Search

Find a vulnerability

Search criteria

    4 vulnerabilities found for Intellispace Cardiovascular (ISCV) by Philips

    CVE-2025-2230 (GCVE-0-2025-2230)

    Vulnerability from nvd – Published: 2025-03-13 18:14 – Updated: 2025-03-17 20:05
    VLAI
    Title
    Philips Intellispace Cardiovascular (ISCV) Improper Authentication
    Summary
    A flaw exists in the Windows login flow where an AuthContext token can be exploited for replay attacks and authentication bypass.
    SSVC
    Exploitation: none Automatable: no Technical Impact: total
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-287 - Improper Authentication
    Assigner
    Impacted products
    Vendor Product Version
    Philips Intellispace Cardiovascular (ISCV) Affected: 0 , ≤ 5.1 (custom)
    Create a notification for this product.
    Credits
    Joe Dillon reported these vulnerabilities to Philips.
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2025-2230",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "total"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2025-03-13T19:33:45.514056Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2025-03-13T19:34:02.967Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "Intellispace Cardiovascular (ISCV)",
              "vendor": "Philips",
              "versions": [
                {
                  "lessThanOrEqual": "5.1",
                  "status": "affected",
                  "version": "0",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Joe Dillon reported these vulnerabilities to Philips."
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "A flaw exists in the Windows login flow where an AuthContext token can \nbe exploited for replay attacks and authentication bypass."
                }
              ],
              "value": "A flaw exists in the Windows login flow where an AuthContext token can \nbe exploited for replay attacks and authentication bypass."
            }
          ],
          "metrics": [
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "LOCAL",
                "availabilityImpact": "NONE",
                "baseScore": 7.7,
                "baseSeverity": "HIGH",
                "confidentialityImpact": "HIGH",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "LOCAL",
                "baseScore": 8.5,
                "baseSeverity": "HIGH",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "NONE",
                "vulnConfidentialityImpact": "HIGH",
                "vulnIntegrityImpact": "HIGH",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-287",
                  "description": "CWE-287 Improper Authentication",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2025-03-17T20:05:56.939Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-072-01"
            },
            {
              "url": "https://www.philips.com/a-w/security/security-advisories.html"
            }
          ],
          "solutions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003ePhilips recommends the following mitigations:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eResolved in ISCV 5.2, which was released in September 2020.\u003c/span\u003e\n\n\u003c/li\u003e\n\n\u003cli\u003ePhilips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\u003c/li\u003e\n\u003cli\u003ePlease contact a local Philips sales (service) representative to learn how to engage this upgrade process.\u003c/li\u003e\n\u003cli\u003eFor managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\u003c/li\u003e\n\u003c/ul\u003e\n\u003cp\u003eRefer to the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.philips.com/a-w/security/security-advisories.html\"\u003ePhilips advisory\u003c/a\u003efor more details.\n\n\u003cbr\u003e\u003c/p\u003e"
                }
              ],
              "value": "Philips recommends the following mitigations:\n\n\n\n  *  \n\nResolved in ISCV 5.2, which was released in September 2020.\n\n\n\n\n  *  Philips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\n\n  *  Please contact a local Philips sales (service) representative to learn how to engage this upgrade process.\n\n  *  For managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\n\n\n\n\nRefer to the  Philips advisory https://www.philips.com/a-w/security/security-advisories.html for more details."
            }
          ],
          "source": {
            "advisory": "ICSMA-25-072-01",
            "discovery": "EXTERNAL"
          },
          "title": "Philips Intellispace Cardiovascular (ISCV) Improper Authentication",
          "x_generator": {
            "engine": "Vulnogram 0.2.0"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2025-2230",
        "datePublished": "2025-03-13T18:14:43.573Z",
        "dateReserved": "2025-03-11T20:14:50.017Z",
        "dateUpdated": "2025-03-17T20:05:56.939Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2025-2229 (GCVE-0-2025-2229)

    Vulnerability from nvd – Published: 2025-03-13 18:17 – Updated: 2025-03-17 20:04
    VLAI
    Title
    Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials
    Summary
    A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations.
    SSVC
    Exploitation: none Automatable: no Technical Impact: total
    CISA Coordinator (v2.0.3)
    CWE
    Assigner
    Impacted products
    Vendor Product Version
    Philips Intellispace Cardiovascular (ISCV) Affected: 0 , ≤ 4.1 (custom)
    Create a notification for this product.
    Credits
    Joe Dillon reported these vulnerabilities to Philips.
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2025-2229",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "total"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2025-03-13T19:30:31.475698Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2025-03-13T19:30:42.059Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "Intellispace Cardiovascular (ISCV)",
              "vendor": "Philips",
              "versions": [
                {
                  "lessThanOrEqual": "4.1",
                  "status": "affected",
                  "version": "0",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Joe Dillon reported these vulnerabilities to Philips."
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "A token is created using the username, current date/time, and a fixed \nAES-128 encryption key, which is the same across all installations."
                }
              ],
              "value": "A token is created using the username, current date/time, and a fixed \nAES-128 encryption key, which is the same across all installations."
            }
          ],
          "metrics": [
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "LOCAL",
                "availabilityImpact": "NONE",
                "baseScore": 7.7,
                "baseSeverity": "HIGH",
                "confidentialityImpact": "HIGH",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "LOCAL",
                "baseScore": 8.5,
                "baseSeverity": "HIGH",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "NONE",
                "vulnConfidentialityImpact": "HIGH",
                "vulnIntegrityImpact": "HIGH",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-1391",
                  "description": "CWE-1391",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2025-03-17T20:04:56.100Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-072-01"
            },
            {
              "url": "https://www.philips.com/a-w/security/security-advisories.html"
            }
          ],
          "solutions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003ePhilips recommends the following mitigations:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eResolved in ISCV 4.2 build 20589, which was released in May 2019.\u003c/span\u003e\n\n\u003c/li\u003e\n\n\u003cli\u003ePhilips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\u003c/li\u003e\n\u003cli\u003ePlease contact a local Philips sales (service) representative to learn how to engage this upgrade process.\u003c/li\u003e\n\u003cli\u003eFor managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\u003c/li\u003e\n\u003c/ul\u003e\n\u003cp\u003eRefer to the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.philips.com/a-w/security/security-advisories.html\"\u003ePhilips advisory\u003c/a\u003efor more details.\n\n\u003cbr\u003e\u003c/p\u003e"
                }
              ],
              "value": "Philips recommends the following mitigations:\n\n\n\n  *  \n\nResolved in ISCV 4.2 build 20589, which was released in May 2019.\n\n\n\n\n  *  Philips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\n\n  *  Please contact a local Philips sales (service) representative to learn how to engage this upgrade process.\n\n  *  For managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\n\n\n\n\nRefer to the  Philips advisory https://www.philips.com/a-w/security/security-advisories.html for more details."
            }
          ],
          "source": {
            "advisory": "ICSMA-25-072-01",
            "discovery": "EXTERNAL"
          },
          "title": "Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials",
          "x_generator": {
            "engine": "Vulnogram 0.2.0"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2025-2229",
        "datePublished": "2025-03-13T18:17:31.256Z",
        "dateReserved": "2025-03-11T20:11:49.638Z",
        "dateUpdated": "2025-03-17T20:04:56.100Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2025-2229 (GCVE-0-2025-2229)

    Vulnerability from cvelistv5 – Published: 2025-03-13 18:17 – Updated: 2025-03-17 20:04
    VLAI
    Title
    Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials
    Summary
    A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations.
    SSVC
    Exploitation: none Automatable: no Technical Impact: total
    CISA Coordinator (v2.0.3)
    CWE
    Assigner
    Impacted products
    Vendor Product Version
    Philips Intellispace Cardiovascular (ISCV) Affected: 0 , ≤ 4.1 (custom)
    Create a notification for this product.
    Credits
    Joe Dillon reported these vulnerabilities to Philips.
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2025-2229",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "total"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2025-03-13T19:30:31.475698Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2025-03-13T19:30:42.059Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "Intellispace Cardiovascular (ISCV)",
              "vendor": "Philips",
              "versions": [
                {
                  "lessThanOrEqual": "4.1",
                  "status": "affected",
                  "version": "0",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Joe Dillon reported these vulnerabilities to Philips."
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "A token is created using the username, current date/time, and a fixed \nAES-128 encryption key, which is the same across all installations."
                }
              ],
              "value": "A token is created using the username, current date/time, and a fixed \nAES-128 encryption key, which is the same across all installations."
            }
          ],
          "metrics": [
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "LOCAL",
                "availabilityImpact": "NONE",
                "baseScore": 7.7,
                "baseSeverity": "HIGH",
                "confidentialityImpact": "HIGH",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "LOCAL",
                "baseScore": 8.5,
                "baseSeverity": "HIGH",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "NONE",
                "vulnConfidentialityImpact": "HIGH",
                "vulnIntegrityImpact": "HIGH",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-1391",
                  "description": "CWE-1391",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2025-03-17T20:04:56.100Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-072-01"
            },
            {
              "url": "https://www.philips.com/a-w/security/security-advisories.html"
            }
          ],
          "solutions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003ePhilips recommends the following mitigations:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eResolved in ISCV 4.2 build 20589, which was released in May 2019.\u003c/span\u003e\n\n\u003c/li\u003e\n\n\u003cli\u003ePhilips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\u003c/li\u003e\n\u003cli\u003ePlease contact a local Philips sales (service) representative to learn how to engage this upgrade process.\u003c/li\u003e\n\u003cli\u003eFor managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\u003c/li\u003e\n\u003c/ul\u003e\n\u003cp\u003eRefer to the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.philips.com/a-w/security/security-advisories.html\"\u003ePhilips advisory\u003c/a\u003efor more details.\n\n\u003cbr\u003e\u003c/p\u003e"
                }
              ],
              "value": "Philips recommends the following mitigations:\n\n\n\n  *  \n\nResolved in ISCV 4.2 build 20589, which was released in May 2019.\n\n\n\n\n  *  Philips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\n\n  *  Please contact a local Philips sales (service) representative to learn how to engage this upgrade process.\n\n  *  For managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\n\n\n\n\nRefer to the  Philips advisory https://www.philips.com/a-w/security/security-advisories.html for more details."
            }
          ],
          "source": {
            "advisory": "ICSMA-25-072-01",
            "discovery": "EXTERNAL"
          },
          "title": "Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials",
          "x_generator": {
            "engine": "Vulnogram 0.2.0"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2025-2229",
        "datePublished": "2025-03-13T18:17:31.256Z",
        "dateReserved": "2025-03-11T20:11:49.638Z",
        "dateUpdated": "2025-03-17T20:04:56.100Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }

    CVE-2025-2230 (GCVE-0-2025-2230)

    Vulnerability from cvelistv5 – Published: 2025-03-13 18:14 – Updated: 2025-03-17 20:05
    VLAI
    Title
    Philips Intellispace Cardiovascular (ISCV) Improper Authentication
    Summary
    A flaw exists in the Windows login flow where an AuthContext token can be exploited for replay attacks and authentication bypass.
    SSVC
    Exploitation: none Automatable: no Technical Impact: total
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-287 - Improper Authentication
    Assigner
    Impacted products
    Vendor Product Version
    Philips Intellispace Cardiovascular (ISCV) Affected: 0 , ≤ 5.1 (custom)
    Create a notification for this product.
    Credits
    Joe Dillon reported these vulnerabilities to Philips.
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2025-2230",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "total"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2025-03-13T19:33:45.514056Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2025-03-13T19:34:02.967Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unaffected",
              "product": "Intellispace Cardiovascular (ISCV)",
              "vendor": "Philips",
              "versions": [
                {
                  "lessThanOrEqual": "5.1",
                  "status": "affected",
                  "version": "0",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "credits": [
            {
              "lang": "en",
              "type": "finder",
              "value": "Joe Dillon reported these vulnerabilities to Philips."
            }
          ],
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "A flaw exists in the Windows login flow where an AuthContext token can \nbe exploited for replay attacks and authentication bypass."
                }
              ],
              "value": "A flaw exists in the Windows login flow where an AuthContext token can \nbe exploited for replay attacks and authentication bypass."
            }
          ],
          "metrics": [
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "LOCAL",
                "availabilityImpact": "NONE",
                "baseScore": 7.7,
                "baseSeverity": "HIGH",
                "confidentialityImpact": "HIGH",
                "integrityImpact": "HIGH",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "LOCAL",
                "baseScore": 8.5,
                "baseSeverity": "HIGH",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "NONE",
                "vulnConfidentialityImpact": "HIGH",
                "vulnIntegrityImpact": "HIGH",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-287",
                  "description": "CWE-287 Improper Authentication",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2025-03-17T20:05:56.939Z",
            "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
            "shortName": "icscert"
          },
          "references": [
            {
              "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-072-01"
            },
            {
              "url": "https://www.philips.com/a-w/security/security-advisories.html"
            }
          ],
          "solutions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003ePhilips recommends the following mitigations:\u003c/p\u003e\n\u003cul\u003e\n\u003cli\u003e\n\n\u003cspan style=\"background-color: rgb(255, 255, 255);\"\u003eResolved in ISCV 5.2, which was released in September 2020.\u003c/span\u003e\n\n\u003c/li\u003e\n\n\u003cli\u003ePhilips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\u003c/li\u003e\n\u003cli\u003ePlease contact a local Philips sales (service) representative to learn how to engage this upgrade process.\u003c/li\u003e\n\u003cli\u003eFor managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\u003c/li\u003e\n\u003c/ul\u003e\n\u003cp\u003eRefer to the \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.philips.com/a-w/security/security-advisories.html\"\u003ePhilips advisory\u003c/a\u003efor more details.\n\n\u003cbr\u003e\u003c/p\u003e"
                }
              ],
              "value": "Philips recommends the following mitigations:\n\n\n\n  *  \n\nResolved in ISCV 5.2, which was released in September 2020.\n\n\n\n\n  *  Philips recommends users upgrade ISCV installed base to the latest \nISCV version (at the time of this publication is 830089 \u2013 IntelliSpace \nCardiovacular 8.0.0.0)\n\n  *  Please contact a local Philips sales (service) representative to learn how to engage this upgrade process.\n\n  *  For managed services users, new releases will be made available upon\n resource availability. Releases are subject to country-specific \nregulations.\n\n\n\n\nRefer to the  Philips advisory https://www.philips.com/a-w/security/security-advisories.html for more details."
            }
          ],
          "source": {
            "advisory": "ICSMA-25-072-01",
            "discovery": "EXTERNAL"
          },
          "title": "Philips Intellispace Cardiovascular (ISCV) Improper Authentication",
          "x_generator": {
            "engine": "Vulnogram 0.2.0"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "assignerShortName": "icscert",
        "cveId": "CVE-2025-2230",
        "datePublished": "2025-03-13T18:14:43.573Z",
        "dateReserved": "2025-03-11T20:14:50.017Z",
        "dateUpdated": "2025-03-17T20:05:56.939Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }