected",
                  "version": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                }
              ]
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "value": "Baxter ExactaMix EM 2400 \u0026 EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. Successful exploitation of this vulnerability may allow an attacker who has gained unauthorized access to system resources, including access to execute software or to view/update files, directories, or system configuration. This could allow an attacker with network access to view sensitive data including PHI."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-259",
                  "description": "USE OF HARD-CODED PASSWORD CWE-259",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2020-06-29T13:54:50.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2020-12016",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "n/a"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Baxter ExactaMix EM 2400 \u0026 EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. Successful exploitation of this vulnerability may allow an attacker who has gained unauthorized access to system resources, including access to execute software or to view/update files, directories, or system configuration. This could allow an attacker with network access to view sensitive data including PHI."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "USE OF HARD-CODED PASSWORD CWE-259"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
                  "refsource": "MISC",
                  "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2020-12016",
        "datePublished": "2020-06-29T13:54:50.000Z",
        "dateReserved": "2020-04-21T00:00:00.000Z",
        "dateUpdated": "2024-08-04T11:48:57.993Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2020-12008 (GCVE-0-2020-12008)

    Vulnerability from cvelistv5 – Published: 2020-06-29 13:53 – Updated: 2024-08-04 11:48
    VLAI
    Summary
    Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system. This could allow an attacker with network access to view sensitive data including PHI.
    Severity
    No CVSS data available.
    CWE
    • CWE-319 - CLEARTEXT TRANSMISSION OF SENSITIVE DATA CWE-319
    Assigner
    References
    Impacted products
    Vendor Product Version
    n/a Baxter ExactaMix EM 2400 & EM 1200 Affected: ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-04T11:48:57.121Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
              "vendor": "n/a",
              "versions": [
                {
                  "status": "affected",
                  "version": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                }
              ]
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "value": "Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system. This could allow an attacker with network access to view sensitive data including PHI."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-319",
                  "description": "CLEARTEXT TRANSMISSION OF SENSITIVE DATA CWE-319",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2020-06-29T13:53:25.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2020-12008",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "n/a"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system. This could allow an attacker with network access to view sensitive data including PHI."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "CLEARTEXT TRANSMISSION OF SENSITIVE DATA CWE-319"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
                  "refsource": "MISC",
                  "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2020-12008",
        "datePublished": "2020-06-29T13:53:25.000Z",
        "dateReserved": "2020-04-21T00:00:00.000Z",
        "dateUpdated": "2024-08-04T11:48:57.121Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2020-12032 (GCVE-0-2020-12032)

    Vulnerability from cvelistv5 – Published: 2020-06-29 13:53 – Updated: 2024-08-04 11:48
    VLAI
    Summary
    Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. This could allow an attacker with network access to view or modify sensitive data including PHI.
    Severity
    No CVSS data available.
    CWE
    • CWE-311 - MISSING ENCRYPTION OF SENSITIVE DATA CWE-311
    Assigner
    References
    Impacted products
    Vendor Product Version
    n/a Baxter ExactaMix EM 2400 & EM 1200 Affected: ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-04T11:48:57.903Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
              "vendor": "n/a",
              "versions": [
                {
                  "status": "affected",
                  "version": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                }
              ]
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "value": "Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. This could allow an attacker with network access to view or modify sensitive data including PHI."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-311",
                  "description": "MISSING ENCRYPTION OF SENSITIVE DATA CWE-311",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2020-06-29T13:53:23.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2020-12032",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "n/a"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. This could allow an attacker with network access to view or modify sensitive data including PHI."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "MISSING ENCRYPTION OF SENSITIVE DATA CWE-311"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
                  "refsource": "MISC",
                  "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2020-12032",
        "datePublished": "2020-06-29T13:53:23.000Z",
        "dateReserved": "2020-04-21T00:00:00.000Z",
        "dateUpdated": "2024-08-04T11:48:57.903Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2020-12020 (GCVE-0-2020-12020)

    Vulnerability from cvelistv5 – Published: 2020-06-29 13:51 – Updated: 2024-08-04 11:48
    VLAI
    Summary
    Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application startup script. Successful exploitation of this vulnerability may allow an attacker to alter the startup script as the limited-access user.
    Severity
    No CVSS data available.
    CWE
    • CWE-668 - EXPOSURE OF RESOURCE TO WRONG SPHERE CWE-668
    Assigner
    References
    Impacted products
    Vendor Product Version
    n/a Baxter ExactaMix EM 2400 & EM 1200 Affected: ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-04T11:48:57.381Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
              "vendor": "n/a",
              "versions": [
                {
                  "status": "affected",
                  "version": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                }
              ]
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "value": "Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application startup script. Successful exploitation of this vulnerability may allow an attacker to alter the startup script as the limited-access user."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-668",
                  "description": "EXPOSURE OF RESOURCE TO WRONG SPHERE CWE-668",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2020-06-29T13:51:49.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2020-12020",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "n/a"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application startup script. Successful exploitation of this vulnerability may allow an attacker to alter the startup script as the limited-access user."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "EXPOSURE OF RESOURCE TO WRONG SPHERE CWE-668"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
                  "refsource": "MISC",
                  "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2020-12020",
        "datePublished": "2020-06-29T13:51:49.000Z",
        "dateReserved": "2020-04-21T00:00:00.000Z",
        "dateUpdated": "2024-08-04T11:48:57.381Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2020-12024 (GCVE-0-2020-12024)

    Vulnerability from cvelistv5 – Published: 2020-06-29 13:51 – Updated: 2024-08-04 11:48
    VLAI
    Summary
    Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access. Successful exploitation of this vulnerability may allow an attacker with physical access to the system the ability to load an unauthorized payload or unauthorized access to the hard drive by booting a live USB OS. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI.
    Severity
    No CVSS data available.
    CWE
    • CWE-284 - IMPROPER ACCESS CONTROL CWE-284
    Assigner
    References
    Impacted products
    Vendor Product Version
    n/a Baxter ExactaMix EM 2400 & EM 1200 Affected: ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "providerMetadata": {
              "dateUpdated": "2024-08-04T11:48:58.000Z",
              "orgId": "af854a3a-2127-422b-91ae-364da2661108",
              "shortName": "CVE"
            },
            "references": [
              {
                "tags": [
                  "x_refsource_MISC",
                  "x_transferred"
                ],
                "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
              }
            ],
            "title": "CVE Program Container"
          }
        ],
        "cna": {
          "affected": [
            {
              "product": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
              "vendor": "n/a",
              "versions": [
                {
                  "status": "affected",
                  "version": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                }
              ]
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "value": "Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access. Successful exploitation of this vulnerability may allow an attacker with physical access to the system the ability to load an unauthorized payload or unauthorized access to the hard drive by booting a live USB OS. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI."
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-284",
                  "description": "IMPROPER ACCESS CONTROL CWE-284",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2020-06-29T13:51:31.000Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "tags": [
                "x_refsource_MISC"
              ],
              "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
            }
          ],
          "x_legacyV4Record": {
            "CVE_data_meta": {
              "ASSIGNER": "ics-cert@hq.dhs.gov",
              "ID": "CVE-2020-12024",
              "STATE": "PUBLIC"
            },
            "affects": {
              "vendor": {
                "vendor_data": [
                  {
                    "product": {
                      "product_data": [
                        {
                          "product_name": "Baxter ExactaMix EM 2400 \u0026 EM 1200",
                          "version": {
                            "version_data": [
                              {
                                "version_value": "ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5"
                              }
                            ]
                          }
                        }
                      ]
                    },
                    "vendor_name": "n/a"
                  }
                ]
              }
            },
            "data_format": "MITRE",
            "data_type": "CVE",
            "data_version": "4.0",
            "description": {
              "description_data": [
                {
                  "lang": "eng",
                  "value": "Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access. Successful exploitation of this vulnerability may allow an attacker with physical access to the system the ability to load an unauthorized payload or unauthorized access to the hard drive by booting a live USB OS. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI."
                }
              ]
            },
            "problemtype": {
              "problemtype_data": [
                {
                  "description": [
                    {
                      "lang": "eng",
                      "value": "IMPROPER ACCESS CONTROL CWE-284"
                    }
                  ]
                }
              ]
            },
            "references": {
              "reference_data": [
                {
                  "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
                  "refsource": "MISC",
                  "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01"
                }
              ]
            }
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2020-12024",
        "datePublished": "2020-06-29T13:51:31.000Z",
        "dateReserved": "2020-04-21T00:00:00.000Z",
        "dateUpdated": "2024-08-04T11:48:58.000Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }