    CVE-2025-24836 (GCVE-0-2025-24836)

    Vulnerability from nvd – Published: 2025-02-13 21:55 – Updated: 2025-02-14 18:11
    VLAI
    Title
    Qardio Heart Health IOS and Android Application and QardioARM A100 Uncaught Exception
    Summary
    With a specially crafted Python script, an attacker could send continuous startMeasurement commands over an unencrypted Bluetooth connection to the affected device. This would prevent the device from connecting to a clinician's app to take patient readings and ostensibly flood it with requests, resulting in a denial-of-service condition.
    SSVC
    Exploitation: none Automatable: no Technical Impact: total
    CISA Coordinator (v2.0.3)
    CWE
    Assigner
    Credits
    Bryan Riggins of Insulet Corporation reported these vulnerabilities to CISA.
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2025-24836",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "total"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2025-02-14T18:11:06.923716Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2025-02-14T18:11:19.931Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "Heart Health IOS Mobile Application",
              "vendor": "Qardio",
              "versions": [
                {
                  "status": "affected",
                  "version": "2.7.4"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "Heart Health Android Mobile Application",
              "vendor": "Qardio",
              "versions": [
                {
                  "status": "affected",
                  "version": "2.5.1"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "QardioARM",
              "vendor": "Qardio",
              "versions": [
                {
                  "status": "affected",
                  "version": "All versions"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Bryan Riggins of Insulet Corporation reported these vulnerabilities to CISA."
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "With a specially crafted Python script, an attacker could send \ncontinuous startMeasurement commands over an unencrypted Bluetooth \nconnection to the affected device. This would prevent the device from \nconnecting to a clinician\u0027s app to take patient readings and ostensibly \nflood it with requests, resulting in a denial-of-service condition."
                }
              ],
              "value": "With a specially crafted Python script, an attacker could send \ncontinuous startMeasurement commands over an unencrypted Bluetooth \nconnection to the affected device. This would prevent the device from \nconnecting to a clinician\u0027s app to take patient readings and ostensibly \nflood it with requests, resulting in a denial-of-service condition."
            }
          ],
          "metrics": [
            {
              "cvssV3_1": {
                "attackComplexity": "HIGH",
                "attackVector": "ADJACENT_NETWORK",
                "availabilityImpact": "HIGH",
                "baseScore": 7.1,
                "baseSeverity": "HIGH",
                "confidentialityImpact": "LOW",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "HIGH",
                "attackRequirements": "NONE",
                "attackVector": "ADJACENT",
                "baseScore": 6.1,
                "baseSeverity": "MEDIUM",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:A/AC:H/AT:N/PR:N/UI:N/VC:L/VI:H/VA:H/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "HIGH",
                "vulnConfidentialityImpact": "LOW",
                "vulnIntegrityImpact": "HIGH",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-248",
                  "description": "CWE-248",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2025-02-13T21:55:12.382Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-044-01"
            },
            {
              "url": "https://www.qardio.com/about-us/#contact"
            }
          ],
          "source": {
            "advisory": "ICSMA-25-044-01",
            "discovery": "EXTERNAL"
          },
          "title": "Qardio Heart Health IOS and Android Application and QardioARM A100 Uncaught Exception",
          "workarounds": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "Qardio has not responded to requests to work with CISA to mitigate these\n vulnerabilities. Users of these affected products are invited to \ncontact \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.qardio.com/about-us/#contact\"\u003eQardio customer support\u003c/a\u003e for additional information.\n\u003cp\u003eUsers should do the following to help mitigate the risk:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003eDisable Bluetooth when not in use.\u003c/li\u003e\n\u003cli\u003eDon\u0027t use this device in public or within Bluetooth range of malicious actors.\u003c/li\u003e\n\u003cli\u003eOnly use trusted mobile apps from trusted providers.\u003c/li\u003e\n\u003c/ul\u003e\n\n\u003cbr\u003e"
                }
              ],
              "value": "Qardio has not responded to requests to work with CISA to mitigate these\n vulnerabilities. Users of these affected products are invited to \ncontact  Qardio customer support https://www.qardio.com/about-us/#contact  for additional information.\nUsers should do the following to help mitigate the risk:\n\n\n\n  *  Disable Bluetooth when not in use.\n\n  *  Don\u0027t use this device in public or within Bluetooth range of malicious actors.\n\n  *  Only use trusted mobile apps from trusted providers."
            }
          ],
          "x_generator": {
            "engine": "Vulnogram 0.2.0"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2025-24836",
        "datePublished": "2025-02-13T21:55:12.382Z",
        "dateReserved": "2025-02-10T15:16:25.257Z",
        "dateUpdated": "2025-02-14T18:11:19.931Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2025-24836 (GCVE-0-2025-24836)

    Vulnerability from cvelistv5 – Published: 2025-02-13 21:55 – Updated: 2025-02-14 18:11
    VLAI
    Title
    Qardio Heart Health IOS and Android Application and QardioARM A100 Uncaught Exception
    Summary
    With a specially crafted Python script, an attacker could send continuous startMeasurement commands over an unencrypted Bluetooth connection to the affected device. This would prevent the device from connecting to a clinician's app to take patient readings and ostensibly flood it with requests, resulting in a denial-of-service condition.
    SSVC
    Exploitation: none Automatable: no Technical Impact: total
    CISA Coordinator (v2.0.3)
    CWE
    Assigner
    Credits
    Bryan Riggins of Insulet Corporation reported these vulnerabilities to CISA.
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2025-24836",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "total"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2025-02-14T18:11:06.923716Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2025-02-14T18:11:19.931Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "Heart Health IOS Mobile Application",
              "vendor": "Qardio",
              "versions": [
                {
                  "status": "affected",
                  "version": "2.7.4"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "Heart Health Android Mobile Application",
              "vendor": "Qardio",
              "versions": [
                {
                  "status": "affected",
                  "version": "2.5.1"
                }
              ]
            },
            {
              "defaultStatus": "unaffected",
              "product": "QardioARM",
              "vendor": "Qardio",
              "versions": [
                {
                  "status": "affected",
                  "version": "All versions"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Bryan Riggins of Insulet Corporation reported these vulnerabilities to CISA."
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "With a specially crafted Python script, an attacker could send \ncontinuous startMeasurement commands over an unencrypted Bluetooth \nconnection to the affected device. This would prevent the device from \nconnecting to a clinician\u0027s app to take patient readings and ostensibly \nflood it with requests, resulting in a denial-of-service condition."
                }
              ],
              "value": "With a specially crafted Python script, an attacker could send \ncontinuous startMeasurement commands over an unencrypted Bluetooth \nconnection to the affected device. This would prevent the device from \nconnecting to a clinician\u0027s app to take patient readings and ostensibly \nflood it with requests, resulting in a denial-of-service condition."
            }
          ],
          "metrics": [
            {
              "cvssV3_1": {
                "attackComplexity": "HIGH",
                "attackVector": "ADJACENT_NETWORK",
                "availabilityImpact": "HIGH",
                "baseScore": 7.1,
                "baseSeverity": "HIGH",
                "confidentialityImpact": "LOW",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "HIGH",
                "attackRequirements": "NONE",
                "attackVector": "ADJACENT",
                "baseScore": 6.1,
                "baseSeverity": "MEDIUM",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:A/AC:H/AT:N/PR:N/UI:N/VC:L/VI:H/VA:H/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "HIGH",
                "vulnConfidentialityImpact": "LOW",
                "vulnIntegrityImpact": "HIGH",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-248",
                  "description": "CWE-248",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2025-02-13T21:55:12.382Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-044-01"
            },
            {
              "url": "https://www.qardio.com/about-us/#contact"
            }
          ],
          "source": {
            "advisory": "ICSMA-25-044-01",
            "discovery": "EXTERNAL"
          },
          "title": "Qardio Heart Health IOS and Android Application and QardioARM A100 Uncaught Exception",
          "workarounds": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "Qardio has not responded to requests to work with CISA to mitigate these\n vulnerabilities. Users of these affected products are invited to \ncontact \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.qardio.com/about-us/#contact\"\u003eQardio customer support\u003c/a\u003e for additional information.\n\u003cp\u003eUsers should do the following to help mitigate the risk:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003eDisable Bluetooth when not in use.\u003c/li\u003e\n\u003cli\u003eDon\u0027t use this device in public or within Bluetooth range of malicious actors.\u003c/li\u003e\n\u003cli\u003eOnly use trusted mobile apps from trusted providers.\u003c/li\u003e\n\u003c/ul\u003e\n\n\u003cbr\u003e"
                }
              ],
              "value": "Qardio has not responded to requests to work with CISA to mitigate these\n vulnerabilities. Users of these affected products are invited to \ncontact  Qardio customer support https://www.qardio.com/about-us/#contact  for additional information.\nUsers should do the following to help mitigate the risk:\n\n\n\n  *  Disable Bluetooth when not in use.\n\n  *  Don\u0027t use this device in public or within Bluetooth range of malicious actors.\n\n  *  Only use trusted mobile apps from trusted providers."
            }
          ],
          "x_generator": {
            "engine": "Vulnogram 0.2.0"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2025-24836",
        "datePublished": "2025-02-13T21:55:12.382Z",
        "dateReserved": "2025-02-10T15:16:25.257Z",
        "dateUpdated": "2025-02-14T18:11:19.931Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }