4 vulnerabilities found for Infinity M300 by Dräger
CVE-2019-25724 (GCVE-0-2019-25724)
Vulnerability from nvd – Published: 2026-06-02 19:11 – Updated: 2026-06-03 18:13
VLAI
Title
Dräger Infinity M300 VG2.x Network-Based Denial of Service
Summary
Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.
Severity
SSVC
Exploitation: none
Automatable: no
Technical Impact: partial
CISA Coordinator (v2.0.3)
CWE
- CWE-400 - Uncontrolled Resource Consumption
Assigner
References
2 references
| URL | Tags |
|---|---|
| https://static.draeger.com/security/download/2019… | vendor-advisory |
| https://www.vulncheck.com/advisories/dr-ger-infin… | third-party-advisory |
Impacted products
1 product
| Vendor | Product | Version | |
|---|---|---|---|
| Dräger | Infinity M300 |
Affected:
VG2.x , ≤ VG2.3.1
(custom)
Unaffected: VG3.0 (custom) |
Date Public
2020-07-24 00:00
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2019-25724",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-06-03T14:37:28.888759Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T14:38:43.975Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unknown",
"product": "Infinity M300",
"vendor": "Dr\u00e4ger",
"versions": [
{
"lessThanOrEqual": "VG2.3.1",
"status": "affected",
"version": "VG2.x",
"versionType": "custom"
},
{
"status": "unaffected",
"version": "VG3.0",
"versionType": "custom"
}
]
}
],
"datePublic": "2020-07-24T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eDr\u00e4ger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.\u003c/p\u003e"
}
],
"value": "Dr\u00e4ger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "ADJACENT",
"baseScore": 7.1,
"baseSeverity": "HIGH",
"exploitMaturity": "NOT_DEFINED",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "NONE",
"vulnIntegrityImpact": "NONE",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 6.5,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-400",
"description": "CWE-400 Uncontrolled Resource Consumption",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T18:13:43.182Z",
"orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"shortName": "VulnCheck"
},
"references": [
{
"tags": [
"vendor-advisory"
],
"url": "https://static.draeger.com/security/download/2019-277-02-M300-VG2x-Security-Advisory.pdf"
},
{
"tags": [
"third-party-advisory"
],
"url": "https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-x-network-based-denial-of-service"
}
],
"source": {
"discovery": "INTERNAL"
},
"title": "Dr\u00e4ger Infinity M300 VG2.x Network-Based Denial of Service"
}
},
"cveMetadata": {
"assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"assignerShortName": "VulnCheck",
"cveId": "CVE-2019-25724",
"datePublished": "2026-06-02T19:11:51.652Z",
"dateReserved": "2026-06-02T19:05:51.385Z",
"dateUpdated": "2026-06-03T18:13:43.182Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
CVE-2019-25721 (GCVE-0-2019-25721)
Vulnerability from nvd – Published: 2026-06-02 17:29 – Updated: 2026-06-03 16:06
VLAI
Title
Dräger Infinity M300 VG2.3.1 Network-Based Denial of Service
Summary
Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.
Severity
SSVC
Exploitation: none
Automatable: no
Technical Impact: partial
CISA Coordinator (v2.0.3)
CWE
- CWE-400 - Uncontrolled Resource Consumption
Assigner
References
2 references
| URL | Tags |
|---|---|
| https://static.draeger.com/security/download/2019… | vendor-advisory |
| https://www.vulncheck.com/advisories/dr-ger-infin… | third-party-advisory |
Impacted products
1 product
| Vendor | Product | Version | |
|---|---|---|---|
| Dräger | Infinity M300 |
Affected:
0 , ≤ VG2.3.1
(custom)
Unaffected: VG2.3.2 (custom) |
Date Public
2019-10-09 00:00
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2019-25721",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-06-03T12:52:13.424163Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T12:52:26.546Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unknown",
"product": "Infinity M300",
"vendor": "Dr\u00e4ger",
"versions": [
{
"lessThanOrEqual": "VG2.3.1",
"status": "affected",
"version": "0",
"versionType": "custom"
},
{
"status": "unaffected",
"version": "VG2.3.2",
"versionType": "custom"
}
]
}
],
"datePublic": "2019-10-09T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eDr\u00e4ger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.\u003c/p\u003e"
}
],
"value": "Dr\u00e4ger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "ADJACENT",
"baseScore": 7.1,
"baseSeverity": "HIGH",
"exploitMaturity": "NOT_DEFINED",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "NONE",
"vulnIntegrityImpact": "NONE",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 6.5,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-400",
"description": "CWE-400 Uncontrolled Resource Consumption",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T16:06:32.703Z",
"orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"shortName": "VulnCheck"
},
"references": [
{
"tags": [
"vendor-advisory"
],
"url": "https://static.draeger.com/security/download/2019-09-13-Cybersecurity-M300-VGxx-Security-advisory-update_v1-3.pdf"
},
{
"tags": [
"third-party-advisory"
],
"url": "https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-network-based-denial-of-service"
}
],
"source": {
"discovery": "INTERNAL"
},
"title": "Dr\u00e4ger Infinity M300 VG2.3.1 Network-Based Denial of Service",
"x_generator": {
"engine": "Vulnogram 1.0.2"
}
}
},
"cveMetadata": {
"assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"assignerShortName": "VulnCheck",
"cveId": "CVE-2019-25721",
"datePublished": "2026-06-02T17:29:08.696Z",
"dateReserved": "2026-06-02T17:23:43.821Z",
"dateUpdated": "2026-06-03T16:06:32.703Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
CVE-2019-25724 (GCVE-0-2019-25724)
Vulnerability from cvelistv5 – Published: 2026-06-02 19:11 – Updated: 2026-06-03 18:13
VLAI
Title
Dräger Infinity M300 VG2.x Network-Based Denial of Service
Summary
Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.
Severity
SSVC
Exploitation: none
Automatable: no
Technical Impact: partial
CISA Coordinator (v2.0.3)
CWE
- CWE-400 - Uncontrolled Resource Consumption
Assigner
References
2 references
| URL | Tags |
|---|---|
| https://static.draeger.com/security/download/2019… | vendor-advisory |
| https://www.vulncheck.com/advisories/dr-ger-infin… | third-party-advisory |
Impacted products
1 product
| Vendor | Product | Version | |
|---|---|---|---|
| Dräger | Infinity M300 |
Affected:
VG2.x , ≤ VG2.3.1
(custom)
Unaffected: VG3.0 (custom) |
Date Public
2020-07-24 00:00
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2019-25724",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-06-03T14:37:28.888759Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T14:38:43.975Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unknown",
"product": "Infinity M300",
"vendor": "Dr\u00e4ger",
"versions": [
{
"lessThanOrEqual": "VG2.3.1",
"status": "affected",
"version": "VG2.x",
"versionType": "custom"
},
{
"status": "unaffected",
"version": "VG3.0",
"versionType": "custom"
}
]
}
],
"datePublic": "2020-07-24T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eDr\u00e4ger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.\u003c/p\u003e"
}
],
"value": "Dr\u00e4ger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "ADJACENT",
"baseScore": 7.1,
"baseSeverity": "HIGH",
"exploitMaturity": "NOT_DEFINED",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "NONE",
"vulnIntegrityImpact": "NONE",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 6.5,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-400",
"description": "CWE-400 Uncontrolled Resource Consumption",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T18:13:43.182Z",
"orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"shortName": "VulnCheck"
},
"references": [
{
"tags": [
"vendor-advisory"
],
"url": "https://static.draeger.com/security/download/2019-277-02-M300-VG2x-Security-Advisory.pdf"
},
{
"tags": [
"third-party-advisory"
],
"url": "https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-x-network-based-denial-of-service"
}
],
"source": {
"discovery": "INTERNAL"
},
"title": "Dr\u00e4ger Infinity M300 VG2.x Network-Based Denial of Service"
}
},
"cveMetadata": {
"assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"assignerShortName": "VulnCheck",
"cveId": "CVE-2019-25724",
"datePublished": "2026-06-02T19:11:51.652Z",
"dateReserved": "2026-06-02T19:05:51.385Z",
"dateUpdated": "2026-06-03T18:13:43.182Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
CVE-2019-25721 (GCVE-0-2019-25721)
Vulnerability from cvelistv5 – Published: 2026-06-02 17:29 – Updated: 2026-06-03 16:06
VLAI
Title
Dräger Infinity M300 VG2.3.1 Network-Based Denial of Service
Summary
Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.
Severity
SSVC
Exploitation: none
Automatable: no
Technical Impact: partial
CISA Coordinator (v2.0.3)
CWE
- CWE-400 - Uncontrolled Resource Consumption
Assigner
References
2 references
| URL | Tags |
|---|---|
| https://static.draeger.com/security/download/2019… | vendor-advisory |
| https://www.vulncheck.com/advisories/dr-ger-infin… | third-party-advisory |
Impacted products
1 product
| Vendor | Product | Version | |
|---|---|---|---|
| Dräger | Infinity M300 |
Affected:
0 , ≤ VG2.3.1
(custom)
Unaffected: VG2.3.2 (custom) |
Date Public
2019-10-09 00:00
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2019-25721",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-06-03T12:52:13.424163Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T12:52:26.546Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unknown",
"product": "Infinity M300",
"vendor": "Dr\u00e4ger",
"versions": [
{
"lessThanOrEqual": "VG2.3.1",
"status": "affected",
"version": "0",
"versionType": "custom"
},
{
"status": "unaffected",
"version": "VG2.3.2",
"versionType": "custom"
}
]
}
],
"datePublic": "2019-10-09T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eDr\u00e4ger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.\u003c/p\u003e"
}
],
"value": "Dr\u00e4ger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "ADJACENT",
"baseScore": 7.1,
"baseSeverity": "HIGH",
"exploitMaturity": "NOT_DEFINED",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "NONE",
"vulnIntegrityImpact": "NONE",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "ADJACENT_NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 6.5,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-400",
"description": "CWE-400 Uncontrolled Resource Consumption",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T16:06:32.703Z",
"orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"shortName": "VulnCheck"
},
"references": [
{
"tags": [
"vendor-advisory"
],
"url": "https://static.draeger.com/security/download/2019-09-13-Cybersecurity-M300-VGxx-Security-advisory-update_v1-3.pdf"
},
{
"tags": [
"third-party-advisory"
],
"url": "https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-network-based-denial-of-service"
}
],
"source": {
"discovery": "INTERNAL"
},
"title": "Dr\u00e4ger Infinity M300 VG2.3.1 Network-Based Denial of Service",
"x_generator": {
"engine": "Vulnogram 1.0.2"
}
}
},
"cveMetadata": {
"assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"assignerShortName": "VulnCheck",
"cveId": "CVE-2019-25721",
"datePublished": "2026-06-02T17:29:08.696Z",
"dateReserved": "2026-06-02T17:23:43.821Z",
"dateUpdated": "2026-06-03T16:06:32.703Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}