Search

Find a vulnerability

Search criteria

    2 vulnerabilities found for Atlan A350 by Dräger

    CVE-2021-4479 (GCVE-0-2021-4479)

    Vulnerability from nvd – Published: 2026-06-02 19:27 – Updated: 2026-06-10 15:59
    VLAI
    Title
    Dräger Atlan A350 1.00 <= 1.01 DoS via Medibus Interface
    Summary
    Dräger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-1286 - Improper Validation of Syntactic Correctness of Input
    Assigner
    References
    Impacted products
    Vendor Product Version
    Dräger Atlan A350 Affected: 1.00 , ≤ 1.01 (custom)
    Create a notification for this product.
    Date Public
    2021-05-04 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2021-4479",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-03T12:50:30.624302Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-03T12:50:37.749Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unknown",
              "product": "Atlan A350",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "lessThanOrEqual": "1.01",
                  "status": "affected",
                  "version": "1.00",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "datePublic": "2021-05-04T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Dr\u00e4ger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "HIGH",
                "attackRequirements": "NONE",
                "attackVector": "NETWORK",
                "baseScore": 6.3,
                "baseSeverity": "MEDIUM",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "LOW",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L",
                "version": "4.0",
                "vulnAvailabilityImpact": "LOW",
                "vulnConfidentialityImpact": "NONE",
                "vulnIntegrityImpact": "NONE",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "HIGH",
                "attackVector": "NETWORK",
                "availabilityImpact": "LOW",
                "baseScore": 4,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "NONE",
                "privilegesRequired": "NONE",
                "scope": "CHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-1286",
                  "description": "Improper Validation of Syntactic Correctness of Input",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-10T15:59:05.558Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security/download/2021-05-04-PSA-21-124-1-Security-Advisory-Atlan.pdf"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/drager-atlan-a350-dos-via-medibus-interface"
            }
          ],
          "source": {
            "discovery": "INTERNAL"
          },
          "title": "Dr\u00e4ger Atlan A350 1.00 \u003c= 1.01 DoS via Medibus Interface",
          "x_generator": {
            "engine": "Vulnogram 1.0.2"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2021-4479",
        "datePublished": "2026-06-02T19:27:00.318Z",
        "dateReserved": "2026-06-02T19:20:57.660Z",
        "dateUpdated": "2026-06-10T15:59:05.558Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }

    CVE-2021-4479 (GCVE-0-2021-4479)

    Vulnerability from cvelistv5 – Published: 2026-06-02 19:27 – Updated: 2026-06-10 15:59
    VLAI
    Title
    Dräger Atlan A350 1.00 <= 1.01 DoS via Medibus Interface
    Summary
    Dräger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.
    SSVC
    Exploitation: none Automatable: no Technical Impact: partial
    CISA Coordinator (v2.0.3)
    CWE
    • CWE-1286 - Improper Validation of Syntactic Correctness of Input
    Assigner
    References
    Impacted products
    Vendor Product Version
    Dräger Atlan A350 Affected: 1.00 , ≤ 1.01 (custom)
    Create a notification for this product.
    Date Public
    2021-05-04 00:00
    Show details on NVD website

    {
      "containers": {
        "adp": [
          {
            "metrics": [
              {
                "other": {
                  "content": {
                    "id": "CVE-2021-4479",
                    "options": [
                      {
                        "Exploitation": "none"
                      },
                      {
                        "Automatable": "no"
                      },
                      {
                        "Technical Impact": "partial"
                      }
                    ],
                    "role": "CISA Coordinator",
                    "timestamp": "2026-06-03T12:50:30.624302Z",
                    "version": "2.0.3"
                  },
                  "type": "ssvc"
                }
              }
            ],
            "providerMetadata": {
              "dateUpdated": "2026-06-03T12:50:37.749Z",
              "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
              "shortName": "CISA-ADP"
            },
            "title": "CISA ADP Vulnrichment"
          }
        ],
        "cna": {
          "affected": [
            {
              "defaultStatus": "unknown",
              "product": "Atlan A350",
              "vendor": "Dr\u00e4ger",
              "versions": [
                {
                  "lessThanOrEqual": "1.01",
                  "status": "affected",
                  "version": "1.00",
                  "versionType": "custom"
                }
              ]
            }
          ],
          "datePublic": "2021-05-04T00:00:00.000Z",
          "descriptions": [
            {
              "lang": "en",
              "value": "Dr\u00e4ger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves."
            }
          ],
          "metrics": [
            {
              "cvssV4_0": {
                "Automatable": "NOT_DEFINED",
                "Recovery": "NOT_DEFINED",
                "Safety": "NOT_DEFINED",
                "attackComplexity": "HIGH",
                "attackRequirements": "NONE",
                "attackVector": "NETWORK",
                "baseScore": 6.3,
                "baseSeverity": "MEDIUM",
                "exploitMaturity": "NOT_DEFINED",
                "privilegesRequired": "NONE",
                "providerUrgency": "NOT_DEFINED",
                "subAvailabilityImpact": "LOW",
                "subConfidentialityImpact": "NONE",
                "subIntegrityImpact": "NONE",
                "userInteraction": "NONE",
                "valueDensity": "NOT_DEFINED",
                "vectorString": "CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L",
                "version": "4.0",
                "vulnAvailabilityImpact": "LOW",
                "vulnConfidentialityImpact": "NONE",
                "vulnIntegrityImpact": "NONE",
                "vulnerabilityResponseEffort": "NOT_DEFINED"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            },
            {
              "cvssV3_1": {
                "attackComplexity": "HIGH",
                "attackVector": "NETWORK",
                "availabilityImpact": "LOW",
                "baseScore": 4,
                "baseSeverity": "MEDIUM",
                "confidentialityImpact": "NONE",
                "integrityImpact": "NONE",
                "privilegesRequired": "NONE",
                "scope": "CHANGED",
                "userInteraction": "NONE",
                "vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L",
                "version": "3.1"
              },
              "format": "CVSS",
              "scenarios": [
                {
                  "lang": "en",
                  "value": "GENERAL"
                }
              ]
            }
          ],
          "problemTypes": [
            {
              "descriptions": [
                {
                  "cweId": "CWE-1286",
                  "description": "Improper Validation of Syntactic Correctness of Input",
                  "lang": "en",
                  "type": "CWE"
                }
              ]
            }
          ],
          "providerMetadata": {
            "dateUpdated": "2026-06-10T15:59:05.558Z",
            "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
            "shortName": "VulnCheck"
          },
          "references": [
            {
              "tags": [
                "vendor-advisory"
              ],
              "url": "https://static.draeger.com/security/download/2021-05-04-PSA-21-124-1-Security-Advisory-Atlan.pdf"
            },
            {
              "tags": [
                "third-party-advisory"
              ],
              "url": "https://www.vulncheck.com/advisories/drager-atlan-a350-dos-via-medibus-interface"
            }
          ],
          "source": {
            "discovery": "INTERNAL"
          },
          "title": "Dr\u00e4ger Atlan A350 1.00 \u003c= 1.01 DoS via Medibus Interface",
          "x_generator": {
            "engine": "Vulnogram 1.0.2"
          }
        }
      },
      "cveMetadata": {
        "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "assignerShortName": "VulnCheck",
        "cveId": "CVE-2021-4479",
        "datePublished": "2026-06-02T19:27:00.318Z",
        "dateReserved": "2026-06-02T19:20:57.660Z",
        "dateUpdated": "2026-06-10T15:59:05.558Z",
        "state": "PUBLISHED"
      },
      "dataType": "CVE_RECORD",
      "dataVersion": "5.2"
    }