    4 vulnerabilities found for Infinity M300 by Dräger

    CVE-2019-25724 (GCVE-0-2019-25724)

    Vulnerability from nvd – Published: 2026-06-02 19:11 – Updated: 2026-06-03 18:13
    VLAI
    Title
    Dräger Infinity M300 VG2.x Network-Based Denial of Service
    Summary
    Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-400 - Uncontrolled Resource Consumption
    Assigner
    References
    Impacted products
    Vendor Product Version
    Dräger Infinity M300 Affected: VG2.x , ≤ VG2.3.1 (custom)
    Unaffected: VG3.0 (custom)
    Create a notification for this product.
    Date Public
    2020-07-24 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2019-25724",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-03T14:37:28.888759Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-03T14:38:43.975Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unknown",
              "product": "Infinity M300",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "lessThanOrEqual": "VG2.3.1",
                  "status": "affected",
                  "version": "VG2.x",
                  "versionType": "custom"
                },
                {
                  "status": "unaffected",
                  "version": "VG3.0",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "datePublic": "2020-07-24T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eDr\u00e4ger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.\u003c/p\u003e"
                }
              ],
              "value": "Dr\u00e4ger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "ADJACENT",
                "baseScore": 7.1,
                "baseSeverity": "HIGH",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "HIGH",
                "vulnConfidentialityImpact": "NONE",
                "vulnIntegrityImpact": "NONE",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "ADJACENT_NETWORK",
                "availabilityImpact": "HIGH",
                "baseScore": 6.5,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "NONE",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-400",
                  "description": "CWE-400 Uncontrolled Resource Consumption",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-03T18:13:43.182Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security/download/2019-277-02-M300-VG2x-Security-Advisory.pdf"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-x-network-based-denial-of-service"
            }
          ],
          "source": {
            "discovery": "INTERNAL"
          },
          "title": "Dr\u00e4ger Infinity M300 VG2.x Network-Based Denial of Service"
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2019-25724",
        "datePublished": "2026-06-02T19:11:51.652Z",
        "dateReserved": "2026-06-02T19:05:51.385Z",
        "dateUpdated": "2026-06-03T18:13:43.182Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2019-25721 (GCVE-0-2019-25721)

    Vulnerability from nvd – Published: 2026-06-02 17:29 – Updated: 2026-06-03 16:06
    VLAI
    Title
    Dräger Infinity M300 VG2.3.1 Network-Based Denial of Service
    Summary
    Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-400 - Uncontrolled Resource Consumption
    Assigner
    References
    Impacted products
    Vendor Product Version
    Dräger Infinity M300 Affected: 0 , ≤ VG2.3.1 (custom)
    Unaffected: VG2.3.2 (custom)
    Create a notification for this product.
    Date Public
    2019-10-09 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2019-25721",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-03T12:52:13.424163Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-03T12:52:26.546Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unknown",
              "product": "Infinity M300",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "lessThanOrEqual": "VG2.3.1",
                  "status": "affected",
                  "version": "0",
                  "versionType": "custom"
                },
                {
                  "status": "unaffected",
                  "version": "VG2.3.2",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "datePublic": "2019-10-09T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eDr\u00e4ger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.\u003c/p\u003e"
                }
              ],
              "value": "Dr\u00e4ger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "ADJACENT",
                "baseScore": 7.1,
                "baseSeverity": "HIGH",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "HIGH",
                "vulnConfidentialityImpact": "NONE",
                "vulnIntegrityImpact": "NONE",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "ADJACENT_NETWORK",
                "availabilityImpact": "HIGH",
                "baseScore": 6.5,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "NONE",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-400",
                  "description": "CWE-400 Uncontrolled Resource Consumption",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-03T16:06:32.703Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security/download/2019-09-13-Cybersecurity-M300-VGxx-Security-advisory-update_v1-3.pdf"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-network-based-denial-of-service"
            }
          ],
          "source": {
            "discovery": "INTERNAL"
          },
          "title": "Dr\u00e4ger Infinity M300 VG2.3.1 Network-Based Denial of Service",
          "x_generator": {
            "engine": "Vulnogram 1.0.2"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2019-25721",
        "datePublished": "2026-06-02T17:29:08.696Z",
        "dateReserved": "2026-06-02T17:23:43.821Z",
        "dateUpdated": "2026-06-03T16:06:32.703Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2019-25724 (GCVE-0-2019-25724)

    Vulnerability from cvelistv5 – Published: 2026-06-02 19:11 – Updated: 2026-06-03 18:13
    VLAI
    Title
    Dräger Infinity M300 VG2.x Network-Based Denial of Service
    Summary
    Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-400 - Uncontrolled Resource Consumption
    Assigner
    References
    Impacted products
    Vendor Product Version
    Dräger Infinity M300 Affected: VG2.x , ≤ VG2.3.1 (custom)
    Unaffected: VG3.0 (custom)
    Create a notification for this product.
    Date Public
    2020-07-24 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2019-25724",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-03T14:37:28.888759Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-03T14:38:43.975Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unknown",
              "product": "Infinity M300",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "lessThanOrEqual": "VG2.3.1",
                  "status": "affected",
                  "version": "VG2.x",
                  "versionType": "custom"
                },
                {
                  "status": "unaffected",
                  "version": "VG3.0",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "datePublic": "2020-07-24T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eDr\u00e4ger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.\u003c/p\u003e"
                }
              ],
              "value": "Dr\u00e4ger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "ADJACENT",
                "baseScore": 7.1,
                "baseSeverity": "HIGH",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "HIGH",
                "vulnConfidentialityImpact": "NONE",
                "vulnIntegrityImpact": "NONE",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "ADJACENT_NETWORK",
                "availabilityImpact": "HIGH",
                "baseScore": 6.5,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "NONE",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-400",
                  "description": "CWE-400 Uncontrolled Resource Consumption",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-03T18:13:43.182Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security/download/2019-277-02-M300-VG2x-Security-Advisory.pdf"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-x-network-based-denial-of-service"
            }
          ],
          "source": {
            "discovery": "INTERNAL"
          },
          "title": "Dr\u00e4ger Infinity M300 VG2.x Network-Based Denial of Service"
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2019-25724",
        "datePublished": "2026-06-02T19:11:51.652Z",
        "dateReserved": "2026-06-02T19:05:51.385Z",
        "dateUpdated": "2026-06-03T18:13:43.182Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2019-25721 (GCVE-0-2019-25721)

    Vulnerability from cvelistv5 – Published: 2026-06-02 17:29 – Updated: 2026-06-03 16:06
    VLAI
    Title
    Dräger Infinity M300 VG2.3.1 Network-Based Denial of Service
    Summary
    Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-400 - Uncontrolled Resource Consumption
    Assigner
    References
    Impacted products
    Vendor Product Version
    Dräger Infinity M300 Affected: 0 , ≤ VG2.3.1 (custom)
    Unaffected: VG2.3.2 (custom)
    Create a notification for this product.
    Date Public
    2019-10-09 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2019-25721",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-03T12:52:13.424163Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-03T12:52:26.546Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unknown",
              "product": "Infinity M300",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "lessThanOrEqual": "VG2.3.1",
                  "status": "affected",
                  "version": "0",
                  "versionType": "custom"
                },
                {
                  "status": "unaffected",
                  "version": "VG2.3.2",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "datePublic": "2019-10-09T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "supportingMedia": [
                {
                  "base64": false,
                  "type": "text/html",
                  "value": "\u003cp\u003eDr\u00e4ger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.\u003c/p\u003e"
                }
              ],
              "value": "Dr\u00e4ger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "LOW",
                "attackRequirements": "NONE",
                "attackVector": "ADJACENT",
                "baseScore": 7.1,
                "baseSeverity": "HIGH",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "NONE",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
                "version": "4.0",
                "vulnAvailabilityImpact": "HIGH",
                "vulnConfidentialityImpact": "NONE",
                "vulnIntegrityImpact": "NONE",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "LOW",
                "attackVector": "ADJACENT_NETWORK",
                "availabilityImpact": "HIGH",
                "baseScore": 6.5,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "NONE",
                "privilegesRequired": "NONE",
                "scope": "UNCHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-400",
                  "description": "CWE-400 Uncontrolled Resource Consumption",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-03T16:06:32.703Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security/download/2019-09-13-Cybersecurity-M300-VGxx-Security-advisory-update_v1-3.pdf"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-network-based-denial-of-service"
            }
          ],
          "source": {
            "discovery": "INTERNAL"
          },
          "title": "Dr\u00e4ger Infinity M300 VG2.3.1 Network-Based Denial of Service",
          "x_generator": {
            "engine": "Vulnogram 1.0.2"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2019-25721",
        "datePublished": "2026-06-02T17:29:08.696Z",
        "dateReserved": "2026-06-02T17:23:43.821Z",
        "dateUpdated": "2026-06-03T16:06:32.703Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }