Search

Find a vulnerability

Search criteria

    2 vulnerabilities found for Perseus A500 by Dräger

    CVE-2019-25723 (GCVE-0-2019-25723)

    Vulnerability from nvd – Published: 2026-06-02 19:00 – Updated: 2026-06-03 18:01
    VLAI
    Title
    Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface
    Summary
    Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-1286 - Improper Validation of Syntactic Correctness of Input
    Assigner
    References
    Impacted products
    Vendor Product Version
    Dräger Perseus A500 Affected: 2.00 (custom)
    Affected: 2.01 (custom)
    Affected: 2.02 (custom)
    Unaffected: 2.03 (custom)
    Create a notification for this product.
    Date Public
    2020-06-17 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2019-25723",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-03T13:56:22.966552Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-03T13:56:45.789Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unknown",
              "product": "Perseus A500",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "2.00",
                  "versionType": "custom"
                },
                {
                  "status": "affected",
                  "version": "2.01",
                  "versionType": "custom"
                },
                {
                  "status": "affected",
                  "version": "2.02",
                  "versionType": "custom"
                },
                {
                  "status": "unaffected",
                  "version": "2.03",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "datePublic": "2020-06-17T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eDr\u00e4ger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.\u003c/p\u003e"
                }
              ],
              "value": "Dr\u00e4ger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "HIGH",
                "attackRequirements": "NONE",
                "attackVector": "NETWORK",
                "baseScore": 6.3,
                "baseSeverity": "MEDIUM",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "LOW",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L",
                "version": "4.0",
                "vulnAvailabilityImpact": "LOW",
                "vulnConfidentialityImpact": "NONE",
                "vulnIntegrityImpact": "NONE",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "HIGH",
                "attackVector": "NETWORK",
                "availabilityImpact": "LOW",
                "baseScore": 4,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "NONE",
                "privilegesRequired": "NONE",
                "scope": "CHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-1286",
                  "description": "CWE-1286 Improper Validation of Syntactic Correctness of Input",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-03T18:01:04.808Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security/download/2020-06-PSA-20-169-01-Security-Advisory-Perseus-A500-SW20n.pdf"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/dr-ger-perseus-a500-dos-via-medibus-interface"
            }
          ],
          "source": {
            "discovery": "EXTERNAL"
          },
          "title": "Dr\u00e4ger Perseus A500 2.00-2.02 DoS via Medibus Interface"
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2019-25723",
        "datePublished": "2026-06-02T19:00:19.249Z",
        "dateReserved": "2026-06-02T18:56:43.003Z",
        "dateUpdated": "2026-06-03T18:01:04.808Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2019-25723 (GCVE-0-2019-25723)

    Vulnerability from cvelistv5 – Published: 2026-06-02 19:00 – Updated: 2026-06-03 18:01
    VLAI
    Title
    Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface
    Summary
    Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-1286 - Improper Validation of Syntactic Correctness of Input
    Assigner
    References
    Impacted products
    Vendor Product Version
    Dräger Perseus A500 Affected: 2.00 (custom)
    Affected: 2.01 (custom)
    Affected: 2.02 (custom)
    Unaffected: 2.03 (custom)
    Create a notification for this product.
    Date Public
    2020-06-17 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2019-25723",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-03T13:56:22.966552Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-03T13:56:45.789Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unknown",
              "product": "Perseus A500",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "status": "affected",
                  "version": "2.00",
                  "versionType": "custom"
                },
                {
                  "status": "affected",
                  "version": "2.01",
                  "versionType": "custom"
                },
                {
                  "status": "affected",
                  "version": "2.02",
                  "versionType": "custom"
                },
                {
                  "status": "unaffected",
                  "version": "2.03",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "datePublic": "2020-06-17T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eDr\u00e4ger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.\u003c/p\u003e"
                }
              ],
              "value": "Dr\u00e4ger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "HIGH",
                "attackRequirements": "NONE",
                "attackVector": "NETWORK",
                "baseScore": 6.3,
                "baseSeverity": "MEDIUM",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "LOW",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L",
                "version": "4.0",
                "vulnAvailabilityImpact": "LOW",
                "vulnConfidentialityImpact": "NONE",
                "vulnIntegrityImpact": "NONE",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "HIGH",
                "attackVector": "NETWORK",
                "availabilityImpact": "LOW",
                "baseScore": 4,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "NONE",
                "privilegesRequired": "NONE",
                "scope": "CHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-1286",
                  "description": "CWE-1286 Improper Validation of Syntactic Correctness of Input",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-03T18:01:04.808Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security/download/2020-06-PSA-20-169-01-Security-Advisory-Perseus-A500-SW20n.pdf"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/dr-ger-perseus-a500-dos-via-medibus-interface"
            }
          ],
          "source": {
            "discovery": "EXTERNAL"
          },
          "title": "Dr\u00e4ger Perseus A500 2.00-2.02 DoS via Medibus Interface"
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2019-25723",
        "datePublished": "2026-06-02T19:00:19.249Z",
        "dateReserved": "2026-06-02T18:56:43.003Z",
        "dateUpdated": "2026-06-03T18:01:04.808Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }